Boqun Liu1, Bert Poolman1, Arnold J Boersma1. 1. Department of Biochemistry, Groningen Biomolecular Sciences and Biotechnology Institute & Zernike Institute for Advanced Materials, University of Groningen , Nijenborgh 4, 9747 AG Groningen, The Netherlands.
Abstract
Knowledge of the ionic strength in cells is required to understand the in vivo biochemistry of the charged biomacromolecules. Here, we present the first sensors to determine the ionic strength in living cells, by designing protein probes based on Förster resonance energy transfer (FRET). These probes allow observation of spatiotemporal changes in the ionic strength on the single-cell level.
Knowledge of the ionic strength in cells is required to understand the in vivo biochemistry of the charged biomacromolecules. Here, we present the first sensors to determine the ionic strength in living cells, by designing protein probes based on Förster resonance energy transfer (FRET). These probes allow observation of spatiotemporal changes in the ionic strength on the single-cell level.
The ionic
strength influences
a wide array of electrostatic interactions in the cell.[1,2] To understand the role of the ionic strength in cell physiology,
a sensor to quantify the ionic strength in vivo is
needed. The ionic strength is the effective ion concentration and
screens electrostatic interactions of (macro)molecules, which is obviously
of crucial importance to the cell. For example, the ionic strength
determines the structure of intrinsic disordered proteins,[3] activity of enzymes,[4] protein aggregation,[5] quinary assemblies/phase
separations,[6] protein binding to (poly)nucleic
acids,[7] the catalytic function of riboswitches,[8] and many other processes. The ionic strength
governs the cell volume by activating the channels and transporters
involved.[9,10] The ionic strength can vary, however, depending
on the extracellular environment, due to, for example, abrupt changes
in medium osmolarity and fluctuations with intracellular events such
as metabolite or polynucleotide synthesis.[11]To quantify the ionic strength, one cannot simply use the
concentration
of ions, because the effective ion concentration (i.e., thermodynamic
activity) is altered due to interaction of the ions with each other
and the biomolecules. Indeed, a subset of ions has a higher affinity
for macromolecules, such as magnesium, while others are more loosely
associated. Their activity coefficient may approach that in the extracellular
medium.[12] Such effects result in a strong
dependence of electrostatic screening on the identity of the ions,
roughly following the Hofmeister series.[13,14] Thus, although it is possible to determine the concentration of
a particular ion with, for example, flame photometry of dry cell mass,[15] this will not immediately reveal the ionic strength;
that is, the effective concentration of an ion depends on its environment.
In addition, these techniques will not yield information on population
heterogeneity or rapid temporal changes in the ionic strength. Probes
to infer the ionic strength from the activity of a membrane protein[9] or the fluorescence of a fluorescein–BSA
conjugate[16] are complicated to use and
interpret, and they have only been applied in vesicles and isolated
mitochondria but not in living cells.Here, we fill this void
by designing the first probes that sense
the ionic strength in living cells. The sensors are based on FRET
and are genetically encoded, because these properties allow monitoring
analytes with high spatiotemporal precision inside living cells. Fluorescent
protein probes can be affected by ions nonspecifically,[17] and FRET sensors selective for a specific ion
have been developed, for example, zinc, calcium, or choride.[18−20] We construct our sensors to contain a positively and a negatively
charged α-helix, whose electrostatic attraction will depend
on the ionic strength (Figure ). At the N- and C-termini of the helices are two fluorescent
proteins, mCerulean3 and mCitrine, that form a FRET pair.[21] Increasing the ionic strength will decrease
the attraction between the helices, which will result in a decrease
in FRET efficiency.
Figure 1
Design of ionic strength sensors and sensing concept.
Ions screen
the attraction between the positively and negatively charged helices,
reducing the FRET efficiency.
Design of ionic strength sensors and sensing concept.
