Literature DB >> 28758853

One-year clinical outcomes of patients implanted with a Resolute Onyx™ zotarolimus-eluting stent.

Chor Cheung Tam1, Kelvin Chan1, Simon Lam1, Arthur Yung1, Yui Ming Lam1, Carmen Chan1, David Siu1, Hung Fat Tse1.   

Abstract

Objective To evaluate the 1-year clinical outcomes of patients who received the Resolute Onyx™ stent. Methods This was a single-centre, retrospective registry analysis that reviewed the clinical data from all patients who were implanted with a Resolute Onyx™ stent between March 2015 and February 2016. Clinical follow-up was performed at 1 year post-implantation. Results A total of 252 patients received a Resolute Onyx™ stent and two patients were lost to follow-up. The mean age of the cohort was 66.9 years and 113 (45.2%) had diabetes mellitus. Thirty-eight patients (15.2%) had left main disease and 73 (29.2%) had three-vessel disease. A total of 175 patients (70.0%) had small vessel disease (<2.75 mm) and 210 (84.0%) had long lesions (>20 mm). The 1-year target lesion failure was 4.4% (11 of 250), cardiovascular death occurred in eight patients (3.2%), ischaemia-driven target lesion revascularization was undertaken in five patients (2.0%) and stent thrombosis occurred in one patient (0.4%). Conclusion The Resolute Onyx™ stent showed a favourable 1-year clinical performance in a real-world population.

Entities:  

Keywords:  Drug-eluting stent; coronary artery disease; zotarolimus-eluting stent

Mesh:

Substances:

Year:  2017        PMID: 28758853      PMCID: PMC6011307          DOI: 10.1177/0300060517717826

Source DB:  PubMed          Journal:  J Int Med Res        ISSN: 0300-0605            Impact factor:   1.671


Introduction

Drug-eluting stents (DES) are widely used for the treatment of obstructive coronary artery disease. Despite concerns related to very late stent thrombosis associated with first-generation DES, continuing advancement in stent design, polymer and drug-eluting technology has improved the efficacy and safety of second-generation DES.[1,2] The Resolute™ Integrity zotarolimus-eluting stent (R-ZES; Medtronic Inc., Santa Rosa, CA, USA) has been extensively studied in the Resolute Global Clinical Program,[3] which demonstrated favourable clinical outcomes in an all-comer population. The latest version of R-ZES, Resolute Onyx™, was launched in November 2014. This study investigated the 1-year clinical outcomes of patients who had received this new technology.

Patients and methods

Study design and population

This was a single-centre retrospective registry analysis. From the launch of Resolute Onyx™, all patients treated in the Division of Cardiology, Queen Mary Hospital, University of Hong Kong, Hong Kong who had been fitted with a Resolute Onyx™ stent between March 2015 and February 2016 were identified in the hospital database and recruited to the analysis. There were no exclusion criteria except patients lost to follow-up or those who could not be contacted for assessment at 1 year after stent implantation. Baseline demographics, procedural details and follow-up data were recorded. All treatment and clinical decision-making processes were subject to the attending cardiologists. The study was approved by the Institutional Ethics Committee of Queen Mary Hospital, University of Hong Kong, Hong Kong (no. UW 17-026). Written informed consent was waived by the Institutional Ethics Committee as this was a retrospective analysis without any alteration in patient management.

Study endpoints

The primary endpoint was 1-year target lesion failure (TLF), defined as a composite of cardiac death, myocardial infarction (not clearly attributed to a non-target vessel), or a clinically indicated target lesion revascularization by percutaneous or surgical methods at 12 months. All clinical outcomes were defined according to the Academic Research Consortium (ARC).[4] Deaths were considered cardiac unless an unequivocal non-cardiac cause could be established. Secondary endpoints included individual components of the primary outcomes: target vessel myocardial infarction; clinically-driven target lesion revascularization (TLR), defined as repeat revascularization within the stented segment including 5 mm proximal and distal border zones; clinically-driven target vessel revascularization (TVR) and stent thrombosis defined according to the ARC as definite, probable, or possible.[4] A revascularization was considered clinically-driven if repeat angiography showed a diameter stenosis ≥50% with at least one of the following: (i) history of recurrent angina pectoris, presumably related to the target vessel; (ii) objective signs of ischaemia at rest or during exercise test by electrocardiogram, presumably related to the target vessel; (iii) abnormal test results of invasive functional diagnostic test (fractional flow reserve); or (iv) a TLR with a diameter stenosis ≥70% even in the absence of the aforementioned ischaemic signs or symptoms.

