Silent navigator-triggered silent MRI of the abdomen.

Yuji Iwadate1, Atsushi Nozaki1, Yoshinobu Nunokawa2, Shigeo Okuda3, Masahiro Jinzaki3, Hiroyuki Kabasawa1.   

Abstract

PURPOSE: To develop and demonstrate the feasibility of a silent respiratory navigator technique for prospective triggering, which was incorporated into a three-dimensional radial zero-echo-time sequence for respiratory navigated silent abdominal imaging.
METHODS: A nonselective hard excitation radiofrequency pulse was used for the navigator sequence with a derated readout gradient, to avoid generation of high levels of acoustic noise. The acquired navigator signals were processed in real time and used for prospective triggering of the zero-echo-time sequence. Ten healthy volunteers were scanned using the proposed and conventional techniques at 1.5 T. An acoustic noise measurement with A-weighted continuous equivalent sound pressure level was also performed.
RESULTS: The sound pressure-level values of the background noise, zero-echo-time imaging, conventional, and silent navigators were 68.3, 68.4, 102.5, and 69.4 dBA, respectively. Excellent correlation with correlation coefficients greater than 0.9 was observed between the bellows signals and displacement values calculated from the navigators. Sharpness of the portal vein of both conventional and silent navigator-triggered images was significantly higher than those of nontriggered images.
CONCLUSIONS: The silent navigator-triggered zero-echo-time technique is feasible and might improve image quality and workflow of abdominal MRI of patients who are prone to acoustic noise. Magn Reson Med 79:2170-2175, 2018.
© 2017 International Society for Magnetic Resonance in Medicine. © 2017 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  abdominal imaging; navigator; respiratory motion; silent MRI

Mesh:

Year:  2017        PMID: 28782173     DOI: 10.1002/mrm.26869

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  2 in total

1.  Qualitative and quantitative analysis of 3D T1 Silent imaging.

Authors:  Francesca Di Giuliano; Silvia Minosse; Eliseo Picchi; Valentina Ferrazzoli; Valerio Da Ros; Massimo Muto; Chiara Adriana Pistolese; Francesco Garaci; Roberto Floris
Journal:  Radiol Med       Date:  2021-06-15       Impact factor: 3.469

2.  Comparison of Silent Navigator Waveform Generation Methods.

Authors:  Yuji Iwadate; Atsushi Nozaki; Yoshinobu Nunokawa; Shigeo Okuda; Hiroyuki Kabasawa; Masahiro Jinzaki
Journal:  Magn Reson Med Sci       Date:  2019-05-30       Impact factor: 2.471

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.