| Literature DB >> 28894737 |
Cristiana Vidali1, Mara Severgnini2, Monica Urbani3, Licia Toscano4, Alfredo Perulli5, Marina Bortul3.
Abstract
INTRODUCTION: Failure Mode Effects and Criticalities Analysis (FMECA) represents a prospective method for risk assessment in complex medical practices. Our objective was to describe the application of FMECA approach to intraoperative electron beam radiotherapy (IOERT), delivered using a mobile linear accelerator, for the treatment of early breast cancer as an anticipated boost.Entities:
Keywords: FMECA; intraoperative electron beam radiotherapy; patient safety; quality assurance; risk assessment
Year: 2017 PMID: 28894737 PMCID: PMC5581388 DOI: 10.3389/fmed.2017.00138
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
FLOW CHART: processes identified in intraoperative electron beam radiotherapy (IOERT) treatment in Trieste.
| Phases | Professional figures | Procedure | Initial risk ranking | Revised risk ranking 2013–2014 | Final revised risk ranking 2015–2016 |
|---|---|---|---|---|---|
| Risk priority number (RPN) | RPN | RPN | |||
| 1 | Physicist | MOBETRON transfer, start-up and warm-up | 45 | 5 | 5 |
| 30 | 10 | 10 | |||
| 36 | 24 | 24 | |||
| 2 | Physicist | MOBETRON daily Mechanical check | 30 | 10 | 10 |
| 3 | Physicist | MOBETRON daily dosimetric quality control | 60 | 30 | 5 |
| 4 | Nurse of the operating room | Set-up of dedicated IOERT operating table and immobilization device | 24 | 4 | 4 |
| 5 | Nurse of the operating room | Preparation of IOERT instruments | 12 | 2 | 2 |
| 6 | Anesthesiologist | Quality control of the anesthesiology instruments | 12 | 2 | 2 |
| 16 | 4 | 4 | |||
| 7 | Anesthesiologist | Patient monitoring and camera functionality start-up | 12 | 2 | 2 |
| 16 | 4 | 4 | |||
| 8 | Nurse of the recovery room | Patient recovery | 12 | 3 | 3 |
| 8 | 2 | 2 | |||
| 12 | 3 | 3 | |||
| 9 | Nurse of the operating room | Patient transfer on IOERT operating table | 60 | 30 | 10 |
| 10 | Surgeon | Surgical procedure and margin check | 45 | 10 | 10 |
| 11 | Surgeon–Nurse of the operating room | Delivery of the specimen of quadrantectomy to the Pathmology Department and margin assessment | 60 | 10 | 5 |
| 80 | 10 | 5 | |||
| 12 | Surgeon | Margin assessment communication | 45 | 10 | 10 |
| 13 | Surgeon | Preparation of the breast flap | 60 | 10 | 10 |
| 14 | Radiation Oncologist–Surgeon | Definition of clinical target volume | 64 | 16 | 16 |
| 15 | Radiation Oncologist–Surgeon | Target thickness evaluation | 48 | 24 | 4 |
| 16 | Radiation Oncologist | Applicator and shielding disk selection | 36 | 8 | 8 |
| 17 | Radiation Oncologist | Dose prescription | 45 | 20 | 20 |
| Beam energy selection | 45 | 20 | 20 | ||
| 18 | Physicist | Gafchromic film for | 60 | 32 | 4 |
| 19 | Radiation Oncologist–Surgeon | Applicator placement | 64 | 36 | 24 |
| 20 | Radiation Oncologist–Surgeon | Shielding disk and applicator alignment | 80 | 60 | 30 |
| 21 | Radiation oncologist–Surgeon | Applicator connection with docking mirror and support | 27 | 9 | 9 |
| 22 | Nurse of the Operating room | Operating table and surgical theater protection with sterilized cover | 45 | 15 | 15 |
| 23 | Anesthesiologist | Movement of operating table and anesthesiologist instruments toward the MOBETRON | 30 | 5 | 5 |
| 24 | Radiotherapy (RT) technician | Alignment between applicator and MOBETRON gantry | 45 | 30 | 30 |
| 25 | RT technician | SOFT-DOCKING | 45 | 10 | 5 |
| 45 | 45 | 30 | |||
| 26 | Physicist | Operating room exit | 27 | 12 | 12 |
| 27 | Physicist | Mobile shielding set-up | 12 | 6 | 6 |
| 28 | Physicist | Monitor unit calculation | 45 | 10 | 10 |
| 29 | RT technician | Data entry | 45 | 20 | 20 |
| 30 | RT technician | Physical delivery of radiation dose (start button pressed) | 48 | 24 | 24 |
| 31 | Anesthesiologist | Anesthesiologist monitoring during irradiation | 36 | 24 | 24 |
| 32 | Physicist | Confirmation of dose delivery | 60 | 20 | 20 |
| 33 | Physicist | Shielding removal | 12 | 6 | 6 |
| 34 | RT technician | MOBETRON removal from the operating table | 45 | 30 | 30 |
| 35 | Radiation Oncologist–Surgeon | Removal of all IOERT devices | 8 | 2 | 2 |
| 36 | Surgeon | Check of treatment area | 20 | 5 | 5 |
| 37 | Radiation Oncologist–Surgeon–Physicist–RT technician | Treatment recording, reporting (in patient report) and signing | 36 | 8 | 8 |
| 38 | RT technician | MOBETRON transfer in the pretreatment area and machine switchoff | 36 | 24 | 24 |
| 39 | Medical Physics and Clinical Engineering Departments | Machine maintenance | 24 | 12 | 8 |
Risk Analysis.
