Hideki Shiihara1, Taiji Sakamoto2, Takehiro Yamashita1, Naoko Kakiuchi1, Hiroki Otsuka1, Hiroto Terasaki1, Shozo Sonoda1.
Abstract
This study was performed to compare the area of the foveal avascular zone (FAZ-area) obtained by three optical coherence tomography angiography (OCTA) instruments. This was a cross-sectional, non-interventional study of twenty-seven healthy right eyes. The superficial and deep FAZ-area was measured manually with three OCTA instruments: Triton (Topcon), RS3000 (Nidek), and CIRRUS (Zeiss). The intra-rater, inter-rater, and inter-instrument correlation coefficients (CC) were assessed. The intra-rater and inter-rater CC were significantly high for the superficial and deep FAZ-areas (P < 0.001). The inter-instrument CC (95% confidence interval) for the superficial FAZ-area was 0.920 (0.803-0.965) for Triton vs RS3000, 0.899 (0.575-0.965) for RS3000 vs CIRRUS, and was 0.963 (0.913-0.983) for CIRRUS vs Triton (P < 0.001). For the deep FAZ-area, the inter-instrument CC was 0.813 (0.633-0.910) for Triton vs RS3000, 0.694 (0.369-0.857) for RS3000 vs CIRRUS, and 0.679 (0.153-0.872) for CIRRUS vs Triton (P < 0.001). The superficial FAZ-area (mm2) was 0.264 ± 0.071 with Triton, 0.278 ± 0.072 with RS3000 and 0.257 ± 0.066 with CIRRUS. For deep FAZ-area, it was 0.617 ± 0.175 with Triton, 0.646 ± 0.178 with RS3000 and 0.719 ± 0.175 with CIRRUS. The FAZ-area from these instruments was clinically interchangeable. However, the absolute values of FAZ-area are significantly different among them. These differences must be considered in comparing the FAZ-areas from different OCTA instruments.Entities:
Mesh:
Year: 2017 PMID: 28851930 PMCID: PMC5575252 DOI: 10.1038/s41598-017-09255-5
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Intra-rater comparison.
| FAZ | Triton | RS3000 | CIRRUS | ||||||
|---|---|---|---|---|---|---|---|---|---|
| ICC (95%CI) | CV | P value | ICC | CV | P value | ICC | CV | P value | |
| Superficial | 0.987 (0.972–0.994) | 0.26 | <0.001 | 0.987 (0.972–0.994) | 0.28 | <0.001 | 0.991 (0.981–0.996) | 0.26 | <0.001 |
| Deep | 0.990 (0.978–0.995) | 0.29 | <0.001 | 0.991 (0.980–0.996) | 0.28 | <0.001 | 0.995 (0.989–0.998) | 0.24 | <0.001 |
FAZ, foveal avascular zone; ICC, intraclass correlation coefficients; CV, coefficient of variation.
Inter-rater comparison.
| FAZ | Triton | RS3000 | CIRRUS | ||||||
|---|---|---|---|---|---|---|---|---|---|
| ICC (95%CI) | CV | P value | ICC | CV | P value | ICC | CV | P value | |
| Superficial | 0.973 (0.941–0.987) | 0.28 | <0.001 | 0.964 (0.922–0.983) | 0.28 | <0.001 | 0.986 (0.969–0.994) | 0.26 | <0.001* |
| Deep | 0.992 (0.983–0.996) | 0.28 | <0.001 | 0.981 (0.960–0.996) | 0.27 | <0.001 | 0.986 (0.969–0.994) | 0.25 | <0.001 |
FAZ, foveal avascular zone; ICC, intraclass correlation coefficients; CV, coefficient of variation.
Comparison of two instruments for superficial FAZ.
| Mean Difference, mm2 ± SD | Range of Difference, mm2 | 95% CI | P value Wilcoxon signed rank test | ICC | 95% CI | P value | |
|---|---|---|---|---|---|---|---|
| Triton -RS3000 | −0.013 ± 0.026 | −0.076 to 0.045 | −0.024 to −0.003 | 0.012 | 0.92 | 0.803–0.965 | <0.001 |
| RS3000 -CIRRUS | 0.021 ± 0.025 | −0.030 to 0.064 | 0.011 to 0.031 | <0.001 | 0.899 | 0.575–0.965 | <0.001 |
| CIRRUS -Triton | −0.007 ± 0.180 | −0.048 to 0.026 | −0.015 to 0.000 | 0.049 | 0.963 | 0.913–0.983 | <0.001 |
CI, confidence interval; ICC, intraclass correlation coefficients.
Comparison of two instruments for deep FAZ.
| Mean Difference mm2 ± SD | Range of Difference mm2 | 95% CI | P value Wilcoxon signed rank test | ICC | 95% CI | P value | |
|---|---|---|---|---|---|---|---|
| Triton -RS3000 | −0.030 ± 0.108 | −0.344 to 0.142 | −0.073 to 0.013 | 0.230 | 0.813 | 0.633–0.910 | <0.001 |
| RS3000 -CIRRUS | −0.073 ± 0.129 | −0.382 to 0.245 | −0.124 to −0.022 | 0.006 | 0.694 | 0.369–0.857 | <0.001 |
| CIRRUS -Triton | 0.103 ± 0.117 | −0.216 to 0.372 | 0.056 to 0.149 | <0.001 | 0.679 | 0.153–0.872 | <0.001 |
CI, confidence interval; ICC, intraclass correlation coefficients.
Figure 1Bland-Altman plot analyses for the superficial FAZ. There was a fixed bias among the three instruments but there was no proportional bias between any two instruments.
Figure 2Bland-Altman plot analysis for the deep FAZ. There was a fixed bias between the RS3000 and CIRRUS and between the CIRRUS and Triton. However, there was no fixed bias between the Triton and RS3000. There was no proportional bias between any two machines.
Figure 3Comparison of 3 × 3 images of superficial foveal avascular zone (FAZ) of the same eye obtained by the Triton, RS3000, and CIRRUS optical coherence tomography angiographic (OCTA) instruments. Representative 3 × 3 retina OCTA images obtained from the same eye with the Triton instrument (top left with red bordering), RS3000 (top middle with green bordering) and CIRRUS (top right with yellow bordering). The superimposed FAZ images are shown in the bottom row. Left; Triton and RS3000, middle; RS3000 and CIRRUS, right; CIRRUS and Triton. The CIRRUS FAZ image is always the largest area among the three instruments.
Scanning Protocols of OCTA instruments.
| Optical source wave length (nm) | Scan speed (A-Scan/sec) | Axial resolution (μm) | Transversal resolution (μm) | Axial adjustment | |
|---|---|---|---|---|---|
| Triton | 1,050 | 100,000 | 8 | 20 | available |
| RS3000 | 880 | 53,000 | 7 | 20 | available |
| CIRRUS | 840 | 68,000 | 5 | 15 | unavailable |