Study device

Resolute Onyx™ (Medtronic Inc., Santa Rosa, CA, USA) is a zotarolimus-eluting stent designed for higher procedural success and better acute performance. The design is based on its predecessor, the Resolute™ Integrity drug-eluting stent, which was extensively studied in the Resolute Global Clinical Program.[3] In brief, the stent is manufactured using continuous sinusoid technology that employs a single strand of wire to form a sinusoidal pattern of crowns, which is laser fused at selected crowns. This distinctive wire-forming facilitates smooth tracking in tortuous vessels and allows excellent conformability.[5] The drug polymer BioLinx is highly biocompatible with a hydrophilic surface designed to minimize inflammation.[6] The polymer elutes the drug zotarolimus, which is a synthetic analogue of sirolimus, over 180 days, with >85% of the zotarolimus dose released by day 60.[7] The latest version Resolute Onyx™ utilizes core-wire technology in which the cobalt alloy incorporates a platinum iridium core for increased radiopacity. Moreover, the strut thickness is further reduced to 81 µm (0.0032 inches) with maintained radial strength and the crossing profile is 0.037 mm for the 2.5 mm stent. The Resolute Onyx™ stents used in patients included in this registry study ranged from 2.0 mm to 5.0 mm diameters and the stent lengths were up to 38 mm.

Statistical analyses

All statistical analyses were performed using the SPSS® statistical package, version 19.0 (SPSS Inc., Chicago, IL, USA) for Windows®. Continuous variables are as mean ± SD. Categorical values are presented as absolute values or percentages. Comparison of continuous data was performed by one-way analysis of variance, Mann–Whitney U-test or independent-samples t-test depending on data distribution. χ2-test and Fisher’s exact test were used to compare categorical data. A two-sided P-value of ≤0.05 was considered statistically significant.

Results

From March 2015 to February 2016, 252 patients were implanted with Resolute Onyx™ stents. Two patients were lost to follow-up, so 250 patients completed the 1-year assessment and were entered into this analysis. A total of 577 Resolute Onyx™ stents were implanted. Baseline demographics are presented in Table 1; 89 patients (35.6%) had type 2 diabetes mellitus (T2DM), 24 (9.6%) had type 1 diabetes mellitus and 21 patients (8.4%) with end-stage renal failure (ESRF) were on dialysis (haemodialysis and peritoneal dialysis). Interventional procedures and lesion details are presented in Tables 2 and 3, respectively. A total of 83 patients (33.2%) presented with acute coronary syndrome including ST elevation myocardial infarction and eight patients (3.2%) had cardiogenic shock. With regard to lesion characteristics, 38 patients (15.2%) had left main involvement, 73 (29.2%) had three-vessel disease; and 175 (70.0%) had small vessel diseases with Resolute Onyx™ 2.5 mm (n = 164), 2.25 mm (n = 111) and 2.0 mm (n = 42) stents implanted. At 1 year, 228 of 239 surviving patients (95.4%) were on dual antiplatelet therapy.
Table 1.

Demographic and clinical characteristics of patients who were implanted with Resolute Onyx™ stents between March 2015 and February 2016.

CharacteristicCohort who received Resolute Onyx™ stents n = 250
Age, years66.9 ± 10.7
Sex, males185 (74.0)
Type 2 diabetes mellitus89 (35.6)
Type 1 diabetes mellitus24 (9.6)
Hypertension166 (66.4)
Previous myocardial infarction38 (15.2)
Previous percutaneous coronary intervention88 (35.2)
Previous coronary bypass grafting10 (4.0)
End stage renal failure on dialysis21 (8.4)

Data presented as mean ± SD or n of patients (%).

Table 2.

Procedural characteristics of patients who were implanted with Resolute Onyx™ stents between March 2015 and February 2016.

Procedural characteristicCohort who received Resolute Onyx™ stents n = 250
ST elevation myocardial infarction22 (8.8)
Non-ST elevation myocardial infarction/ unstable angina61 (24.4)
Stable angina167 (66.8)
Cardiogenic shock8 (3.2)
Class III/IV heart failure11 (4.4)
Left ventricular ejection fraction <35%22 (8.8)
Intravascular image guided intervention (intravascular ultrasound/ optic coherence tomography)96 (38.4)

Data presented as n of patients (%).

Table 3.

Lesion characteristics of patients who were implanted with Resolute Onyx™ stents between March 2015 and February 2016.

Lesion characteristicCohort who received Resolute Onyx™ stents n = 250
Left main disease38 (15.2)
One-vessel disease84 (33.6)
Two-vessel disease93 (37.2)
Three-vessel disease73 (29.2)
Small vessel <2.75 mm175 (70.0)
Long lesion >20 mm210 (84.0)
Chronic total occlusion29 (11.6)
In-stent restenosis34 (13.6)
Bifurcation (two stents)11 (4.4)

Data presented as n of patients (%).