Ions screen
the attraction between the positively and negatively charged helices,
reducing the FRET efficiency.To induce charge in the helices, we inserted six glutamates
in
an alanine background in one helix and six lysines in the other helix
(Figure ). The charged
amino acids are in an i + 5 spacing, a staircase-like
configuration, ensuring that all sides of the helix are covered with
charges. This prevents the helix from having a charged patch as well
as preventing metal ion chelation. We further reasoned that the high
intramolecular concentration of the oppositely charged α-helix
will outcompete electrostatic interactions with most of the charged
cellular biomacromolecules. Because the presence of specific ion effects
could be dependent on the identity of the amino acids in the helix,
we designed sensors with glutamate or aspartate residues, and lysine
or arginine residues, and thereby alter, e.g., salt-bridge stability.[22] With these design elements, we avoid selectivity
for specific ions in the sensors, allowing for ionic strength determination
in living cells.To determine the ionic strength sensitivity
of the sensors, we
performed extensive in vitro tests on isolated probes
(Figure A, Supporting Information Figure S1). We tested
the three ionic strength sensors, which are the KE (lysine-glutamate),
the RE (arginine-glutamate), and the RD (arginine-aspartate) sensors,
and compared the findings with a sensor that contains neutral helices,
the E6G2 probe.[23] We found that the FRET
efficiencies of the ionic strength sensors in 10 mM NaPi buffer were
much higher than that of the E6G2 probe, confirming that the helices
attract each other. The addition of potassium chloride to the ionic
strength sensors led to a decrease in FRET efficiency (Figure B), with the highest sensitivity
between 0 and 300 mM KCl. The E6G2 probe was not sensitive, showing
that we probe the ionic strength. Salt does not affect other neutral
sensors with higher FRET efficiencies,[23] and hence the FRET efficiency does not relate to the ionic strength
dependence. We added a wide range of salts to test whether specific
ion effects interfere with the readouts (Figure C). We find that the sensitivity to the cations
of KCl, NaCl, MgCl2, and LiCl is the same when the ratios
are plotted versus the ionic strength (Supporting Information Figure S3). In contrast, the readout had a dependence
on the identity of the anion (Supporting Information Figure S2). The order of sensitivity followed the Hofmeister
series, that is, the measure to what extent ions are hydrated, which
is generally more dependent on the anion. This deviation from ideal
behavior depended on the probe and qualitatively followed the intramolecular
salt bridge strength between the two helices;[22] the RD probe has the lowest salt-bridge strength and is least affected
by the nonideal behavior, as judged from the spread in deviation between
the different ions. Likely due to the lower salt-bridge strength,
the RD probe also has a lower FRET efficiency. Together, the dependence
on ion identity indicates that we probe the effective ion concentration
(or ionic strength) rather than total ion concentration. In all cases,
the E6G2 probe was insensitive to any ion.
Figure 2
In vitro characterization of the ionic strength
sensors. (a) Fluorescence emission spectra of the RD sensor titrated
with potassium chloride. (b) Dependence of the mCitrine/mCerulean3
ratio of the EK, RD, RK, and E6G2 sensors on the potassium chloride
concentration. (c) Dependence of the mCitrine/mCerulean3 ratio of
the RD sensor on the ion identity. All experiments are in 10 mM NaPi,
at pH 7.4. Error bars indicate the standard deviation of three biological
replicates.
