| Literature DB >> 28788753 |
Anna Zamir, Paul C Diemoz, Fabio A Vittoria, Charlotte K Hagen, Marco Endrizzi, Alessandro Olivo.
Abstract
In this paper we present a single-image phase retrieval algorithm for multi-material samples, developed for the edge illumination (EI) X-ray phase contrast imaging method. The theoretical derivation is provided, along with any assumptions made. The algorithm is evaluated quantitatively using both simulated and experimental results from a computed tomography (CT) scan using the EI laboratory implementation. Qualitative CT results are provided for a biological sample containing both bone and soft-tissue. Using a single EI image per projection and knowledge of the complex refractive index, the algorithm can accurately retrieve the interface between a given pair of materials. A composite CT slice can be created by splicing together multiple CT reconstructions, each retrieved for a different pair of materials.Year: 2017 PMID: 28788753 DOI: 10.1364/OE.25.011984
Source DB: PubMed Journal: Opt Express ISSN: 1094-4087 Impact factor: 3.894