Pressure wire versus microcatheter for FFR measurement: a head-to-head comparison.

Christophe Pouillot1, Stephane Fournier, Jens Glasenapp, Goeffray Rambaud, Karim Bougrini, Richard Vi Fane, Christophe Geyer, Julien Adjedj.   

Abstract

AIMS: Recently developed microcatheters can be used instead of a pressure wire for fractional flow reserve (FFR) measurement. We sought to assess the haemodynamic and clinical impact of using a larger profile device to measure FFR. METHODS AND
RESULTS: Our prospective registry included 77 consecutive patients who underwent invasive FFR measurement of intermediate coronary stenoses between June 2015 and July 2016. FFR values were obtained first using a pressure wire only (FFRw), second using a Navvus microcatheter (FFRMC), and finally using the wire with the microcatheter still in the stenosis (FFRw-MC) during intravenous adenosine infusion. Eighty-eight stenoses were suitable for a thorough head-to-head comparison. Mean FFRw (0.83±0.08) was significantly higher than mean FFRMC (0.80±0.10) and FFRw-MC (0.80±0.10). Mean FFRMC and FFRw-MC did not differ significantly. Bland-Altman analysis showed a bias of -0.03±0.05 for lower FFRMC values compared to FFRw values. Using a threshold of 0.80 for FFR, the indication for revascularisation would have differed when based on FFRMC versus FFRw in 20/88 (23%) of the lesions and 18/77 (23%) of the patients.
CONCLUSIONS: FFR measured using a microcatheter overestimates stenosis severity, leading to erroneous indication for revascularisation in a sizeable proportion of cases.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 28804057     DOI: 10.4244/EIJ-D-17-00238

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  3 in total

1.  Prognostic value of post-percutaneous coronary intervention diastolic pressure ratio.

Authors:  K Masdjedi; L J C van Zandvoort; T Neleman; I Kardys; J Ligthart; W K Den Dekker; R Diletti; F Zijlstra; N M Van Mieghem; J Daemen
Journal:  Neth Heart J       Date:  2022-04-07       Impact factor: 2.854

2.  Fractional flow reserve derived from microcatheters versus standard pressure wires: a stenosis-level meta-analysis.

Authors:  Henry Seligman; Matthew J Shun-Shin; Anushkumar Vasireddy; Christopher Cook; Yousif Y Ahmad; James Howard; Sayan Sen; Rasha Al-Lamee; Sukhjinder Nijjer; Daniel Chamie; Justin Davies; Jamil Mayet; Darrel P Francis; Ricardo Petraco
Journal:  Open Heart       Date:  2019-03-25

3.  Extent of the difference between microcatheter and pressure wire-derived fractional flow reserve and its relation to optical coherence tomography-derived parameters.

Authors:  Yoshiki Matsuo; Yasutsugu Shiono; Kuninobu Kashiyama; Yasushi Ino; Takahiro Nishi; Kosei Terada; Hiroki Emori; Daisuke Higashioka; Yosuke Katayama; Amir Khalifa Mahfouz; Teruaki Wada; Suwako Fujita; Masahiro Takahata; Kunihiro Shimamura; Manabu Kashiwagi; Akio Kuroi; Atsushi Tanaka; Takeshi Hozumi; Takashi Kubo; Takashi Akasaka
Journal:  Int J Cardiol Heart Vasc       Date:  2020-03-13
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.