Abstract

BACKGROUND: Cardiac implantable electronic device therapy (CIED) has revolutionized treatment for advanced heart failure. Most patients considered for orthotopic heart transplantation (OHT) are treated with implantable cardioverter defibrillators, cardiac resynchronization therapy, or both. These CIEDs are surgically extracted at the time of transplant. Occasionally, CIEDs are incompletely removed. Little is known about the outcomes of post-OHT patients with retained CIED fragments.
METHODS: We identified 200 consecutive patients that underwent OHT at our institution between April 2006 and December 2014 and performed a retrospective analysis of available radiographic images and clinical records. Chest radiographs prior to and following OHT were reviewed for the presence of CIED or retained CIED fragments. The outcomes of patients with retained CIED fragments that had subsequent magnetic resonance imaging (MRI) studies performed were further investigated.
RESULTS: One hundred eighty of 200 patients were identified as having CIED prior to OHT, of which 29 had retained CIED fragments after OHT. Most retained CIED fragments originated from superior vena cava defibrillator coils. There were no adverse events in the retained CIED fragment cohort, and survival was unaffected. Ten patients with retained CIED fragments safely underwent a total of 28 MRIs after OHT, all of diagnostic quality.
CONCLUSION: Retained CIED fragments are not associated with adverse events or increased mortality after OHT. Diagnostic MRI has been safely performed in patients with retained CIED fragments after incomplete device extraction. Retrieval of these fragments prior to MRI does not appear warranted given the demonstrated safety and preserved image quality in this population.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28760213     DOI: 10.1016/j.ahj.2017.04.011

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  4 in total

Review 1.  [Safe MRI examinations in patients with pacemakers and ICD].

Authors:  D Beitzke
Journal:  Radiologe       Date:  2019-10       Impact factor: 0.635

2.  Safety and image quality of cardiovascular magnetic resonance imaging in patients with retained epicardial pacing wires after heart transplantation.

Authors:  Constantin Gatterer; Marie-Elisabeth Stelzmüller; Andreas Kammerlander; Andreas Zuckermann; Martin Krššák; Christian Loewe; Dietrich Beitzke
Journal:  J Cardiovasc Magn Reson       Date:  2021-03-15       Impact factor: 5.364

3.  Subcutaneous implantable cardioverter defibrillator in patients awaiting cardiac transplantation or left ventricular assist device for refractory heart failure: a feasible alternative to transvenous device?

Authors:  Federico Migliore; Giacomo Cavalli; Tomaso Bottio; Pietro De Franceschi; Emanuele Bertaglia; Gino Gerosa; Sabino Iliceto
Journal:  ESC Heart Fail       Date:  2018-01-27

4.  Safety of MRI in patients with retained cardiac leads.

Authors:  Bach T Nguyen; Bhumi Bhusal; Amir Ali Rahsepar; Kate Fawcett; Stella Lin; Daniel S Marks; Rod Passman; Donny Nieto; Richard Niemzcura; Laleh Golestanirad
Journal:  Magn Reson Med       Date:  2021-12-27       Impact factor: 3.737

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.