Abstract
The objective of our study was to compare adhesion of methicillin-resistant Staphylococcus pseudintermedius (MRSP) to stainless steel (SS) and to tantalum (TA) canine limb salvage endoprosthesis implants in an in vitro experimental study. The median of the mean log10 colony forming units/mL for adherent MRSP was 4.96 (range: 4.63 to 6.33) for the TA endoprosthesis and 4.31 (range: 3.86 to 5.05) for the SS endoprosthesis (P = 0.009). Although the trabecular and porous design of the TA endoprosthesis provides mechanical benefits over the SS endoprosthesis, it may increase the risk of developing infection due to higher levels of bacterial adherence.Entities:
Mesh:
Year: 2017 PMID: 28878420 PMCID: PMC5556484