Abstract

PURPOSE: This research investigated whether the Sprotte needle causes less leakage of CSF than the Quincke needle in the artificial spinal cord.
METHODS: The changes in intradural pressure, extradural pressure, and leaked volume of CSF were evaluated following puncture with Sprotte and Quincke needles in the artificial spinal cord.
RESULTS: The decrease in intradural pressure was 9.7±1.8 mm H2O with the Sprotte needle and 20.5±2.7 mm H2O with the Quincke needle (P<0.05). The volume of leakage of artificial CSF was 2.0±0.3 ml with the Sprotte needle and 3.3 ±0.3 ml with the Quincke needle (P<0.01). The extradural pressure increase was 166.1±8.2 mm H2O with the Sprotte needle and 186.8±13.2 mm H2O with the Quincke needle (P<0.05).
CONCLUSION: The Sprotte needle produces less CSF leakage than the Quincke needle.

Entities:  

Keywords:  Spinal anesthesia; Spinal cord; Spinal fluid

Year:  1997        PMID: 28921068     DOI: 10.1007/BF02480746

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  7 in total

1.  Long-term follow-up of patients who received 10,098 spinal anesthetics; syndrome of decreased intracranial pressure (headache and ocular and auditory difficulties).

Authors:  L D VANDAM; R D DRIPPS
Journal:  J Am Med Assoc       Date:  1956-06-16

2.  Headache after spinal anesthesia: experiments with a new spinal needle.

Authors:  S HARALDSON
Journal:  Anesthesiology       Date:  1951-05       Impact factor: 7.892

3.  Sprotte needle for obstetric anesthesia: decreased incidence of post dural puncture headache.

Authors:  B K Ross; H S Chadwick; J J Mancuso; C Benedetti
Journal:  Reg Anesth       Date:  1992 Jan-Feb

4.  Postdural puncture headache: comparison of 25-gauge Whitacre and Quincke needles.

Authors:  J Buettner; K P Wresch; R Klose
Journal:  Reg Anesth       Date:  1993 May-Jun

5.  [An "atraumatic" universal needle for single-shot regional anesthesia: clinical results and a 6 year trial in over 30,000 regional anesthesias].

Authors:  G Sprotte; R Schedel; H Pajunk; H Pajunk
Journal:  Reg Anaesth       Date:  1987-07

6.  Postdural puncture headache and spinal needle design. Metaanalyses.

Authors:  S Halpern; R Preston
Journal:  Anesthesiology       Date:  1994-12       Impact factor: 7.892

7.  Posture and post-spinal headache. A controlled trial in 80 obstetric patients.

Authors:  E A Thornberry; T A Thomas
Journal:  Br J Anaesth       Date:  1988-02       Impact factor: 9.166

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.