Yong Wang1, Xiang-Mei Chen2, Guang-Yan Cai1, Wen-Ge Li3, Ai-Hua Zhang4, Li-Rong Hao5, Ming Shi6, Rong Wang7, Hong-Li Jiang8, Hui-Min Luo9, Dong Zhang1, Xue-Feng Sun1.
Abstract
OBJECTIVE: To evaluate the in vivo and in vitro performance of a China-made dialysis machine (SWS-4000).Entities:
Keywords: China-made; Hemodialysis machine; Performance; Safety
Mesh:
Year: 2017 PMID: 28768540 PMCID: PMC5541422 DOI: 10.1186/s12938-017-0387-y
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Comparison of changes in Kt/V and URR values after dialysis between the two types of dialysis machines
| SWS-4000 | Fresenius 4008 | |||
|---|---|---|---|---|
| Week 1 | Week 4 | Week 1 | Week 4 | |
| Kt/V | 1.34 ± 0.49 | 1.39 ± 0.66 | 1.36 ± 0.44 | 1.42 ± 0.64 |
| URR | 0.66 ± 0.08 | 0.66 ± 0.95 | 0.66 ± 0.08 | 0.67 ± 0.09 |
Fig. 1Comparison of changes in Kt/V values after dialysis between the China-made and imported dialysis machines
Fig. 2Comparison of changes in URR values after dialysis between the China-made and imported dialysis machines
Comparison of changes in blood biochemical parameters after dialysis between the two types of dialysis machines
| SWS-4000 | Fresenius 4008 | |||
|---|---|---|---|---|
| Week 1 | Week 4 | Week 1 | Week 4 | |
| Cr decline | 586.8 ± 168.8 | 599.4 ± 183.8 | 576.4 ± 171.3 | 590.9 ± 174.9 |
| K+ decline | 1.5 ± 0.6 | 1.5 ± 0.6 | 1.5 ± 0.5 | 1.4 ± 0.6 |
| HCO3 − change | 4.7 ± 3.5 | 4.9 ± 3.5 | 5.3 ± 3.8 | 6.3 ± 7.3 |
Fig. 3Comparison of blood Cr changes after dialysis between the China-made and imported dialysis machines
Fig. 4Comparison of blood K+ changes after dialysis between the China-made and imported dialysis machines
Ultrafiltration errors for the two types of dialysis machines at different time points under different ultrafiltration rate settings
| Test time | Ultrafiltration rate (mL/h) | Actual error (ml) | |
|---|---|---|---|
| SWS-4000 | Fresenius 4008 | ||
| Week 1 | 200 | 7.9 ± 12.0 | 6.6 ± 8.6 |
| 400 | 10.5 ± 17.9 | 4.4 ± 6.4 | |
| 600 | 4.9 ± 5.4 | 6.4 ± 10.0 | |
| 1000 | 20.5 ± 23.2 | 14.1 ± 3.8 | |
| Week 12 | 200 | 8.2 ± 11.8 | 6.8 ± 9.6 |
| 400 | 8.4 ± 12.5 | 4.9 ± 9.7 | |
| 600 | 16.6 ± 27.5 | 8.8 ± 12.5 | |
| 1000 | 14.1 ± 15.4 | 14.6 ± 14.0 | |
| Week 23 | 200 | 10.7 ± 14.5 | 10.6 ± 13.7 |
| 400 | 16.0 ± 20.0 | 6.6 ± 13.2 | |
| 600 | 11.3 ± 13.2 | 8.8 ± 13.7 | |
| 1000 | 16.8 ± 16.3 | 25.1 ± 27.9 | |
Errors of ion concentrations in proportioned dialysate for the two types of dialysis machines at different time points
| Test time | Ion (mmol/L) | Actual error (ml) | |
|---|---|---|---|
| SWS-4000 | Fresenius 4008 | ||
| Week 1 | K+ | 0.12 ± 0.10 | 0.09 ± 0.10 |
| Na+ | 1.40 ± 1.74 | 1.45 ± 1.37 | |
