Recording natural head position using an accelerometer and reconstruction from computed tomographic images.

Il Kyung Park1,2, Keun Young Lee3, Yeong Kon Jeong1,2, Rae Hyong Kim2, Dae Gun Kwon2, Sunghee Yeon2, Kyung-Hwan Kwon1.   

Abstract

OBJECTIVES: The concept of natural head position (NHP) was first introduced by Broca in 1862, and was described as a person's stable physiologic position "when a man is standing and his visual axis is horizontal." NHP has been used routinely for clinical examination; however, a patient's head position is random during cone-beam computed tomography (CBCT) acquisition. To solve this problem, we developed an accelerometer to record patients' NHP and reproduce them for CBCT images. In this study, we also tested the accuracy and reproducibility of our accelerometer.
MATERIALS AND METHODS: A total of 15 subjects participated in this study. We invented an accelerometer that measured acceleration on three axes and that could record roll and pitch calculations. Recorded roll and pitch data for each NHP were applied to a reoriented virtual image using three-dimensional (3D) imaging software. The data between the 3D models and the clinical photos were statistically analyzed side by side. Paired t-tests were used to statistically analyze the measurements.
RESULTS: The average difference in the angles between the clinical photograph and the 3D model was 0.04° for roll and 0.29° for pitch. The paired ttests for the roll data (P=0.781) and the pitch data (P=0.169) showed no significant difference between the clinical photographs and the 3D model (P>0.05).
CONCLUSION: By overcoming the limitations of previous NHP-recording techniques, our new method can accurately record patient NHP in a time-efficient manner. Our method can also accurately transfer the NHP to a 3D virtual model.

Entities:  

Keywords:  Cone-beam computed tomography; Surgical diagnostic techniques; Three-dimensional imaging

Year:  2017        PMID: 28875140      PMCID: PMC5583200          DOI: 10.5125/jkaoms.2017.43.4.256

Source DB:  PubMed          Journal:  J Korean Assoc Oral Maxillofac Surg        ISSN: 1225-1585


I. Introduction

The concept of natural head position (NHP) was first introduced by Broca1 in 1862, and was described as a stable physiologic position “when a man is standing and his visual axis is horizontal.” In 1958, NHP was described more specifically as “a standardized and reproducible orientation of the head in space when one is focusing on a distant point at eye level”2. Physiologically, head position is controlled by vestibuloocular and vestibulospinal reflexes, as well as by inner ear otolithic gravitational responses that provide interactions among eye position, head position, and muscles3. According to previous studies, NHP can be influenced by nasorespiratory function456, occlusion789, craniofacial skeletal morphologic features, and visual feedback deprivation10. Based on these findings, NHP might change with emotional state, age, personal characteristics, environmental conditions, and cervical discomfort11. NHP has been used routinely for clinical examination in medicine and dentistry by plastic and maxillofacial surgeons, as well as by orthodontists. Since NHP provides a coordinate reference system for measurements and therefore has vital importance for orthodontic diagnosis, surgical design, postoperative outcome evaluation, and developmental tracking in patients with dentomaxillofacial deformities, it has drawn the interest of many experts121314. Previous studies have mainly focused on how to record NHP with two-dimensional (2D) cephalograms1516171819. Unfortunately, traditional methods for recording NHP are limited to 2D cephalograms and are not appropriate for three-dimensional (3D) models. Clinicians use 3D intracranial reference planes in place of 3D NHP models2021. Now, with the advent of 3D technology, many authors have introduced new methods for recording 3D NHP with cone-beam computed tomography (CBCT) and integrating it into computer-assisted 3D diagnosis and evaluation. However, because patient head position is random during CBCT acquisition, it is difficult to place each person in his or her NHP. To solve this problem, we developed an accelerometer to record patient NHP and reproduce it in CBCT images. We also tested the accuracy and reproducibility of our accelerometer.