| Severity (S) | Occurrence (O) | Detectability (D) |
|---|---|---|
| 1.No damage | 1.Extremely unlikely | 1.Almost always detected |
| 2.Minimal damage | 2.Low probability | 2.Great probability to be detected |
| 3.Moderate damage in the short term | 3.Moderate probability, it occasionally occurs | 3.Moderate probability to be detected |
| 4.Main damage in the long term | 4.Great probability, it repeatedly occurs | 4.Low probability to be detected |
| 5.Permanent damage | 5.Very high probability, almost inevitable | 5.Very low probability to be detected, remote |
FMECA worksheet: high risk class steps.
| Phases | Professional figures | Procedure | Failure mode | Failure effects | Failure causes | Initial risk ranking | Corrective actions 2013–2014 | Revised risk ranking | Corrective actions 2015–2016 | Final revised risk ranking | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| S | O | D | Risk priority number (RPN) | S | O | D | RPN | S | O | D | RPN | ||||||||
| 3 | Physicist | Pre treatment quality control and authorization | Wrong measure, insertion or reading of a delivery parameter | Wrong delivered dose | Violation of protocol limits for quality assurance | 5 | 3 | 4 | 60 | Adherence to instructions in quality assurance protocol | 5 | 2 | 3 | 30 | Adherence to instructions in quality assurance protocol—wider experience | 5 | 1 | 1 | 5 |
| 9 | Nurse | Patient positioning on the operating table | Patient displacement | Patient displacement, slipping and fall off | Non-adherence to guidelines/protocol | 5 | 4 | 3 | 60 | Adherence to Hospital Policies and Procedures | 5 | 2 | 3 | 30 | Adherence to Hospital Policies and Procedures—wider experience | 5 | 1 | 2 | 10 |
| 11 | Surgeon–Nurse | Delivery of surgical specimen to the Pathology and/or Radiology Department | Transcription error | Wrong patient identification | Communication defect | 5 | 4 | 3 | 60 | Procedure adherence | 5 | 2 | 1 | 10 | Procedure adherence-wider experience | 5 | 1 | 1 | 5 |
| Wrong therapeutic decision | Communication defect | 5 | 4 | 4 | 80 | Surgeon confirmation required | 5 | 2 | 1 | 10 | Procedure adherence-wider experience | 5 | 1 | 1 | 5 | ||||
| 13 | Surgeon | Breast flap preparation | Tissue devitalization | Post operatory morbidity-reintervention | Fat necrosis | 5 | 4 | 3 | 60 | Careful visual control of tissue vitality | 5 | 2 | 1 | 10 | Careful visual control of tissue vitality | 5 | 2 | 1 | 10 |
| 14 | Radiation Oncologist–Surgeon | Treatment area definition | Wrong treatment area definition | Wider or smaller treated area | Wrong visual evaluation of the tumor bed to be irradiated | 4 | 4 | 4 | 64 | Preventive evaluation of tumor dimensions (Mammography, Ultrasound, MRI); intraoperative evaluation | 4 | 2 | 2 | 16 | Preventive evaluation of tumor dimensions (Mammography, Ultrasound, MRI); intraoperative evaluation | 4 | 2 | 2 | 16 |
| 18 | Physicist | Preparation of gafchromic film and placement on the shielding disk | (1)Inadequate placement | Wrong measure of the delivered dose | Wrong observance of the “ | 4 | 3 | 5 | 60 | Observance of the “ | 4 | 2 | 4 | 32 | Observance of the “ | 4 | 1 | 1 | 4 |
| 19 | Radiation Oncologist–Surgeon | Applicator placement | Absent or incomplete adherence of the applicator to the tumor bed | Non-homogeneous irradiation | Air gap presence, blood accumulation, very curved tumor bed | 4 | 4 | 4 | 64 | Accurate visual control, correct placement of the patient on the operating table | 4 | 3 | 3 | 36 | Accurate visual control, correct placement of the patient on the operating table, double check | 4 | 3 | 2 | 24 |
| 20 | Radiation Oncologist–Surgeon | Alignment of the shielding disk | Misalignment of the shielding disk | Unintended normal tissues irradiation below the tumor bed | Low accuracy in the alignment | 5 | 4 | 4 | 80 | Selection of a disk much larger than the applicator size | 5 | 3 | 4 | 60 | New shielding set-up and check with the ultrasound probe | 5 | 3 | 2 | 30 |
| 32 | Physicist | Verification of the correctness of the parameters related to the delivered treatment | (1) Transcription error | Wrong evaluation of the delivered treatment | Oversight or non-observance of the procedures | 5 | 3 | 4 | 60 | Check list and double check between Physicist and Radiation Oncologist | 5 | 2 | 2 | 20 | Check list and double check between Physicist and Radiation Oncologist | 5 | 2 | 2 | 20 |
Figure 1Disk placement behind a patient’s breast parenchyma before IOERT.
Figure 2Sketch of the shielding disk, showing the stack of PMMA and copper layers.
Figure 3New shielding set-up with the surgical elastic band.