Abstract
An assay has been developed and validated for the routine monitoring of vecuronium in plasma. It consists of solid-phase extraction using C18 disposables as sample pre-treatment, high-performance liquid chromatography and post-column ion-pair extraction with fluorimetric detection. The fluorescent anion 9,10-dimethoxyanthracene-2-sulphonate (DAS) is used as the counter ion. The detection limit is ca. 5 ng/ml in plasma with a signal-to-noise ratio of 10. The assay is also applicable for monitoring vecuronium and its potential metabolites in other biological media, e.g., urine, bile and tissue (liver, kidney) homogenates.Entities:
Mesh:
Substances:
Year: 1987 PMID: 2892850 DOI: 10.1016/0378-4347(87)80412-7
Source DB: PubMed Journal: J Chromatogr