| Cl− | 1.40 ± 1.40 | 1.35 ± 0.94 | |
| Ga2+ | 0.06 ± 0.05 | 0.04 ± 0.05 | |
| HCO3 − | 1.69 ± 1.73 | 1.45 ± 0.91 | |
| Week 4 | K+ | 0.11 ± 0.10 | 0.11 ± 0.11 |
| Na+ | 1.89 ± 1.80 | 2.65 ± 3.07 | |
| Cl− | 2.06 ± 2.20 | 2.43 ± 2.04 | |
| Ga2+ | 0.06 ± 0.05 | 0.05 ± 0.05 | |
| HCO3 − | 1.54 ± 1.45 | 1.96 ± 1.82 | |
| Week 8 | K+ | 0.09 ± 0.10 | 0.13 ± 0.11 |
| Na+ | 1.37 ± 1.53 | 1.95 ± 1.18 | |
| Cl− | 1.55 ± 1.97 | 1.84 ± 0.87 | |
| Ga2+ | 0.11 ± 0.11 | 0.05 ± 0.05 | |
| HCO3 − | 2.06 ± 1.41 | 1.29 ± 1.25 | |
| Week 12 | K+ | 0.10 ± 0.11 | 0.10 ± 0.10 |
| Na+ | 1.93 ± 1.43 | 1.70 ± 1.63 | |
| Cl− | 2.05 ± 1.70 | 1.23 ± 1.03 | |
| Ga2+ | 0.05 ± 0.05 | 0.03 ± 0.03 | |
| HCO3 − | 1.70 ± 1.92 | 1.14 ± 0.89 | |
| Week 16 | K+ | 0.10 ± 0.11 | 0.13 ± 0.11 |
| Na+ | 3.08 ± 1.74 | 2.26 ± 1.73 | |
| Cl− | 1.98 ± 1.91 | 1.60 ± 1.18 | |
| Ga2+ | 0.08 ± 0.09 | 0.06 ± 0.06 | |
| HCO3 − | 1.24 ± 0.90 | 1.15 ± 1.58 | |
| Week 20 | K+ | 0.17 ± 0.15 | 0.15 ± 0.16 |
| Na+ | 2.04 ± 1.33 | 1.88 ± 1.51 | |
| Cl− | 3.05 ± 2.66 | 1.66 ± 1.77 | |
| Ga2+ | 0.07 ± 0.06 | 0.03 ± 0.04 | |
| HCO3 − | 2.36 ± 1.85 | 2.04 ± 2.69 | |
| Week 24 | K+ | 0.13 ± 0.10 | 0.16 ± 0.24 |
| Na+ | 3.20 ± 2.64 | 2.03 ± 2.42 | |
| Cl− | 1.93 ± 2.22 | 2.25 ± 1.91 | |
| Ga2+ | 0.10 ± 0.07 | 0.06 ± 0.05 | |
| HCO3 − | 1.61 ± 1.78 | 2.52 ± 2.48 | |
Errors of heparin injection rate for the two types of dialysis machines at different time points under different injection settings
| Test time | Heparin pump (mL/h) | Actual error (ml) | |
|---|---|---|---|
| SWS-4000 | Fresenius 4008 | ||
| Week 1 | 2 | 0.0 | 0.0 |
| 4 | 0.04 ± 0.13 | 0.03 ± 0.08 | |
| Week 12 | 2 | 0.0 | 0.0 |
| 4 | 0.04 ± 0.07 | 0.01 ± 0.05 | |
| Week 23 | 2 | 0.10 ± 0.21 | 0.0 |
| 4 | 0.13 ± 0.23 | 0.08 ± 0.19 | |
Errors of blood pump flow rate for the two types of dialysis machines at different time points under different settings
| Test time | Blood pump flow rate (mL/min) | Actual error (ml) | |
|---|---|---|---|
| SWS-4000 | Fresenius 4008 | ||
| Week 2 | 100 | 69.6 ± 128.7 | 41.8 ± 78.2 |
| 200 | 146.9 ± 212.3 | 79.5 ± 76.1 | |
| 300 | 307.5 ± 320.5 | 86.0 ± 86.5* | |
| 400 | 385.4 ± 451.8 | 287.8 ± 334.3 | |
| Week 13 | 100 | 80.4 ± 118.1 | 20.6 ± 37.6 |
| 200 | 175.0 ± 191.3 | 118.8 ± 155.1 | |
| 300 | 326.3 ± 501.3 | 133.4 ± 179.2 | |
| 400 | 475.8 ± 517.9 | 258.2 ± 213.5 | |
| Week 24 | 100 | 95.7 ± 115.6 | 65.0 ± 80.7 |
| 200 | 220.6 ± 253.3 | 118.1 ± 158.5 | |
| 300 | 449.0 ± 291.7 | 139.4 ± 121.6* | |
| 400 | 736.0 ± 701.5 | 328.1 ± 402.6 | |
* P < 0.05, vs. SWS-4000 at the same flow rate
Fig. 5The structural design of SWS-4000 dialysis machine
Fig. 6The structural design of Fresenius 4008 dialysis machine