Automated cell counts on CSF samples: A multicenter performance evaluation of the GloCyte system.

E A Hod1,2, C Brugnara3, M Pilichowska4, L M Sandhaus5, H S Luu6, S K Forest1,2, J C Netterwald1, G M Reynafarje2, A Kratz1,2.   

Abstract

OBJECTIVES: Automated cell counters have replaced manual enumeration of cells in blood and most body fluids. However, due to the unreliability of automated methods at very low cell counts, most laboratories continue to perform labor-intensive manual counts on many or all cerebrospinal fluid (CSF) samples. This multicenter clinical trial investigated if the GloCyte System (Advanced Instruments, Norwood, MA), a recently FDA-approved automated cell counter, which concentrates and enumerates red blood cells (RBCs) and total nucleated cells (TNCs), is sufficiently accurate and precise at very low cell counts to replace all manual CSF counts.
METHODS: The GloCyte System concentrates CSF and stains RBCs with fluorochrome-labeled antibodies and TNCs with nucleic acid dyes. RBCs and TNCs are then counted by digital image analysis. Residual adult and pediatric CSF samples obtained for clinical analysis at five different medical centers were used for the study. Cell counts were performed by the manual hemocytometer method and with the GloCyte System following the same protocol at all sites. The limits of the blank, detection, and quantitation, as well as precision and accuracy of the GloCyte, were determined.
RESULTS: The GloCyte detected as few as 1 TNC/μL and 1 RBC/μL, and reliably counted as low as 3 TNCs/μL and 2 RBCs/μL. The total coefficient of variation was less than 20%. Comparison with cell counts obtained with a hemocytometer showed good correlation (>97%) between the GloCyte and the hemocytometer, including at very low cell counts.
CONCLUSIONS: The GloCyte instrument is a precise, accurate, and stable system to obtain red cell and nucleated cell counts in CSF samples. It allows for the automated enumeration of even very low cell numbers, which is crucial for CSF analysis. These results suggest that GloCyte is an acceptable alternative to the manual method for all CSF samples, including those with normal cell counts.
© 2017 John Wiley & Sons Ltd.

Entities:  

Keywords:  automated cell counters; cerebrospinal fluid testing; laboratory testing; red cell counts; white cell counts

Mesh:

Year:  2017        PMID: 28880455     DOI: 10.1111/ijlh.12728

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  3 in total

Review 1.  Technological features of blast identification in the cerebrospinal fluid: A systematic review of flow cytometry and laboratory haematology methods.

Authors:  John L Frater; Cara Lunn Shirai; Jonathan R Brestoff
Journal:  Int J Lab Hematol       Date:  2022-07-04       Impact factor: 3.450

2.  Automated Analysis of Cerebrospinal Fluid Cells Using Commercially Available Blood Cell Analysis Devices-A Critical Appraisal.

Authors:  Manfred Wick; Catharina C Gross; Hayrettin Tumani; Brigitte Wildemann; Martin Stangel
Journal:  Cells       Date:  2021-05-18       Impact factor: 6.600

3.  On-Chip Cell Staining and Counting Platform for the Rapid Detection of Blood Cells in Cerebrospinal Fluid.

Authors:  Yujin Lee; Byeongyeon Kim; Sungyoung Choi
Journal:  Sensors (Basel)       Date:  2018-04-07       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.