Renate List1, Marina Hitz1, Michael Angst1, William R Taylor1, Silvio Lorenzetti2.
Abstract
Force plates are common assessment tools used in biomechanics to measure ground reaction forces during motion capture or strength exercises. While the accuracy of the resulting kinetic data is critical for accurate gait analysis, factors such as plate mounting are thought to influence the measured point of force application (PFA) of the ground reaction forces. The aim of this study was therefore to present an accurate in-situ PFA calibration method. In addition, the approach was evaluated by examining the required PFA corrections over a period of ten years, which included plate remounting and a change of foundation. The in-situ PFA calibration was performed on six plates by assessing the locations of up to 98 points of force application per plate using an instrumented pole. Application of the in-situ PFA calibration approach reduced the root mean square errors by up to approximately 60% compared to the manufacturers calculation. Correction coefficients were strongly dependent on the individual platform as well as the location of the applied force on the plate. Remounting of the plates altered the plate coefficient corrections considerably, while changes over time were notable but not as extensive. We therefore recommend that plates should be recalibrated after remounting, but also at least every 5 years, in order to ensure the preservation of an accurate PFA.Entities:
Keywords: Ground reaction force plates; In-situ calibration; Point of force application
Mesh:
Year: 2017 PMID: 28763716 DOI: 10.1016/j.gaitpost.2017.07.111
Source DB: PubMed Journal: Gait Posture ISSN: 0966-6362 Impact factor: 2.840