In vitro characterization of the ionic strength
sensors. (a) Fluorescence emission spectra of the RD sensor titrated
with potassium chloride. (b) Dependence of the mCitrine/mCerulean3
ratio of the EK, RD, RK, and E6G2 sensors on the potassium chloride
concentration. (c) Dependence of the mCitrine/mCerulean3 ratio of
the RD sensor on the ion identity. All experiments are in 10 mM NaPi,
at pH 7.4. Error bars indicate the standard deviation of three biological
replicates.We investigated sensitivity
to a wide variety of other parameters
to make the transition to in vivo measurements. Similar
to the E6G2 and GE crowding probes,[21] mCitrine
induced pH sensitivity in all the probes only below pH 7.0 (Supporting Information Figure S4). The zwitterionic
glycine betaine and neutral small molecules sorbitol and sucrose hardly
influenced the probes (Supporting Information Figure S5). The common intracellular organophosphates fructose
bisphoshate, ATP, ADP, and AMP influenced the probes as can be expected
from their charge, albeit the sensitivity to the nucleotides was somewhat
higher (Supporting Information Figure S6). Glutathione influenced the probes according to its net negative
charge (Supporting Information Figure S7). The sensitivity to potassium glutamate was less than that to potassiumchloride, as expected from the Hofmeister series, with the RD probe
again deviating the least (Supporting Information Figure S8). The sensors associated with the polyelectrolyte
DNA in the absence of KCl, but not in the presence of 100 mM KCl (Supporting Information Figure S9). Macromolecular
crowding induced by Ficoll 70 compressed all the probes (Supporting Information Figure S10) and hence
should be taken into account when measuring in live cells. The ionic
strength probes are temperature sensitive in the absence of salt,
but the temperature sensitivity decreased with increasing salt concentration
(Supporting Information Figure S11). Increasing
the sensor concentration itself did not affect the ratio (Supporting Information Figure S12), ruling out
interference from self-association. Hence, the sensors are sensitive
to the ionic strength mainly, but other parameters such as macromolecular
crowding, pH, and temperature need to be controlled for.To
demonstrate ionic strength sensing in vivo,
we expressed the probes in the mammalian cell line HEK293 and imaged
the cells by scanning confocal microscopy. The cells were subsequently
imaged by excitation of the mCerulean3 at 405 nm, and the emission
ratio of mCitrine (505–750 nm) over mCerulean3 (450–505
nm), after background subtraction, was monitored (Figure A,B). The ratios obtained were
very homogeneous over all the cells. When comparing the different
sensors, we found that RD gave lower ratios than the KE and RE probes,
similar to what we observed in vitro (Supporting Information Figure S14). To quantify
the readouts, we calibrated the probes in the cell by clamping the
internal potassium concentration by titrating the external medium
with potassium chloride in the presence of the ionophores valinomycin
plus nigericin. These ionophores equilibrate the protons and potassium
over the membrane, providing a calibration curve of the corresponding
sensor with a known concentration of potassium. All the ion sensors
sensed the potassium concentration in a similar manner to that observed in vitro. The ratios for the GE crowding sensor remained
unchanged, indicating that with the calibration method we did not
alter the crowding and prevent osmotic pressure differences over the
membrane by ionophore-assisted ion equilibration. The cells did change
shape during the calibration procedure (Figure A). We subsequently used the calibration
curves to quantify the ionic strength from the readout in a regular
growth medium and found that the ionic strength was comparable to
∼130 mM for the RD and RE sensors and ∼110 mM for the
KE sensor (Figure B, Supporting Information Figure S14).
These values are in the same range as the ion concentrations reported
for the monovalent ions in various cells, and the concentration of
“free” ions is somewhat lower.[24] The total ion concentration would be 140 mM and 12 mM for potassium
and sodium, respectively. Divalent cations were not monitored because
they are mostly bound to the biopolymers. Organophosphates form complexes
with magnesium and may bind nonspecifically to proteins,[25−27] and the molecules thus contribute less to the ionic strength than
is expected from their total concentrations. Single cell measurements
as a function of time indicate that the ionic strength is determined
with a precision better than 10 mM (that is, changes in mCitrine/mCerulean3
of less than 0.02). Thus, we conclude that the probes function in
HEK293 cells, and we can determine the ionic strength in vivo.
Figure 3
Ionic strength determination in vivo and observation
of dynamic changes observed in scanning fluorescence confocal microscopy
of HEK293 cells expressing the RD sensor. (a) Fluorescence and mCitrine/mCerulean3
ratio images of sensor calibration with KCl, nigericin, and valinomycin.
(b) Calibration curve of the mCitrine/mCerulean3 ratio versus potassium
concentration; the red line is the mCitrine/mCerulean3 ratio of cells
in growth medium (red shading is the standard deviation within a population, n = 20). (c) Fluorescence and mCitrine/mCerulean3 ratio
images of cells before and 10 min after addition of 450 mM sorbitol.
See Supporting Figure S13 for more detail.
