Literature DB >> 28867520

Comparison of the iAssist Handheld Guidance System to Conventional Instruments for Mechanical Axis Restoration in Total Knee Arthroplasty.

Matthew C Kinney1, Krishna R Cidambi2, Dustyn L Severns3, Francis B Gonzales1.   

Abstract

BACKGROUND: Recent advances in total knee arthroplasty (TKA) include an intelligent instrument system designed to provide intraoperative guidance to reduce mechanical alignment errors. Internal position-sensing technology is integrated into microelectronic pods that attach to cutting blocks. The purpose of this prospective, randomized study was to determine whether this iAssist system enables the surgeon to make more accurate bone resections and better restore the mechanical axis compared to conventional instruments in TKA.
METHODS: We randomized patients undergoing TKA into 2 groups. Group I (n = 25) underwent TKA assisted by the iAssist guidance system, group II (n = 25) underwent TKA using conventional instruments. Preoperative and postoperative mechanical axes were measured from full-length lower extremity radiographs to evaluate alignment. Additional surgical parameters were also assessed, including tourniquet time and blood loss.
RESULTS: Patient demographics and preoperative mechanical axis alignments were similar between the groups. Postoperatively, 4.0% of patients had greater than 3° of tibial or femoral component mal-alignment in the guidance-assisted cohort, compared with 36.0% in the conventional group (P < .05). Additionally, group I showed significant improvement in variance seen in both the femoral mechanical axis (1.65° ± 0.17° vs 2.23° ± 0.33°, P < .005) and tibial mechanical axis (1.28° ± 0.13° vs 1.71° ± 0.24°, P < .005) compared to group II. There were no significant differences in tourniquet time (P = .86) or blood loss (P = .39) between groups.
CONCLUSION: Use of the iAssist system in TKA results in an improved postoperative mechanical axis and decreased alignment variability compared to conventional instruments, without significantly increasing operative time.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  computer navigation; handheld navigation; mechanical axis alignment; primary total knee arthroplasty; surgical technique

Mesh:

Year:  2017        PMID: 28867520     DOI: 10.1016/j.arth.2017.06.004

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  9 in total

1.  Comparison of a novel handheld accelerometer-based navigation system and conventional instrument for performing distal femoral resection in total knee arthroplasty: a randomized controlled trial.

Authors:  Xingquan Xu; Peilai Liu; Zhenfeng Yuan; Dawei Wang; Qunshan Lu; Zhe Zhang; Qing Jiang; Dongquan Shi
Journal:  Ann Transl Med       Date:  2019-11

2.  Current role of handheld navigation system in total knee arthroplasty: where we are?

Authors:  Jong-Heon Kim; Woo-Sung Kim; Hyun-Soo Ok; Seong Hwan Kim
Journal:  Ann Transl Med       Date:  2020-03

Review 3.  Clinical outcomes associated with robotic and computer-navigated total knee arthroplasty: a machine learning-augmented systematic review.

Authors:  Quinlan D Buchlak; Joe Clair; Nazanin Esmaili; Arshad Barmare; Siva Chandrasekaran
Journal:  Eur J Orthop Surg Traumatol       Date:  2021-06-25

4.  Total knee arthroplasty in the next decade: is navigation necessary?

Authors:  Graham S Goh; Ming Han Lincoln Liow
Journal:  Ann Transl Med       Date:  2020-06

5.  Accelerometer-Based Navigation in Total Knee Arthroplasty for the Management of Extra-Articular Deformity and Retained Femoral Hardware: Analysis of Component Alignment.

Authors:  Andrea Cozzi Lepri; Matteo Innocenti; Fabrizio Matassi; Marco Villano; Roberto Civinini; Massimo Innocenti
Journal:  Joints       Date:  2019-10-11

6.  Trends in Operative Time and Short-Term Outcomes After Conventional and Navigated Total Knee Arthroplasty.

Authors:  Troy K Sekimura; Alexander Upfill-Brown; Peter P Hsiue; Amir Khoshbin; Erik N Zeegen; Alexandra I Stavrakis
Journal:  Arthroplast Today       Date:  2021-04-12

Review 7.  Accelerometer-based navigation vs. conventional techniques for total knee arthroplasty (TKA): a systematic review and meta-analysis of randomized controlled trials.

Authors:  Juntan Li; Yuqi Zhang; Xiang Gao; Tianxu Dou; Xu Li
Journal:  Arthroplasty       Date:  2022-09-02

8.  Are There Differences in Accuracy or Outcomes Scores Among Navigated, Robotic, Patient-specific Instruments or Standard Cutting Guides in TKA? A Network Meta-analysis.

Authors:  Pierre-Alban Bouché; Simon Corsia; Agnès Dechartres; Matthieu Resche-Rigon; Rémy Nizard
Journal:  Clin Orthop Relat Res       Date:  2020-09       Impact factor: 4.755

9.  Comparison of iASSIST Navigation System with Conventional Techniques in Total Knee Arthroplasty: A Systematic Review and Meta-Analysis of Radiographic and Clinical Outcomes.

Authors:  Jun-Tan Li; Xiang Gao; Xu Li
Journal:  Orthop Surg       Date:  2019-11-22       Impact factor: 2.071

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.