Effects of altering heel wedge properties on gait with the Intrepid Dynamic Exoskeletal Orthosis.

Andrea J Ikeda1,2, John R Fergason1, Jason M Wilken1,2.   

Abstract

BACKGROUND: The Intrepid Dynamic Exoskeletal Orthosis is a custom-made dynamic response carbon fiber device. A heel wedge, which sits in the shoe, is an integral part of the orthosis-heel wedge-shoe system. Because the device restricts ankle movement, the system must compensate to simulate plantarflexion and allow smooth forward progression during gait.
OBJECTIVES: To determine the influence of wedge height and durometer on the walking gait of individuals using the Intrepid Dynamic Exoskeletal Orthosis. STUDY
DESIGN: Repeated measures.
METHODS: Twelve individuals walked over level ground with their Intrepid Dynamic Exoskeletal Orthosis and six different heel wedges of soft or firm durometer and 1, 2, or 3 cm height. Center of pressure velocity, joint moments, and roll-over shape were calculated for each wedge.
RESULTS: Height and durometer significantly affected time to peak center of pressure velocity, time to peak internal dorsiflexion and knee extension moments, time to ankle moment zero crossing, and roll-over shape center of curvature anterior-posterior position. Wedge height had a significant influence on peak center of pressure velocity, peak dorsiflexion moment, time to peak knee extension moment, and roll-over shape radius and vertical center of curvature.
CONCLUSION: Changes in wedge height and durometer systematically affected foot loading. Participants preferred wedges which produced ankle moment zero crossing timing, peak internal knee extension moment timing, and roll-over shape center of curvature anterior-posterior position close to that of able-bodied individuals. Clinical relevance Adjusting the heel wedge is a simple, straightforward way to adjust the orthosis-heel wedge-shoe system. Changing wedge height and durometer significantly alters loading of the foot and has great potential to improve an individual's gait.

Entities:  

Keywords:  Biomechanics of prosthetic/orthotic devices; IDEO; biomechanics; lower limb orthotics; orthotics; rehabilitation; rehabilitation of orthoses users

Mesh:

Year:  2017        PMID: 28870146     DOI: 10.1177/0309364617728116

Source DB:  PubMed          Journal:  Prosthet Orthot Int        ISSN: 0309-3646            Impact factor:   1.895