Demographic and clinical characteristics of patients who were implanted with Resolute Onyx™ stents between March 2015 and February 2016. Data presented as mean ± SD or n of patients (%). Procedural characteristics of patients who were implanted with Resolute Onyx™ stents between March 2015 and February 2016. Data presented as n of patients (%). Lesion characteristics of patients who were implanted with Resolute Onyx™ stents between March 2015 and February 2016. Data presented as n of patients (%). Table 4 shows the 1-year clinical outcomes. The primary endpoint of 1-year TLF was observed in 11 patients (4.4%). One patient suffered from ARC probable subacute stent thrombosis. When patients were stratified according to vessel size, the demographics and 1-year clinical outcomes were similar for patients with and without small vessel disease (Table 5). Of the 21 patients with ESRF who were receiving dialysis during the index procedure, five patients (23.8%) suffered from TLF within 1 year.
Table 4.

The 1-year clinical outcomes of patients who were implanted with Resolute Onyx™ stents between March 2015 and February 2016.

Clinical outcomeCohort who received Resolute Onyx™ stents n = 250
Target lesion failure11 (4.4)
Target vessel failure13 (5.2)
Cardiovascular death8 (3.2)
Target vessel myocardial infarction5 (2.0)
Clinically-indicated target lesion revascularization5 (2.0)
Clinically-indicated target vessel revascularization7 (2.8)
Stent thrombosis (definite/probable)1 (0.4)

Data presented as n of patients (%).

Table 5.

Comparison of patients with or without small vessel disease who were implanted with Resolute Onyx™ stents between March 2015 and February 2016.

CharacteristicPatients with small vessel disease n = 175Patients without small vessel disease n = 75
Age, years67.2 ± 10.366.3 ± 11.5
Sex, males132 (75.4)53 (70.7)
Diabetes mellitus82 (46.9)31 (41.3)
Target lesion failure8 (4.6)3 (4.0)
Target vessel failure10 (5.7)3 (4.0)
Cardiovascular death7 (4.0)1 (1.3)
Target vessel myocardial infarction3 (1.7)2 (2.7)
Target lesion revascularization3 (1.7)2 (2.7)
Target vessel revascularization5 (2.9)2 (2.7)
Stent thrombosis (definite/probable)1 (0.6)0 (0.0)

Data presented as mean ± SD or n of patients (%).

No significant between-group differences (P > 0.05).

The 1-year clinical outcomes of patients who were implanted with Resolute Onyx™ stents between March 2015 and February 2016. Data presented as n of patients (%). Comparison of patients with or without small vessel disease who were implanted with Resolute Onyx™ stents between March 2015 and February 2016. Data presented as mean ± SD or n of patients (%). No significant between-group differences (P > 0.05).

Discussion

To the best of our knowledge, this is the first report of the medium-term clinical performance of the latest evolution of R-ZES, the Resolute Onyx™ stent. These present data demonstrated that the Resolute Onyx™ stent achieved similar efficacy and safety compared with most contemporary DES with a 1-year TLF rate of 4.4%.[1] In the Resolute All Comers trial,[8] the Resolute zotarolimus-eluting stent and an everolimus-eluting stent had 1-year TLF rates of 8.2% and 8.3%, respectively. Although this current retrospective registry has its inherent limitations, this post-marketing study allowed the analysis of all patients including those with ultra-high-risk characteristics. For example, this current cohort included patients with cardiogenic shock, patients with ESRF receiving dialysis, a high proportion of patients with T2DM and even patients who presented with cardiac arrest requiring haemodynamic support. This current cohort reflects real-world coronary intervention practices, which is difficult to study in prospective randomized clinical trials. In this present study, many of the TLFs occurred in those ultra-high-risk patients and this further provides insight into the use of the Resolute Onyx™ stent in a real-world population. In the present study, 45% of patients had diabetes mellitus and 70% had small vessel disease. Small vessel stenting has traditionally been associated with higher rates of restenosis and repeat revascularizations.[9,10] The degree of neointimal proliferation is relatively independent of the reference vessel size and hence the proportion of cross-sectional luminal narrowing is higher in small vessels compared with large vessels.[11] The problem is further complicated by the presence of T2DM that predisposes to neointimal hyperplasia while patients with T2DM tend to have diffuse small vessel disease.[12] These current data suggest that the Resolute Onyx™ stent had similar outcomes in small and large vessel diseases, which is similar to its predecessor stent as demonstrated by the Resolute Global Clinical Program small vessel subgroup analysis.[13] Even in ultra-small vessels implanted with a 2.0 mm Resolute Onyx™ stent, the results were satisfactory and further studies will be required to investigate the optimal treatment of ultra-small vessel diseases. Of the patients in the current study who experienced TLF within 1 year, a significant proportion were patients with ESRF receiving dialysis (five of 11 patients). Chronic kidney disease (CKD) is known to be associated with increased mortality and morbidity from coronary artery disease.[14] Particularly in those with ESRF on dialysis, who are frequently excluded from clinical trials, treatment failures are common after revascularization no matter whether percutaneous coronary intervention (PCI) or coronary bypass grafting are performed.[15] Even with second-generation DES, the 1-year TLF rate approached 15% and more than 30% needed repeat revascularization in longer term follow-up.[16-18] In this current registry, five out of 21 ESRF patients (24%) suffered from TLF within 1 year, which is quite a high proportion. With the aging population, increasing prevalence of T2DM, CKD and better renal support, the incidence of coronary artery disease in patients with ESRF is set to increase. Further advancements in coronary revascularization technology including PCI and stenting is imminently needed to address this future healthcare demand. This study had several limitations. First, a lack of a control group prevents comparison with other stents. Secondly, the small sample size and low event rates render conclusive outcome and independent predictor analyses difficult. Furthermore, the retrospective study design was subject to selection bias, although it allowed recruitment of a real-world PCI practice cohort and only two patients were lost to follow-up. In conclusion, the Resolute Onyx™ stent showed a favourable 1-year clinical performance in a real-world population.
  18 in total

