Response to letter re: Carotid atherosclerotic plaques standardized uptake values: methodological issues on reproducibility and accuracy.

Nicola Giannotti1, Martin J O'Connell2,3, Shane J Foley4, Marie C Galligan5, Peter J Kelly3,6, Jonathan P McNulty4.   

Abstract

Entities:  

Year:  2017        PMID: 28879508      PMCID: PMC5587506          DOI: 10.1186/s13550-017-0309-9

Source DB:  PubMed          Journal:  EJNMMI Res        ISSN: 2191-219X            Impact factor:   3.138


× No keyword cloud information.

Background

Dear Editor, We would like to thank Dr. Siamak Sabour for his letter and comments relating to our recently published article [1]. In the manuscript, an investigation was conducted on whether or not carotid atherosclerotic plaque standardized uptake values (SUVs) are consistent and reproducible across software packages; therefore, the purpose of the analysis performed was to measure the reproducibility, rather than validity, of SUV measurements between two software packages (OsiriX MD® version 6.5.2, Pixmeo© SARL, Geneva, Switzerland and AquariusNet iNtuitionTM version 4.4.11, TeraRecon, Foster City, CA, USA) (Table 1).
Table 1

SUV measurements (mean and standard deviation) by location and software together with differences (mean and standard deviation)

Common carotid (CC) and internal carotid (IC) arteriesTeraRecon values (mean ± SD)Osirix values (mean ± SD)Difference in values (mean ± SD)
SUV mean bifurcation left1.5 ± 0.361.58 ± 0.430.08 ± 0.003
SUV mean bifurcation right1.56 ± 0.381.67 ± 0.460.11 ± 0.006
SUV mean CC left1.08 ± 0.591.17 ± 0.60.09 ± 0.004
SUV mean CC right1.11 ± 0.611.23 ± 0.640.12 ± 0.007
SUV mean IC left1.56 ± 0.451.71 ± 0.570.15 ± 0.01
SUV mean IC right1.66 ± 0.381.72 ± 0.440.06 ± 0.002
SUV max bifurcation left2.54 ± 0.652.5 ± 0.70.04 ± 0.008
SUV max bifurcation right2.6 ± 0.692.62 ± 0.730.02 ± 0.002
SUV max CC left2.09 ± 0.912.06 ± 0.880.03 ± 0.001
SUV max CC right2.14 ± 0.972.16 ± 0.990.02 ± 0.001
SUV max IC left2.52 ± 0.672.57 ± 0.840.05 ± 0.001
SUV max IC right2.6 ± 0.692.58 ± 0.740.02 ± 0.001
SUV min bifurcation left0.71 ± 0.30.87 ± 0.360.16 ± 0.012
SUV min bifurcation right0.76 ± 0.30.94 ± 0.380.18 ± 0.016
SUV min CC left0.45 ± 0.360.56 ± 0.460.11 ± 0.006
SUV min CC right0.48 ± 0.390.62 ± 0.480.14 ± 0.01
SUV min IC left0.86 ± 0.341 ± 0.430.14 ± 0.01
SUV min IC right0.84 ± 0.291 ± 0.340.16 ± 0.012
SUV measurements (mean and standard deviation) by location and software together with differences (mean and standard deviation)

Conclusions

We acknowledge that the p values reported in the manuscript previously submitted are dependent on the study sample size, and may not provide sufficient support of measurement reliability. Thus, we will now provide the intra-class coefficient (ICC) for the relevant variables (see Tables 2, 3, and 4) which was found to be supportive of our initial findings.
Table 2

ICC for SUVs mean with 95% confidence intervals

LocationICCConfidence limit minimumConfidence limit maximum
SUV mean bifurcation left0.8430.8060.872
SUV mean bifurcation right0.7870.7330.828
SUV mean CC left0.9060.8870.922
SUV mean CC right0.9240.8580.924
SUV mean IC left0.790.7130.842
SUV mean IC right0.8280.7270.883
Table 3

ICC for SUVs max with 95% confidence intervals

LocationICCConfidence limit minimumConfidence limit maximum
SUV max bifurcation left0.8260.8030.846
SUV max bifurcation right0.8170.7930.838
SUV max CC left0.830.8080.85
SUV max CC right0.8910.8030.891
SUV max IC left0.7520.6990.795
SUV max IC right0.7910.7210.838
Table 4

ICC for SUVs min with 95% confidence intervals

LocationICCConfidence limit minimumConfidence limit maximum
SUV min bifurcation left0.6270.4940.773
SUV min bifurcation right0.6350.4340.748
SUV min CC left0.740.6940.783
SUV min CC right0.650.5450.705
SUV min IC left0.7880.6560.858
SUV min IC right0.4280.270.546
ICC for SUVs mean with 95% confidence intervals ICC for SUVs max with 95% confidence intervals ICC for SUVs min with 95% confidence intervals As expected, higher agreements (ICC) were found among SUV mean and maximum measurements. Effect size measurements also show that SUV max measurements were similar when compared (differences in mean values within the range: 0.02–0.05). Once again we thank Dr. Sabour for his contribution to the important discussion around SUV measurements across software packages.
  1 in total

1.  Carotid atherosclerotic plaques standardised uptake values: software challenges and reproducibility.

Authors:  Nicola Giannotti; Martin J O'Connell; Shane J Foley; Peter J Kelly; Jonathan P McNulty
Journal:  EJNMMI Res       Date:  2017-04-28       Impact factor: 3.138