| Literature DB >> 28835757 |
Pilar I Beato-Víbora1, S Alejo-González2.
Abstract
INTRODUCTION: Interfering antibodies are capable of causing potentially misleading results in automated thyroid hormone immunoassays. CASEEntities:
Keywords: Antibody; Free Thyroxine; Immunoassay; Interference
Year: 2016 PMID: 28835757 PMCID: PMC5554610 DOI: 10.5812/ijem.37792
Source DB: PubMed Journal: Int J Endocrinol Metab ISSN: 1726-913X
Thyroid Hormone Levels
| TSH (µUI/mL), | FT4 (pg/mL), | FT4 (pg/mL), | FT4 (pg/mL), | FT4 (pg/mL), | FT3 (pg/mL), | TT4 (pg/mL), | TT4 (pg/mL), | TT4 (pg/mL), | |
|---|---|---|---|---|---|---|---|---|---|
| [0.5 - 5] | [8 - 18] | [8 - 18] | [9.3 - 17] | [8 - 18] | [2.3 - 4.2] | [4.5 - 10.9] | [4.5 - 10.9] | [4.5 - 10.9] | |
| Before PEG | Double-Dilution Test | After PEG | Before PEG | Double-Dilution Test | After PEG | ||||
|
| |||||||||
| Method 1 | 20.95 | 7.7 | - | - | - | - | |||
|
| |||||||||
| Method 1 | 1.486 | 12.3 | - | - | - | - | |||
|
| |||||||||
| Method 1 | 2.903 | 14.5 | - | - | - | - | |||
|
| |||||||||
| Method 1 | 7.396 | 51.7 | (2) 48; (4) 61; (8) 82 | - | 11 | 2.8 | 67.9 | (2) > 30; (4) 93.6; (8) 93.6 | 3.8 |
| Method 2 | 9.333 | - | - | 7.4 | - | 2.97 |
Abbreviations: TSH, serum thyrotropin; FT4, free thyroxin; FT3, free triiodothyronine; TT4, total thyroxin; Method 1, Siemens Healthcare Reagents Kit, Advia centaur; Method 2, Roche Cobas “e”, Roche Diagnostics Corporation; PEG, polyethylene glycol; Double serial dilution from the primary specimen were analyzed at the same run in an Advia centaur platform.