| Literature DB >> 28944210 |
Jaroslavas Belovickis1, Aliaksei Kurylenka2, Vadim Murashko2.
Abstract
We report on the impact of direct ultraviolet germicidal irradiation (UVGI) on reflective optics, used in the excimer laser system Allegretto Eye-Q. The aim of our work was to confirm our hypothesis based on long-rate observations of obtained anomalies in post-operative results that are attributed to degradation of reflective optics upon ultraviolet radiation. The presence of direct UVGI coupled with humidity in the operating environment caused merging anomalies and unwanted post-operative correction values. Ultraviolet-A radiation caused a similar effect on the reflective cover of the mirrors.Entities:
Keywords: excimer lasers; laser-assisted in situ keratomileusis; mirror degradation; post-operative anomalies; ultraviolet radiation
Year: 2017 PMID: 28944210 PMCID: PMC5596236 DOI: 10.18240/ijo.2017.09.22