The NeutraClear® needleless connector is equally effective against catheter colonization compared to MicroClave®.

María Guembe1,2, María Jesús Pérez Granda2,3,4, Raquel Cruces1,2, Luis Alcalá1, Emilio Bouza1,3,5.   

Abstract

INTRODUCTION: Neutral-valve closed-system connectors can reduce the frequency of catheter colonization. Commercially available closed system connectors need to be tested and compared with each other to assess how they protect against contamination. We aimed to compare, in vitro, the efficacy of connectors NeutraClear® and MicroClave® against contamination under conditions of daily clinical practice.
METHODS: The model consisted of a set of 200 blood culture bottles (BCBs) with a cannula inserted (100 closed with NeutraClear® and 100 closed with MicroClave®) that were assessed in two experiments while instilling 1 mL of saline: manipulation based on the standard of care and manipulation using gloves impregnated with a 0.05 McFarland Staphylococcus aureus solution. The BCBs were incubated in a BACTEC System at 37°C under continuous shaking for up to 7 days. When a bottle turned positive, 100 µL of the fluid was cultured. The positivity rate and time to positivity of the BCB in each experiment was compared.
RESULTS: In the aseptic model in the NeutraClear® and MicroClave® groups, only 1 BCB and 2 BCBs were positive, respectively, (p = 0.55). In the contaminated model, all BCBs were positive in both groups at the end of the incubation time. We did not find differences for the MTP between NeutraClear® and MicroClave® (36.04 vs. 20.13 hours, p = 0.09).
CONCLUSIONS: The NeutraClear® needleless connector proved to be as efficient as the MicroClave® connector in the prevention of catheter colonization and migration of S. aureus from the surface to the inside of the hub in an in vitro model.

Entities:  

Mesh:

Year:  2017        PMID: 28777419     DOI: 10.5301/jva.5000775

Source DB:  PubMed          Journal:  J Vasc Access        ISSN: 1129-7298            Impact factor:   2.283


  1 in total

1.  Even the Simplest Devices May Malfunction: Split Septum Design Revisited.

Authors:  Rotem Naftalovich; Steven Char; Andrew J Iskander; Daniel Naftalovich
Journal:  Biomed Instrum Technol       Date:  2021-01-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.