New procedure of quantitative mapping of Ti and Al released from dental implant and Mg, Ca, Fe, Zn, Cu, Mn as physiological elements in oral mucosa by LA-ICP-MS.

Adam Sajnóg1, Anetta Hanć1, Ryszard Koczorowski2, Danuta Barałkiewicz3.   

Abstract

A new procedure for determination of elements derived from titanium implants and physiological elements in soft tissues by laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) is presented. The analytical procedure was developed which involved preparation of in-house matrix matched solid standards with analyte addition based on certified reference material (CRM) MODAS-4 Cormorant Tissue. Addition of gelatin, serving as a binding agent, essentially improved physical properties of standards. Performance of the analytical method was assayed and validated by calculating parameters like precision, detection limits, trueness and recovery of analyte addition using additional CRM - ERM-BB184 Bovine Muscle. Analyte addition was additionally confirmed by microwave digestion of solid standards and analysis by solution nebulization ICP-MS. The detection limits are in range 1.8μgg-1 to 450μgg-1 for Mn and Ca respectively. The precision values range from 7.3% to 42% for Al and Zn respectively. The estimated recoveries of analyte addition line within scope of 83%-153% for Mn and Cu respectively. Oral mucosa samples taken from patients treated with titanium dental implants were examined using developed analytical method. Standards and tissue samples were cryocut into 30µm thin sections. LA-ICP-MS allowed to obtain two-dimensional maps of distribution of elements in tested samples which revealed high content of Ti and Al derived from implants. Photographs from optical microscope displayed numerous particles with µm size in oral mucosa samples which suggests that they are residues from implantation procedure.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Elemental analysis; Implants; LA-ICP-MS; Method validation; Quantitative analysis; Soft tissues

Mesh:

Substances:

Year:  2017        PMID: 28842005     DOI: 10.1016/j.talanta.2017.07.058

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  3 in total

Review 1.  The unfavorable role of titanium particles released from dental implants.

Authors:  Zilan Zhou; Quan Shi; Jie Wang; Xiaohang Chen; Yujia Hao; Yuan Zhang; Xing Wang
Journal:  Nanotheranostics       Date:  2021-03-10

Review 2.  Latest Trends in Surface Modification for Dental Implantology: Innovative Developments and Analytical Applications.

Authors:  Francesca Accioni; Juan Vázquez; Manuel Merinero; Belén Begines; Ana Alcudia
Journal:  Pharmaceutics       Date:  2022-02-21       Impact factor: 6.321

3.  [Aluminium release of glitter particles in removable orthodontic appliances].

Authors:  Lena Wepner; Harald Andreas Färber; Anna Weber; Andreas Jaensch; Ludger Keilig; Florian Andreas Heuser; Christoph Peter Bourauel
Journal:  Bundesgesundheitsblatt Gesundheitsforschung Gesundheitsschutz       Date:  2021-06-11       Impact factor: 1.513

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.