The Effect of Pulsing on Transverse Ultrasound Efficiency and Chatter.

Alex J Wright1, Austin D Bohner1, Ashlie A Bernhisel1, Brian Zaugg1, William R Barlow1, Jeff H Pettey1, Randall J Olson2.   

Abstract

PURPOSE: To evaluate the effects of micropulse, long pulse, and continuous ultrasound on transverse ultrasound using Abbott Medical Optics' (AMO) WhiteStar Signature Pro with the Ellips FX handpiece.
DESIGN: In vitro laboratory study.
METHODS: This study was conducted at the John A. Moran Eye Center Laboratory, University of Utah, Salt Lake City, Utah, USA. Porcine lenses were hardened in formalin for 2 hours and equilibrated in basic salt solution (BSS) over a 24-hour period. The lenses were then cubed in 2.0 × 2.0-mm pieces. These pieces were stored in BSS until the time of experimentation. The AMO WhiteStar Signature Pro machine (Abbott Medical Optics) with the Ellips FX handpiece and a 0.9-mm bent Dewey tip with a 30-degree bevel (Microsurgical Technology Inc) were used for phacoemulsification. Three runs of 20 lenses each were performed, measuring efficiency and chatter. Transverse ultrasound varied in the 3 runs and included continuous, 6 ms on/off micropulse, and 50 ms on/off long pulse.
RESULTS: Micropulse was more efficient than long pulse by 43% (P = .00003) and continuous by 42% (P = .000387). There were also less chatter events with micropulse than with long-pulse and continuous ultrasound. However, this difference did not reach significance.
CONCLUSION: The 6 ms on and 6 ms off micropulse transverse 3-dimensional ultrasound is more efficient and produces fewer chatter events than both long-pulse and continuous ultrasound.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28860045     DOI: 10.1016/j.ajo.2017.08.012

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  1 in total

1.  Optimization of Phacoemulsification Tip Gauge on the Oertli CataRhex3 in an in vitro Setting.

Authors:  John S Peterson; Michael R Christensen; Caitlynn Cooper; Christina Mamalis; Ashlie A Bernhisel; Brian Zaugg; William R Barlow; Emilie Ungricht; Randall J Olson; Jeff H Pettey
Journal:  Clin Ophthalmol       Date:  2022-04-08
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.