Comparison of Point-of-Care Versus Laboratory Troponin Testing in an Emergency Department Setting.

Michael Juliano1, Courtney Wason2.   

Abstract

BACKGROUND: There have been concerns regarding troponin results accuracy between point-of-care (POC) testing preformed in an emergency department (ED) setting and laboratory testing. The purpose of this study was to compare the results of cardiac troponin I testing as concerns POC testing and laboratory analysis as a way to show that these results are interchangeable.
METHODS: A retrospective chart review was performed from October 2012 through September 2013 to identify all patients who presented to the ED that received both a POC and laboratory troponin associated with the same blood draw.
FINDINGS: A total of 189 patients met inclusion criteria. Three laboratory samples were hemolyzed and not available for comparison. Of the remaining 186 samples, when a positive cutoff value of 0.034 ng/mL was used: 37 POC tests were positive and 149 POC tests were negative, sensitivity 0.85 (95% confidence interval [CI] = 0.69-0.94), specificity 0.98 (95% CI = 0.94-0.99), positive predictive value 91.9% (95% CI = 77-97.9%), and negative predictive value 96% (95% CI = 91.1-98.4%). When a POC troponin positive cutoff value of 0.12 ng/mL was used: 20 POC tests were positive and 164 were negative: sensitivity 1.0 (95% CI = 0.8-1.0), specificity 0.99 (95% CI = 0.95-1.0), positive predictive value 91% (95% CI = 69-98.4%), and negative predictive value 100% (95% CI = 97.1-100%). DISCUSSION: In an ED setting; POC testing is accurate and correlates well with laboratory testing. Considering that the POC analysis takes a fraction of time to yield results, the validity of its data is important. The results of this study show that the POC system is a suitable test for rapid evaluation of patients presenting to the ED. This correlation increased with a higher testing positive cutoff value (0.12 versus 0.034 ng/mL). These data can assist the emergency physician with rapid identification of evidence of cardiac injury and be assured that the results are accurate. RECOMMENDATIONS: POC lab testing should be considered whenever possible to provide the fastest troponin results. Reprint &
Copyright © 2017 Association of Military Surgeons of the U.S.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28810994     DOI: 10.7205/MILMED-D-16-00387

Source DB:  PubMed          Journal:  Mil Med        ISSN: 0026-4075            Impact factor:   1.437


  5 in total

1.  Modern diagnostics in emergency medicine.

Authors:  Jan Niederdöckl; Nina Buchtele; Michael Schwameis; Hans Domanovits
Journal:  Wien Klin Wochenschr       Date:  2021-03       Impact factor: 1.704

Review 2.  Evaluation of Molecularly Imprinted Polymers for Point-of-Care Testing for Cardiovascular Disease.

Authors:  Brian Regan; Fiona Boyle; Richard O'Kennedy; David Collins
Journal:  Sensors (Basel)       Date:  2019-08-09       Impact factor: 3.576

3.  Performance of ivisen IA-1400, a new point-of-care device with an internal centrifuge system, for the measurement of cardiac troponin I levels.

Authors:  Ha Nui Kim; Soo-Young Yoon
Journal:  J Clin Lab Anal       Date:  2021-03-17       Impact factor: 2.352

4.  Comparison of Point-of-Care and Highly Sensitive Laboratory Troponin Testing in Patients Suspicious of Acute Myocardial Infarction and Its Efficacy in Clinical Outcome.

Authors:  Sahand Mohammadzadeh; Nasim Matani; Neda Soleimani; Hamed Bazrafshan Drissi
Journal:  Cardiol Res Pract       Date:  2022-02-24       Impact factor: 1.866

5.  Point-of-Care Diagnosis of Acute Myocardial Infarction in Central Vietnam: International Exchange, Needs Assessment, and Spatial Care Paths.

Authors:  Gerald J Kost; Amanullah Zadran; Thuan T B Duong; Tung T Pham; An V D Ho; Nhan V Nguyen; Irene J Ventura; Layma Zadran; Mykhaylo V Sayenko; Kelly Nguyen
Journal:  Point Care       Date:  2018-08-27
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.