II. Materials and Methods

This study was conducted at the Departments of Oral and Maxillofacial Surgery and/or Orthodontics, College of Dentistry, Wonkwang University (Iksan, Korea) from April to October 2016. Fifteen patients (7 males and 8 females; mean age, 25.2 years) participated in this study. Their chief complaint was facial asymmetry. The study was approved by the Institutional Review Board of Wonkwang University (w1410/002-001), and informed consent was obtained prior to patient enrollment. We invented a small accelerometer (3×3×2 cm, 60 g) using the TinyDuino Platform (TinyCircuits, Akron, OH, USA) with a Nordic BLE nRF8001 chip and a Bosch BMA250 chip. The horizontal axis and vertical axis of the device are indicated with 2 wires.(Fig. 1) The Nordic BLE nRF8001 chip enables the device to communicate with the investigator's smartphone through an application (nRF UART v2.0) using Bluetooth technology (ver. 4.0. Bosch BMA250 chip). The accelerometer has the following characteristics—digital resolution: 10 bits; measurement ranges: ±2 g, ±4 g, ±8 g, ±16 g; sensitivity: 2 g: 256 LSB/g, 4 g: 128 LSB/g, 8 g: 64 LSB/g, 16 g: 32 LSB/g, zero-g offset (over lifetime): ±80 mg measures 3-axis acceleration, and roll and pitch are calculated every 0.5 seconds using an inverse trigonometric function. The last five data points in the average were used as the patient NHP (roll and pitch).(Fig. 2)
Fig. 1

The accelerometer and three lingual buttons. We made a small accelerometer (3×3×2 cm, 60 g) using the TinyDuino Platform (TinyCircuits, USA) with a Nordic BLE nRF8001 bluetooth 4.0 chip and Bosch BMA250 3-axis accelerator chip on it. Roll and pitch are calculated per 0.5 second with inverse trigonometric function.

Fig. 2

Measuring roll and pitch with an application. Roll and pitch data of the device were transmitted to android smartphone application (nRF UART v2.0). Averaged last five data were used in this study.

The device was attached to each subject's forehead with skin tape, and three lingual buttons were placed in the preauricular area and on both zygomatic areas.(Fig. 3, 4) Subject NHP (roll and pitch) was recorded using a mirror-guided technique.(Fig. 5) Frontal and lateral clinical photos were taken, and then a CBCT with a standard Frankfort horizontal (F-H) plane-based position was taken.
Fig. 3

Clinical photo (lateral side). Two lingual buttons, one on the periauricular and another on the right zygomatic area are shown. To compare the pitch value, we made a line connecting those buttons, and the angle was calculated.

Fig. 4

Clinical photo (frontal side). Two lingual buttons on the bilateral zygomatic area are shown. To compare the roll value, we made a line between those buttons, and the angle was measured.

Fig. 5

Clinical photo room. The mirror on the wall was used to orient patient's natural head position by mirror-guided technique. The mirror is 150 cm far from the chair. The studio-stand on the bottom of the picture was set to be parallel to the gravity line, vertically, and perpendicular to the gravity line, horizontally by using a fluid level.

Those CT data were reconstructed with OnDemand3D ver. 1.0 (Cybermed Inc., Tustin, CA, USA), and reorientation was performed using the device's horizontal and vertical axes. (Fig. 6) Recorded NHP roll and pitch data were applied to the reoriented virtual image. On the virtually recreated NHP 3D model, we made a line between two lingual buttons present on each the frontal and lateral side. The angles between those lines and the horizontal line were calculated.(Fig. 7, 8) The same procedure was conducted on the clinical photos. These angles between the 3D models and the clinical photos were statistically analyzed side by side.
Fig. 6

Three-dimensional (3D) model reorientation. The cone-beam computed tomography data were reconstructed to 3D model by virtual imaging software (OnDemand3D ver. 1.0; Cybermed Inc., USA), and the 3D model was reoriented as the axis' of the device and the computed tomography images' were parallel to each other.

Fig. 7

Recreated natural head position (NHP) on three-dimensional (3D) simulator, frontal side. The 3D model was re-oriented to the NHP based on the roll and pitch data acquired by the device. Then the angle between two lingual buttons on both zygoma in the frontal side was calculated to compare the roll value.

