| Literature DB >> 28900364 |
Alessandro De Stefano1, Liliana Mannucci1, Renato Massoud1, Sergio Bernardini1, Claudio Cortese1.
Abstract
INTRODUCTION: Lipoprotein-associated phospholipase A2 (Lp-PLA2) is a marker of vascular inflammation associated with coronary heart disease and stroke. We evaluated analytical performance of the PLAC® Activity Test on Siemens Dimension Vista® 1500 analyzer and measured Lp-PLA2 activity in Italian adults to establish reference intervals (RIs) and evaluate correlation with circulating lipids and age.Entities:
Keywords: Darapladib; Lp-PLA2; assay validation; cardiovascular disease; reference intervals
Mesh:
Substances:
Year: 2017 PMID: 28900364 PMCID: PMC5575650 DOI: 10.11613/BM.2017.030701
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Precision evaluation of the lipoprotein-associated phospholipase A2 activity assay on the Dimension Vista analyzer
| 121 (1.5) | 1.2 | - | 121 (2.4) | 2.0 | - | |
| 295 (2.5) | 0.8 | - | 301 (4.1) | 1.4 | - | |
| 98 (0.6) | 0.6 | - | 99 (0.9) | 0.9 | - | |
| 200 (2.9) | 1.4 | - | 201 (2.8) | 1.4 | - | |
| 254 (2.4) | 0.9 | - | 258 (3.0) | 1.2 | - | |
| 98 – 295 (0.6 - 2.9) | 0.6 - 1.4 | 1.3 - 1.4 | 99 – 301 (0.9 – 4.1) | 0.9 - 2.0 | 1.9 - 2.6 | |
| Lp-PLA2 - lipoprotein-associated phospholipase A2. Lp-PLA2 activities are presented as mean (standard deviation). CV – coefficient of variation. For intra-assay imprecision evaluation, 20 replicates of each sample were measured in a single assay run. For inter-assay imprecision evaluation, 4 replicates of each sample were measured on 5 different days. Ranges of intra- and inter-assay %CVs obtained by the manufacturer on a Beckman Coulter AU400 analyzer (package insert) are also shown. | ||||||
Linearity evaluation of the lipoprotein-associated phospholipase A2 activity assay on the Dimension Vista analyzer
| 299 | 96 | 1.01 | - 1.13 | 0.999 | Slope: 0.98 - 1.03 | |
| 316 | 97 | 1.01 | - 0.17 | 0.999 | Intercept: - 6.8 - 10.7 | |
| 328 | 97 | 1.01 | 3.10 | 0.998 | R2: 0.995 - 0.999 | |
| Three pairs of serum samples with high and low Lp-PLA2 activities were used to obtain 11 Lp-PLA2 activity levels in a serial recovery assay. Lp-PLA2 activity was measured in duplicate in each level and mean values were compared to expected activity values. Linearity results obtained by the manufacturer on a Beckman Coulter AU400 analyzer (package insert) are also shown. | ||||||
Figure 1Comparison of lipoprotein-associated phospholipase A2 (Lp-PLA2) activity values (N = 40) obtained on VISTA and AU400. (A) In the Bland-Altman plot the solid line shows mean difference, while the dashed lines show the ± 1.96 SD. (B) In the Passing-Bablok regression analysis, the dashed lines show the 95% confidence interval (CI), while dotted line represents identity (X = Y).
Figure 2Hook effect evaluation of the lipoprotein-associated phospholipase A2 activity assay on the Dimension Vista analyser. Substrate depletion hooks back measured ODs only at recombinant lipoprotein-associated phospholipase A2 (Lp-PLA2) activities > 1000 U/L.
Main clinical and laboratory characteristics of the study cohort of healthy donors
| 123 | |
|---|---|
| 127 | |
| 37 (18 - 70) | |
| 4.7 ± 0.5 | |
| 0.9 ± 0.3 | |
| 4.3 ± 0.6 | |
| 2.8 ± 0.5 | |
| 1.4 ± 0.3 | |
| Age is presented as median (min-max). Results are presented as mean ± SD. | |
Lipoprotein-associated phospholipase A2 activity values in adult healthy subjects
| Entire cohort | Males | Females | |
| 71 | 87 | 71 | |
| 93 | 107 | 84 | |
| 106 | 135 | 97 | |
| 144 | 171 | 128 | |
| 164 | 187 | 145 | |
| 183 | 205 | 165 | |
| 203 | 224 | 179 | |
| 220 | 241 | 192 | |
| 254 | 260 | 217 | |
| 262 | 265 | 225 | |
| 290 | 290 | 262 | |
| 182 | 203 | 161 | |
| 44 | 40 | 37 | |
| 93-262 | 107-265 | 84-225 | |
| Lp-PLA2 - lipoprotein-associated phospholipase A2. | |||