(d) mCitrine/mCerulean3 ratio upon addition of sorbitol as a function
of time. The data of 20 individual cells and the average are shown.
Error bars are standard deviation within a single population of cells.
The standard deviations of the averages of independent biological
repeats are smaller (±0.03, n = 4).
Ionic strength determination in vivo and observation
of dynamic changes observed in scanning fluorescence confocal microscopy
of HEK293 cells expressing the RD sensor. (a) Fluorescence and mCitrine/mCerulean3
ratio images of sensor calibration with KCl, nigericin, and valinomycin.
(b) Calibration curve of the mCitrine/mCerulean3 ratio versus potassium
concentration; the red line is the mCitrine/mCerulean3 ratio of cells
in growth medium (red shading is the standard deviation within a population, n = 20). (c) Fluorescence and mCitrine/mCerulean3 ratio
images of cells before and 10 min after addition of 450 mM sorbitol.
See Supporting Figure S13 for more detail.
(d) mCitrine/mCerulean3 ratio upon addition of sorbitol as a function
of time. The data of 20 individual cells and the average are shown.
Error bars are standard deviation within a single population of cells.
The standard deviations of the averages of independent biological
repeats are smaller (±0.03, n = 4).To further demonstrate the potential of the sensors in
vivo, we monitored changes in ionic strength over time after
an osmotic upshift by adding 450 mM sorbitol to HEK293 cells (Figure C,D). We expected
the short-term response to be an initial increase in crowding and
ionic strength, followed by a regulatory volume increase by uptake
of potassium and chloride from the medium, retaining the abnormally
high ionic strength but decreasing the crowding.[11] We assume that the response mechanism of the sensor is
much faster than the biological events. We saw with the GE crowding
probe that the crowding indeed increased with 450 mM sorbitol (Supporting Information Figure S15), as we observed
previously, after which a slow decrease in crowding took place. In
the time frame of ∼20 min, the crowding was not yet fully recovered.
The RD sensor showed that the ionic strength increased with an osmotic
upshift and remained at this level. It showed no apparent sensitivity
to in vivo crowding, contrary to the KE and RE sensors
that showed an initial increase in FRET directly after the osmotic
upshift. Interestingly, an increase in ionic strength of ∼190
mM to 320 mM can be inferred with the RD calibration curve, which
is similar to the expected increase in osmolarity simply based on
equating the sum of K and Cl to the number of sorbitol molecules added.
We could monitor single cells in time and found that the distribution
of FRET values increased after the osmotic upshift and the adaptation
process that followed. This shows that the cells are affected differently
by the osmotic upshift, which could, for example, be related to the
cell cycle[15] or intrinsic variation in
fitness in the population of cells that we analyzed. Hence, the probes
can sense the ionic strength on the single-cell level during dynamic
changes in the ionic strength.In conclusion, we present here
the first sensors to determine the
ionic strength that function in living cells. Taking into account
changes in macromolecular crowding and pH, the sensors allow facile
determination of the intracellular ionic strength on the single-cell
level in changing environments or intracellular conditions. Given
the wide variety of processes that are influenced by the ionic strength,
these sensors will aid future investigations on the importance of
the ionic strength during a wide variety of conditions such as cell
volume regulation, disease, and environmental stresses.
Methods
Expression and Purification of the Ionic
Strength Sensors
Synthetic genes (Supporting Information) encoding the ionic strength sensors in pRSET-A
were obtained from
GeneArt. The plasmids were transformed into the E. coli strain BL21(DE3) pLysS (Promega). The cells were grown to an OD600
of 0.6 at 37 °C and shaking at 200 rpm in LB medium (1.0% bactotryptone,
0.5% yeast extract, 1% NaCl) with 1 mg mL–1 ampicillin,
after which the cells were induced overnight with 0.1 mM isopropyl-β-d-thiogalactoside (IPTG) at 25 °C and shaking at 200 rpm.
