How accurate are TheraMon® microsensors at measuring intraoral wear-time? Recorded vs. actual wear times in five volunteers.

Catherine A Brierley1,2, Philip E Benson1, Jonathan Sandler2.   

Abstract

BACKGROUND: The TheraMon® microsensor is the most recent device developed to measure the wear-time of removable appliances. The accuracy has not been validated intraorally.
OBJECTIVES: To determine 1) if the TheraMon® microsensor accurately records time when fixed intraorally, and 2) the effect of the intraoral location on the recorded time.
METHODS: A prospective pilot study, using a convenience sample, was carried out in a UK hospital orthodontic department. Five non-patient volunteers wore microsensors positioned palatal to an upper molar, and buccal to a lower molar for 7 days. Differences between actual amount of wear and the wear recorded by each device were calculated. Differences between sites were also examined.
RESULTS: The mean daily wear-time recorded by the upper and lower microsensors combined was 23 hours (95% CI 22.6-23.4), which is a mean under-recording of 4% (CI 2.5-5.8%). The maximum daily under-reporting of wear times was 5.5 hours. Microsensors in the lower buccal sulcus recorded wear-times that were closer to actual wear-times.
CONCLUSIONS: Assumptions made by the TheraMon® microsensors software lead to under-reporting of intraoral wear-time, particularly when placed palatally. These discrepancies could be significant in both clinical practice and research. Adjustment of the microsensor software parameters would improve accuracy, irrespective of the intraoral location.

Entities:  

Keywords:  Microsensor; removable appliance; wear-time

Mesh:

Year:  2017        PMID: 28826346     DOI: 10.1080/14653125.2017.1365220

Source DB:  PubMed          Journal:  J Orthod        ISSN: 1465-3125


  4 in total

1.  Frankel 2 appliance versus the Modified Twin Block appliance for Phase 1 treatment of Class II division 1 malocclusion in children and adolescents: A randomized clinical trial.

Authors:  Ciara Campbell; Declan Millett; Niamh Kelly; Marie Cooke; Michael Cronin
Journal:  Angle Orthod       Date:  2019-10-15       Impact factor: 2.079

2.  The Use of Microsensors to Assess the Daily Wear Time of Removable Orthodontic Appliances: A Prospective Cohort Study.

Authors:  Marek Nahajowski; Joanna Lis; Michał Sarul
Journal:  Sensors (Basel)       Date:  2022-03-22       Impact factor: 3.576

3.  Determining the performance of a temperature sensor embedded into a mouthguard.

Authors:  Leonardo de Almeida E Bueno; William Milnthorpe; Jeroen H M Bergmann
Journal:  BDJ Open       Date:  2022-08-01

4.  Orthodontic retention: what's on the horizon?

Authors:  Simon J Littlewood; Oyku Dalci; Calogero Dolce; L Shannon Holliday; Sasan Naraghi
Journal:  Br Dent J       Date:  2021-06-11       Impact factor: 1.626

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.