BACKGROUND: In revision total hip arthroplasty (THA), proximal femoral bone loss creates a challenge of achieving adequate stem fixation. The purpose of this study was to examine the outcomes of a monoblock, splined, tapered femoral stem in revision THA.
METHODS: Outcomes of revision THA using a nonmodular, splined, tapered femoral stem from a single surgeon were reviewed. With a minimum of 2-year follow-up, there were 68 cases (67 patients). Paprosky classification was 3A or greater in 85% of the cases. Preoperative and postoperative Harris Hip Scores (HHS), radiographic subsidence and osseointegration, limb length discrepancy, complications, and reoperations were analyzed.
RESULTS: The Harris Hip Score improved from 37.4 ± SD 19.4 preoperatively to 64.6 ± SD 21.8 at final follow-up (P < .001). There were 16 revision procedures-8 for septic indications and 8 for aseptic indications. Subsidence occurred at a rate of 3.0% and dislocation at 7.4%. Limb length discrepancy of more than 1 cm after revision was noted in 13.6% of patients. Bone ingrowth was observed in all but 4 patients (94.1%). At 4-year follow-up, Kaplan-Meier estimated survival was 72.9% (95% confidence interval [CI] 57.0-83.8) for all causes of revision, 86.6% (95% CI 72.0-93.9) for all aseptic revision, and 95.5% (95% CI 86.8-98.5) for aseptic femoral revision.
CONCLUSION: Although complications were significant, revision for femoral aseptic loosening occurred in only 3 patients. Given the ability of this monoblock splined tapered stem to adequately provide fixation during complex revision THA, it remains a viable option in the setting of substantial femoral bone defects.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Wagner-SL stem; femoral component; monoblock stem; nonmodular stem; proximal femoral bone loss; revision total hip arthroplasty

Mesh:

Substances:

Year:  2017        PMID: 28803814     DOI: 10.1016/j.arth.2017.06.045

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  2 in total

1.  Femoral revision for periprosthetic fracture in total hip arthroplasty.

Authors:  Luke G Menken; Jose A Rodriguez
Journal:  J Clin Orthop Trauma       Date:  2019-12-14

2.  Differences between proximal bone remodeling in femoral revisions for aseptic loosening and periprosthetic fractures using the Wagner SL stem.

Authors:  Gábor Friebert; Csaba Gombár; András Bozó; Ilona Polyák; Ádám Brzózka; Krisztián Sisák
Journal:  BMC Musculoskelet Disord       Date:  2021-02-17       Impact factor: 2.362

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.