Mitsumasa Hata1, M Shiono1, Y Orime1, K Nakata1, A Sezai1, H Yamada1, T Saito1, Y Sezai1.
Abstract
To estimate coronary microcirculation during left heart bypass (LHB), we performed an experimental comparison study of LHB and intraaortic balloon pumping (1ABP). LHB was performed with a BioMedicus BP-80 pump supporting half of the flow of cardiac output whereas the IABP was pumped in a 1:1 mode for cardiogenic shock in a swine model. Coronary circulations were analyzed by electromagnetic flowmeter, pulsed Doppler velocimeter, and laser Doppler flowmeter. Left ventricular end-diastolic pressure (LVEDP) was reduced significantly by LHB. Although there was no significant difference in epicardial flow between the LHB and IABP groups, endocardial flow was increased significantly by LHB. In the LHB group, the systolic reverse wave of the coronary velocity called a myocardial invalid circulation was reduced remarkably. There was a significant inverse correlation between endocardial flow and LVEDP. These results suggested that LHB was more effective for myocardial microcirculation than was IABP. © 1996 International Society for Artificial Organs.Entities:
Keywords: Coronary microcirculation-; Coronary velocity; Intraaortic balloon pump-; Left heart bypass-
Year: 1996 PMID: 28868713 DOI: 10.1111/j.1525-1594.1996.tb04501.x