| Literature DB >> 28824402 |
Elena Ratti1, Shani Waninger2, Chris Berka2, Giulio Ruffini3, Ajay Verma1.
Abstract
Objectives: To compare quantitative EEG signal and test-retest reliability of medical grade and consumer EEG systems.Entities:
Keywords: clinical trials; consumer EEG; electroencephalogram; electrophysiology; neurophysiology; quantitative EEG
Year: 2017 PMID: 28824402 PMCID: PMC5540902 DOI: 10.3389/fnhum.2017.00398
Source DB: PubMed Journal: Front Hum Neurosci ISSN: 1662-5161 Impact factor: 3.169
Study population demographics and baseline characteristics.
| Age (years) | 25 | 23 | 27 | 37 | 23 | 27 ± 5.8 |
| Gender (male) | Male | Male | Female | Male | Female | 60% |
| Handedness (right) | Right | Right | Right | Right | Right | 100% |
| Education (years) | 17 Undergraduate Degree | 17 Undergraduate degree | 17 Undergraduate degree | 19 Advanced Degree | 17 Undergraduate degree | 17.4 ± 0.9 |
| Caffeine consumption (number of coffees/day) | 2–3 | 1–2 | 1 | 1–2 | 0–1 | 1.8 ± 0.8 |
| Alcohol consumption (number of drinks/day) | 0 | 0 | 1 | 0–2 | 0–1 | 0.8 ± 0.8 |
| Smoking | No | No | No | No | No | 0% |
Categorical variables: N (%); Continuous variables: Mean ± Standard Deviation (SD).
Figure 1Anatomical channel distribution of the multi- and limited-channel EEG systems. Representation of the shared and unique channels among the four EEG systems evaluated. The common channel to all the four EEG systems, Fp1 (blue); channels unique to the Muse system (red); channels shared between the multi-channel systems, B-Alert and Enobio (green).
EEG systems comparisons.
| Channels | 2 | 1 | 20 | 20 |
| Sampling rate | 220 Hz | 512 Hz | 256 Hz | 500 Hz |
| Wet/dry electrodes | Dry | Dry | Wet | Wet |
| Signal quality check | Yes | Yes | Yes | Yes |
| Impedance check | n/a | n/a | Yes | Yes |
| Setup time (minutes) | 5 | 3 | 20–25 | 20–25 |
Figure 2Extracts of EO raw data from B-Alert (A) and Enobio (B) Muse (C) and MindWave (D) within the same participant during the same visit.
Figure 3Extracts of EC raw data from B-Alert (A) and Enobio (B) Muse (C) and MindWave (D) within the same participant during the same visit.
Figure 4Fp1 Power Spectral Densities (PSDs) from all four EEG systems in the EO condition at Visit 1 (A) and Visit 2 (B) and EC condition at Visit 1 (C) and Visit 2 (D).
Figure 5Test-retest ratios for EC (A) and EO (B) condition.