The accuracy of static pressure measurement with water-filled urodynamic systems.

Andrew Gammie1.   

Abstract

AIMS: To quantify the inaccuracy of pressure measurement by water-filled systems, with a view to recommending better practice.
METHODS: Papers examining quality of measurements were reviewed, and standards documents examined. Technical data from manufacturers were gathered and summarized. Measurements were made on two different urodynamic systems. Individual inaccuracies were combined using the root sum of squares method.
RESULTS: The total inaccuracy in measurement may reach up to 9.6 cmH2 O on pdet , but with good practice this may be mitigated to 1.3 cmH2 O.
CONCLUSIONS: Water-filled systems, while being the recommended method for urodynamic measurements, are subject to a potential inaccuracy during poor use of up to approximately 10 cmH2 O on pdet . Good practice can reduce this inaccuracy to a tolerable level, that is, approximately 1 cmH2 O.
© 2017 Wiley Periodicals, Inc.

Entities:  

Keywords:  error; good practice; inaccuracy; water-filled

Mesh:

Substances:

Year:  2017        PMID: 28763564     DOI: 10.1002/nau.23358

Source DB:  PubMed          Journal:  Neurourol Urodyn        ISSN: 0733-2467            Impact factor:   2.696


  2 in total

1.  Head-to-head comparison of pressures during full cystometry, with clinical as well as in-depth signal-analysis, of air-filled catheters versus the ICS-standard water-filled catheters.

Authors:  Peter F W M Rosier
Journal:  Neurourol Urodyn       Date:  2021-08-07       Impact factor: 2.367

2.  Good urodynamic practice: Pressure signal quality immediately after catheter insertion for cystometry with a water-filled pressure transducer system and its relevance for the ICS zero procedure.

Authors:  Peter F W M Rosier
Journal:  Neurourol Urodyn       Date:  2020-11-03       Impact factor: 2.367

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.