Literature DB >> 28847446

Sequential treatment with ProACT™ device implantation after male sling failure for male urinary incontinence.

M G Baron1, C Delcourt2, F-X Nouhaud2, A Gillibert3, C Pfister2, P Grise2, J-N Cornu2.   

Abstract

OBJECTIVES: There is no strong evidence for second line therapy after male sling failure for post-prostatectomy urinary incontinence (PPUI). We report the outcomes after periurethral balloons implantation for persistence or recurrence of mild PPI symptoms after male sling implantation.
METHODS: All patients implanted of a ProACT™ device (Uromedica, Inc., MN, USA) following I-STOP transobturator male sling (TOMS) failure, in a tertiary reference center between 2009 and 2016, were included. Patients were evaluated by 24-hour pad-test before and after implantation, and after each balloon repressurizing procedure. PGI-I and Likert scale patient satisfaction were estimated during a telephone interview conducted in 2016. Objective and subjective cure of urinary incontinence were defined by a 24-hour pad-test<8g and the use of zero or one pad per day, respectively.
RESULTS: Fourteen patients were included. Median follow-up was 34months [4-89]. Objective and subjective cure were 29% (n=4) and 57% (n=8), respectively. Median pad-test decreased from 95g [IQR: 130] to 34g [IQR: 83] (P=0.022). ProACT™ significantly decreased median pad-test by a factor 2.73 [1.19-6.29]. Eighty-eight percent patients were feeling a little better, much better or very much better and 77% were satisfied or very satisfied at the end of follow-up. Reoperation rate was 28% (n=4): 3 balloons were changed for caudally migration (n=2) or deflation (n=1) and 1 had a urinary sphincter implanted for severe UI.
CONCLUSIONS: ProACT™ is a safe and efficient treatment that can be used in second line therapy after TOMS failure for PPUI. LEVEL OF EVIDENCE: 4.
Copyright © 2017 Elsevier Masson SAS. All rights reserved.

Entities:  

Keywords:  Ballons ProACT™; Bandelettes I-STOP TOMS; Continence device; I-STOP TOMS; Incontinence urinaire de l’homme; Male urinary incontinence; ProACT™; Prostatectomie radicale; Radical prostatectomy

Mesh:

Year:  2017        PMID: 28847446     DOI: 10.1016/j.purol.2017.08.001

Source DB:  PubMed          Journal:  Prog Urol        ISSN: 1166-7087            Impact factor:   0.915


  2 in total

1.  ATOMS (Adjustable Transobturator Male System) Is an Effective and Safe Second-Line Treatment Option for Recurrent Urinary Incontinence after Implantation of an AdVance/AdVance XP Fixed Male Sling? A Multicenter Cohort Analysis.

Authors:  Fabian Queissert; Keith Rourke; Sandra Schönburg; Alessandro Giammò; Andreas Gonsior; Carmen González-Enguita; Antonio Romero; Andres J Schrader; Francisco Cruz; Francisco E Martins; Juan F Dorado; Javier C Angulo
Journal:  J Clin Med       Date:  2021-12-24       Impact factor: 4.241

2.  Systematic review and meta-analysis comparing Adjustable Transobturator Male System (ATOMS) and Adjustable Continence Therapy (ProACT) for male stress incontinence.

Authors:  Javier C Angulo; Sandra Schönburg; Alessandro Giammò; Francisco J Abellán; Ignacio Arance; David Lora
Journal:  PLoS One       Date:  2019-12-02       Impact factor: 3.240

  2 in total