Fig. 8

Recreated natural head position (NHP) on three-dimensional (3D) simulator. The 3D model was re-oriented to the NHP based on the roll and pitch data acquired by the device. Then the angle between two lingual buttons on periauricula and zygomatic area in the lateral side was calculated to compare the pitch value.

The Kolmogorov-Smirnov test and the Shapiro-Wilk test were used to test for normality. The comparisons of the angles between the 3D models and the clinical photos were conducted with paired t-tests. IBM SPSS ver. 21.0 for Windows (IBM Co., Armonk, NY, USA) was used for all analysis. A two-tailed P-value <0.05 was considered statistically significant for all analyses.

III. Results

The average difference between the angle in the clinical photograph and the angle in the 3D model was 0.04° for the roll and 0.29° for the pitch. Both roll and pitch data were normally distributed, according to the Kolmogorov-Smirnov test and the Shapiro-Wilk test.(Table 1)
Table 1

Tests of normality

Kolmogorov-Smirnov1Shapiro-Wilk
StatisticdfSig.StatisticdfSig.
Roll0.175150.20020.911150.140
Pitch0.129150.20020.942150.407

(df: degree of freedom, Sig.: significance)

1Lilliefors significance correction. 2This is a lower bound of the true significance.

The results from the paired t-tests for the roll data (P=0.781) and the pitch data (P=0.169) indicated that there was no significant difference between the clinical photographs and the 3D model (P>0.05).(Table 2)
Table 2

Paired-samples test

Paired differencestdfSig. (2-tailed)
MeanStd. deviationStd. error mean95% CI of the difference
LowerUpper
Roll−0.04000.54620.1410−0.34250.2625−0.284140.781
Pitch0.29330.78330.2022−0.14040.72711.450140.169

(Std.: standard, CI: confidence interval, df: degree of freedom, Sig.: significance)

IV. Discussion

In this study, we used the mirror position technique to acquire patient NHP. Three methods have been widely used to acquire NHP; the self-balanced position, the mirror position, and the estimated position. The self-balanced position asks the patient to look into horizontal infinity12, and the mirror position asks the patient to look into his or her own eyes in a mirror322. The estimated position is generally thought to be that wherein the head position can be reproduced with less variation when corrections are made by the doctor2. A study that investigated the reproducibility of the roll and pitch of NHP found no significant differences in those three methods in reproducibility for either measure23. Recording patient NHP and reproducing it in a virtual environment is becoming increasingly important for pretreatment planning and posttreatment evaluation. One of the most common methods for orienting a 3D CT model in NHP is to use the clinician's insight or digitized landmarks to the NHP24. This method is more accurate when the craniofacial structures are symmetrical than when the patient has severe facial asymmetries. Recently, several newly developed methods for recording NHP in three dimensions have been introduced, including a digital gyroscope that can record NHP in three dimensions1225. For this method, an electronic device is mounted on a bite jig, and the patient holds the device by biting the bite jig until the 3D photographs are taken with the patient in a NHP. The NHP orientation data are automatically recorded and transferred to the virtual imaging software for 3D CT model reconstruction. However, there is potential for error because the patient is holding a heavy instrument for a long period of time12. The 3dMDface imaging system was used to record sitting NHP32627. For this method, four ink dots were marked on a patient's face to record NHP before taking the 3D photograph. However, because the patient's head is prone to moving while dots are marked on the face, this technique is not reliable. To overcome those difficulties, we developed a new method to record NHP and to transfer the NHP to the 3D CT model. NHP recording methods should not influence head orientation. We designed an accelerometer that is not only smaller (3×3×2 cm), but also lighter (60 g) than any other previous devices. We placed the device on the patient's forehead so that its center of gravity aligned with that of the patient's head. Also, as long as the device was placed on the patient's head, its specific position was not a concern. Because we wanted to assess the device's accuracy in reproducing each patient's individual NHP in 3D images, the device was attached to each patient's forehead and did not move until the clinical photo and CBCT were taken. Because the device can calculate real-time roll and pitch angles (per 0.5 seconds) and record those data automatically, there is no time interval between acquiring and recording NHP. In this study, a clinician attached the device to a patient before taking clinical photographs, so this procedure itself does not affect NHP acquisition or recording. Since it does not require additional devices or procedures, we could record patient NHP at the first visit within a few seconds while clinical photographs were taken. Furthermore, it is possible to reconstruct individual NHP on any diagnostic software that supports coordinate-dependent modeling. Reconstructed NHP on a 3D simulator showed no significant differences (P>0.005) in NHP; all differences (average differences were 0.04° in roll and 0.29° in pitch) were within the established NHP variability of 2°228. However, the small number of patients (n=15) was a limitation of this study. Also, we did not identify NHP yawing, which is a very important factor for diagnosing facial asymmetry, as analyzing data with only pitch and roll can lead to errors in the estimated position in 3D space. In future studies, more patients and yaw analysis should be included. All of our procedures were conducted by one examiner.