The cell lysate was cleared by centrifugation, supplemented with 10
mM imidazole, and purified by nickel-nitrilotriacetic acid Sepharose
chromatography (wash/elution buffer: 20/250 mM imidazole, 50 mM NaPi,
300 mM NaCl, at pH 8.0). The sensor was further purified by Superdex
200 10/300 GL size-exclusion chromatography (Amersham Biosciences)
in 10 mM NaPi, at pH 7.4. Fractions containing pure protein were aliquoted
and stored at −80 °C.
In Vitro Characterization of the Sensors
A 300 μL solution
containing the given salt and 10 mM NaPi
(adjusted the pH to 7.4 after dissolution of the salts) was added
to a 96-well plate (Greiner). The purified sensor was added, and the
fluorescence intensities at 475 and 525 nm were recorded separately
in a Spark 10 M microplate reader with excitation at 420 nm at RT.
A 20 nm bandwidth for excitation and emission was applied, and the
average of 10 measurements of a single well was taken. The background
fluorescence, buffer without sensor, was subtracted.
Transfection
and Imaging of HEK293 Cells
Transfection
and imaging of HEK293 cells was performed as described,[21] with some minor modifications. HEK293 cells
(ATCC CRL-1573, tested for mycoplasma contamination) were cultured
in DMEM (Gibco) supplemented with 10% (v/v) fetal calf serum, 2 mM l-glutamine (Gibco), 100 units mL–1 penicillin
(Invitrogen), and 100 μg mL–1 streptomycin
(Invitrogen). For transfection experiments, HEK293 cells were plated
in eight-well Labtek glass chamber slides (Thermo Scientific) at 6
× 104 cells per well. One day after plating, the cells
were transfected with plasmid DNA coding for the sensors as follows:
lipoplexes composed of 1.5 μL of Lipofectamine 2000 (Invitrogen)
and 0.5 μg of the pcDNA 3.1 vector carrying the corresponding
sensor gene were prepared in 100 μL of serum-free DMEM, according
to the manufacturer’s instructions. A total of 25 μL
of lipoplex solution was added per well and incubated for 4 h at 37
°C and 5% CO2, after which the medium was refreshed.
The next day, the medium was replaced by 200 μL of DMEM with
HEPES without phenol red, and sensor expression in the HEK293 cells
was subsequently analyzed by confocal fluorescence microscopy. The
cells were imaged directly in the eight-well Labtek glass chamber
slides. The slides were mounted on a laser-scanning confocal microscope
(Zeiss LSM 710), at 37 °C. The sensor was excited using a 405
nm LED laser, and the emission was split into a 450–505 nm
channel and 505–797 nm channel. The fluorescence intensity
of the cells was determined in ImageJ for each channel. The backgrounds
for each channel were subtracted and the mCitrine intensity divided
by the mCerulean intensity for each cell. When the mCitrine intensity
was plotted versus the mCerulean intensity, linear fits with R2 > 0.99 were generally obtained.
Authors: Iwo König; Arash Zarrine-Afsar; Mikayel Aznauryan; Andrea Soranno; Bengt Wunderlich; Fabian Dingfelder; Jakob C Stüber; Andreas Plückthun; Daniel Nettels; Benjamin Schuler Journal: Nat Methods Date: 2015-07-06 Impact factor: 28.547
Authors: Roland Moussa; Anna Baierl; Victoria Steffen; Tina Kubitzki; Wolfgang Wiechert; Martina Pohl Journal: J Biotechnol Date: 2014-08-05 Impact factor: 3.307
Authors: Jack W Shepherd; Sarah Lecinski; Jasmine Wragg; Sviatlana Shashkova; Chris MacDonald; Mark C Leake Journal: Methods Date: 2020-11-04 Impact factor: 3.608
Authors: Cody P Aplin; Robert C Miller; Taryn M Kay; Ahmed A Heikal; Arnold J Boersma; Erin D Sheets Journal: Biophys J Date: 2021-02-12 Impact factor: 4.033
Authors: Diana Morzy; Roger Rubio-Sánchez; Himanshu Joshi; Aleksei Aksimentiev; Lorenzo Di Michele; Ulrich F Keyser Journal: J Am Chem Soc Date: 2021-05-07 Impact factor: 15.419