Computational modelling of motion at the bone-implant interface after total knee arthroplasty: The role of implant design and surgical fit.

Noel Conlisk1, Colin R Howie2, Pankaj Pankaj3.   

Abstract

BACKGROUND: Aseptic loosening, osteolysis, and infection are the most commonly reported reasons for revision total knee arthroplasty (TKA). This study examined the role of implant design features (e.g. condylar box, pegs) and stems in resisting loosening, and also explored the sensitivity of the implants to a loose surgical fit due to saw blade oscillation.
METHODS: Finite element models of the distal femur implanted with four different implant types: cruciate retaining (CR), posterior stabilising (PS), total stabilising (TS) with short stem (12mm×50mm), and a total stabilising (TS) with long stem (19mm×150mm) were developed and analysed in this study. Two different fit conditions were considered: a normal fit, where the resections on the bone exactly match the internal profile of the implant, and a loose fit due to saw blade oscillation, characterised by removal of one millimetre of bone from the anterior and posterior surfaces of the distal femur. Frictional interfaces were employed at the bone-implant interfaces to allow relative motions to be recorded.
RESULTS: The results showed that interface motions increased with increasing flexion angle and loose fit. Implant design features were found to greatly influence the surface area under increased motion, while only slightly influencing the values of peak motion. Short uncemented stems behaved similarly to PS implants, while long canal filling stems exhibited the least amount of motion at the interface under any fit condition.
CONCLUSION: In conclusion, long stemmed prostheses appeared less susceptible to surgical cut errors than short stemmed and stemless implants.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Finite element analysis; Implant design features; Primary knee arthroplasty; Revision knee arthroplasty; Saw blade error; Surgical resections

Mesh:

Year:  2017        PMID: 28778499     DOI: 10.1016/j.knee.2017.07.003

Source DB:  PubMed          Journal:  Knee        ISSN: 0968-0160            Impact factor:   2.199


  1 in total

1.  Design, Modeling, and Evaluation of the Eddy Current Sensor Deeply Implanted in the Human Body.

Authors:  Rajas Prakash Khokle; Karu P Esselle; Desmond J Bokor
Journal:  Sensors (Basel)       Date:  2018-11-11       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.