Reliability and Validity of a Smartphone-Paired Pulse Oximeter for Screening of Critical Congenital Heart Defects in Newborns.

Maurice J Huizing1, Eduardo Villamor-Martínez, Ingrid A Chavagne, Ward Y Vanagt, Marc A E Spaanderman, Eduardo Villamor.   

Abstract

BACKGROUND: Barriers to widespread implementation of pulse oximetry screening of critical congenital heart defects (CCHD) in newborns include increasing trends of out-of-hospital births and cost of equipment. In recent years, smartphone-compatible pulse oximeters have appeared on the market, but the validity of such devices in the setting of CCHD screening has not been evaluated.
OBJECTIVES: To compare the performance in CCHD screening of a smartphone-paired pulse oximeter (Masimo iSpO2-Rx™) and a hospital-grade pulse oximeter (Masimo Radical-7™).
METHODS: Preductal (right hand) and postductal (either foot) saturations were determined in a population of 201 term newborns by 2 independent teams, one using the Radical-7 and the other using the iSpO2-Rx. Bland-Altman analysis was applied to calculate mean bias and 95% limits of agreement between the 2 pulse oximeters.
RESULTS: For the preductal oxygen saturation, the mean bias (Radical-7 minus iSpO2-Rx) was -0.08 (SD 1.76) and the lower and upper limits of agreement were -3.52 and 3.36, respectively. For the postductal oxygen saturation, the mean bias was -0.11 (SD 1.68) and the lower and upper limits of agreement were -3.49 and 3.18, respectively. In addition, the iSpO2-Rx provided reliable measurements of saturations below 95% in a group of 12 infants admitted to the neonatal intensive care unit.
CONCLUSIONS: Our data suggest that CCHD screening with the Masimo iSpO2-Rx is feasible and accurate. The use of reliable smartphone-paired pulse oximeters may contribute to the extension of CCHD screening to home births and low resource settings.
© 2017 S. Karger AG, Basel.

Entities:  

Keywords:  Critical congenital heart defects; Pulse oximetry; Screening; Smartphone

Mesh:

Substances:

Year:  2017        PMID: 28768288     DOI: 10.1159/000477294

Source DB:  PubMed          Journal:  Neonatology        ISSN: 1661-7800            Impact factor:   4.035


  2 in total

1.  Reliability of Smartphone Applications for the Quantification of Oxygen Saturation.

Authors:  Almas F Khattak; Susan S Kakakhel; Noman K Wazir; Madiha Khattak; Tania Khattak; Faryal Akbar
Journal:  Cureus       Date:  2021-11-09

2.  Mobile Health for Central Sleep Apnea Screening Among Patients With Stable Heart Failure: Single-Cohort, Open, Prospective Trial.

Authors:  Roderick Willem Treskes; Arie C Maan; Harriette Florence Verwey; Robert Schot; Saskia Lambertha Maria Anna Beeres; Laurens F Tops; Enno Tjeerd Van Der Velde; Martin Jan Schalij; Annelies Margaretha Slats
Journal:  JMIR Cardio       Date:  2019-03-19
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.