V. Conclusion

While overcoming the limitations of previous NHP recording techniques, this device offers a new way to record NHP with accuracy and efficiency. Also, it can accurately transfer NHP to a 3D virtual model of the human head.
  26 in total

1.  Reproducibility of natural head position measured with an inclinometer.

Authors:  Serdar Uşümez; Metin Orhan
Journal:  Am J Orthod Dentofacial Orthop       Date:  2003-04       Impact factor: 2.650

2.  Facial planning for orthodontists and oral surgeons.

Authors:  G William Arnett; Michael J Gunson
Journal:  Am J Orthod Dentofacial Orthop       Date:  2004-09       Impact factor: 2.650

3.  [Longitudinal study of the stability and reproducibility of natural head position in adolescents with different facial types over time].

Authors:  Xin-ping Lin; Stenvik Arild
Journal:  Shanghai Kou Qiang Yi Xue       Date:  2005-06

4.  A cephalometric analysis based on natural head position.

Authors:  A D Viazis
Journal:  J Clin Orthod       Date:  1991-03

5.  Natural head position and growth of the facial part of the skull.

Authors:  Anna M Dubojska; Grazyna Smiech-Slomkowska
Journal:  Cranio       Date:  2013-04       Impact factor: 2.020

6.  Simple technique to achieve a natural head position for cephalography.

Authors:  Chun-Ming Chen; Steven Lai; Y C Tseng; Kun-Tsung Lee
Journal:  Br J Oral Maxillofac Surg       Date:  2008-08-30       Impact factor: 1.651

7.  Simple technique to achieve a natural position of the head for cone beam computed tomography.

Authors:  Janalt Damstra; Zacharias Fourie; Yijin Ren
Journal:  Br J Oral Maxillofac Surg       Date:  2009-11-01       Impact factor: 1.651

8.  A simple method for taking natural-head-position cephalograms.

Authors:  K J Showfety; P S Vig; S Matteson
Journal:  Am J Orthod       Date:  1983-06

9.  Relationship of natural head position to craniofacial morphology.

Authors:  P Leitão; R S Nanda
Journal:  Am J Orthod Dentofacial Orthop       Date:  2000-04       Impact factor: 2.650

10.  Head orientation in CBCT-generated cephalograms.

Authors:  Lucia Cevidanes; Ana E F Oliveira; Alexandre Motta; Ceib Phillips; Brandon Burke; Donald Tyndall
Journal:  Angle Orthod       Date:  2009-09       Impact factor: 2.079

View more
  2 in total

1.  Cephalometric Evaluation of Natural Head Position in Lingayat Population of Karnataka.

Authors:  Inderpreet Singh; K Kiran Kumar; Pradeep Raj; R Haranadh Babu; Narendra Pithani; Altaf H Thekiya
Journal:  J Pharm Bioallied Sci       Date:  2019-02

2.  Recording Natural Head Position Using Cone Beam Computerized Tomography.

Authors:  Tai-Chiu Hsung; Wai-Kan Yeung; Wing-Shan Choi; Wai-Kuen Luk; Yi-Yung Cheng; Yu-Hang Lam
Journal:  Sensors (Basel)       Date:  2021-12-08       Impact factor: 3.576

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.