Cardiac mapping and pulmonary vein isolation using a novel ablation catheter with tip minielectrodes.

Florian Heringhaus1, Tanja Lotz1, Lena Loehr1, Julia Gelep1, Sebastian Lask1, Kaffer Kara1, Andreas Mügge1, Alexander Wutzler1.   

Abstract

BACKGROUND: Pulmonary vein isolation (PVI) is a standard treatment for atrial fibrillation (AF). Identification of gaps in the ablation line is difficult. Tip-ring electrograms from ablation catheters represent relative large areas of myocardial tissue. Recently, an ablation catheter with three minielectrodes (ME) on the catheter tip with closer interelectrode spacing was introduced. The aim of our study was to evaluate the novel electrodes during PVI.
METHODS: PVI was performed with an irrigated ablation catheter equipped with conventional electrodes and three additional radial tip electrodes. Detection of pulmonary vein potentials (PVPs), local signal amplitude, amplitude reduction during ablation, and loss of capture after ablation were compared between the ME and the conventional tip-ring electrodes.
RESULTS: Thirty-one patients (mean age 67.8 ± 10.3 years, 45.2 % men) were included. A total of 306 mapping/lesion points were analyzed. A PVP was significantly more often obtained with the ME compared to the conventional tip-ring electrodes (99.2% vs 83.5%, P < 0.001). Local amplitude was significantly higher on the ME (0.8 ± 0.6 mV vs 0.67 ± 0.46 mV, P  =  0.003). Amplitude reduction during 1 RF pulse was significantly greater on the ME (82.9 ± 19.5% vs. 61.8 ± 26.9%, P < 0.001). During pace mapping, loss of capture after 1 RF pulse was observed significantly more often on ME (98.3% vs 63.3%, P < 0.001).
CONCLUSION: Signal amplitude is higher and sensitivity during PVP mapping and ablation is increased when ME are used. ME may facilitate catheter ablation of AF in the future.
© 2017 Wiley Periodicals, Inc.

Entities:  

Keywords:  catheter ablation; mapping; minielectrodes; pulmonary vein isolation

Mesh:

Year:  2017        PMID: 28846147     DOI: 10.1111/pace.13185

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Ablation of typical atrial flutter using mini electrode measurements for maximum voltage-guided ablation: A randomized, controlled trial.

Authors:  Matthew K Rowe; Andrew Claughton; Jason Davis; Lauren Yee; Gerald C Kaye; Kieran Dauber; John Hill; Paul A Gould
Journal:  J Arrhythm       Date:  2021-12-09
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.