Axial or Helical? Considerations for wide collimation CT scanners capable of volumetric imaging in both modes.

Jack W Lambert1, Elizabeth D Phillips1, Javier E Villanueva-Meyer1, Lorenzo Nardo1, Luca Facchetti1, Robert G Gould1.   

Abstract

PURPOSE: To determine whether axial or helical mode is more appropriate for a 16 cm collimation CT scanner capable of step-and-shoot volumetric axial coverage, in terms of radiation dose, image quality, and scan duration.
METHODS: All scans were performed with a Revolution CT (GE Healthcare) operating at 120 kV and 100 mAs. Using calibrated optically stimulated luminescence detectors, radiation dose along the axial scan profile was evaluated at the isocenter, including the overlap region between two axial sections. This overlap region measures 3 cm in the z-axis at the isocenter and is required to obtain sufficient projection data from the relatively large cone-beam angles. Using an image quality phantom (Gammex Model 464), spatial resolution, CT number uniformity, image noise, and low contrast detectability (LCD) were evaluated under five different conditions: in the middle of a helical acquisition, in the middle of a 16 cm axial section, at both ends of an axial section and in the overlap region between two axial sections. Scan durations and dose length products (DLP) were recorded for prescribed scan lengths of 2.5-100 cm.
RESULTS: The overlap region between two axial sections received a dose 83% higher than the single-exposure region at the isocenter. Within a single axial section, the dose at the anode end was 37% less than at the cathode end due to the anode heel effect. Image noise ranged from a low of 13 HU for the cathode end of an axial section up to 14.7 HU for the anode end (P < 0.001). The LCD was at lower at the anode end of the axial section compared to both the cathode end (P < 0.05) and the overlap location (P < 0.02). The spatial resolution and CT number uniformity were consistent among all conditions. Scan durations were shorter (0.28 s) for the axial mode compared to the helical mode at scan lengths ≤ 16 cm, and longer at scan lengths ≥ 16 cm where more than one table position was required, up to a difference of 13.9 s for a the 100 cm scan length (3.8 s for helical compared to 17.6 s for axial). DLPs were consistent between scan modes; slightly lower in axial mode at shorter scan lengths due to helical overranging, and slightly higher in axial mode at longer scan lengths due to the axial overlap regions.
CONCLUSIONS: To ensure the most consistent radiation dose and image quality along the scan length, we recommend helical mode for scans longer than the 16 cm coverage of a single axial section. For scan lengths ≤ 16 cm, axial scanning is the most practical option, with a shorter scan duration and higher dose efficiency.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  256 detector row; axial mode; image quality; radiation dose; step-and-shoot

Mesh:

Year:  2017        PMID: 28833277     DOI: 10.1002/mp.12525

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Relative Perfusion Differences between Parathyroid Adenomas and the Thyroid on Multiphase 4DCT.

Authors:  Steven P M J Raeymaeckers; Yannick De Brucker; Maurizio Tosi; Nico Buls; Johan De Mey
Journal:  Int J Biomed Imaging       Date:  2022-05-20

Review 2.  High-frequency Multiphase 4DCT for the Detection of Parathyroid Adenomas: A Pictorial Essay.

Authors:  Steven Raeymaeckers; Yannick De Brucker; Maurizio Tosi; Nico Buls; Johan De Mey
Journal:  J Endocr Soc       Date:  2021-07-31

3.  Detection of parathyroid adenomas with multiphase 4DCT: towards a true four-dimensional technique.

Authors:  Steven Raeymaeckers; Yannick De Brucker; Tim Vanderhasselt; Nico Buls; Johan De Mey
Journal:  BMC Med Imaging       Date:  2021-04-07       Impact factor: 1.930

4.  Study on Key Technologies of Virtual Interactive Surgical Simulation for 3D Reconstruction of Medical Images.

Authors:  Feng Qiu; Qingfu Zhu; Ke Jing; Yufei Duan
Journal:  Biomed Res Int       Date:  2022-08-02       Impact factor: 3.246

5.  Measurement accuracy of lung nodule volumetry in a phantom study: Effect of axial-volume scan and iterative reconstruction algorithm.

Authors:  Han Na Lee; Jung Im Kim; So Youn Shin
Journal:  Medicine (Baltimore)       Date:  2020-06-05       Impact factor: 1.817

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.