1.  Acute and long-term outcomes of stenting in coronary vessel > 3.0 mm, 3.0-2.5 mm, and < 2.5 mm.

Authors:  I C Hsieh; C C Chien; H J Chang; M S Chern; K C Hung; F C Lin; D Wu
Journal:  Catheter Cardiovasc Interv       Date:  2001-07       Impact factor: 2.692

2.  Comparison of zotarolimus-eluting and everolimus-eluting coronary stents.

Authors:  Patrick W Serruys; Sigmund Silber; Scot Garg; Robert Jan van Geuns; Gert Richardt; Pawel E Buszman; Henning Kelbaek; Adrianus Johannes van Boven; Sjoerd H Hofma; Axel Linke; Volker Klauss; William Wijns; Carlos Macaya; Philippe Garot; Carlo DiMario; Ganesh Manoharan; Ran Kornowski; Thomas Ischinger; Antonio Bartorelli; Jacintha Ronden; Marco Bressers; Pierre Gobbens; Manuela Negoita; Frank van Leeuwen; Stephan Windecker
Journal:  N Engl J Med       Date:  2010-06-16       Impact factor: 91.245

3.  Performance of the resolute zotarolimus-eluting stent in small vessels.

Authors:  Ronald Caputo; Martin Leon; Patrick Serruys; Franz-Josef Neumann; Alan Yeung; Stephan Windecker; Jorge A Belardi; Sigmund Silber; Ian Meredith; Petr Widimský; Shigeru Saito; Laura Mauri
Journal:  Catheter Cardiovasc Interv       Date:  2014-04-07       Impact factor: 2.692

4.  Everolimus-eluting versus paclitaxel-eluting stents in coronary artery disease.

Authors:  Gregg W Stone; Ali Rizvi; William Newman; Kourosh Mastali; John C Wang; Ronald Caputo; Julie Doostzadeh; Sherry Cao; Charles A Simonton; Krishnankutty Sudhir; Alexandra J Lansky; Donald E Cutlip; Dean J Kereiakes
Journal:  N Engl J Med       Date:  2010-05-06       Impact factor: 91.245

5.  Sirolimus-eluting stents vs bare metal stents for coronary intervention in Japanese patients with renal failure on hemodialysis.

Authors:  Toru Aoyama; Hideki Ishii; Takanobu Toriyama; Hiroshi Takahashi; Hirotake Kasuga; Ryuichiro Murakami; Tetsuya Amano; Tadayuki Uetani; Yoshinari Yasuda; Yukio Yuzawa; Shoichi Maruyama; Seiichi Matsuo; Tatsuaki Matsubara; Toyoaki Murohara
Journal:  Circ J       Date:  2008-01       Impact factor: 2.993

6.  Revascularization in Patients With Multivessel Coronary Artery Disease and Chronic Kidney Disease: Everolimus-Eluting Stents Versus Coronary Artery Bypass Graft Surgery.

Authors:  Sripal Bangalore; Yu Guo; Zaza Samadashvili; Saul Blecker; Jinfeng Xu; Edward L Hannan
Journal:  J Am Coll Cardiol       Date:  2015-09-15       Impact factor: 24.094

7.  Impact of polymer hydrophilicity on biocompatibility: implication for DES polymer design.

Authors:  Ayala Hezi-Yamit; Carol Sullivan; Jennifer Wong; Laura David; Mingfei Chen; Peiwen Cheng; David Shumaker; Josiah N Wilcox; Kishore Udipi
Journal:  J Biomed Mater Res A       Date:  2009-07       Impact factor: 4.396

8.  Short- and long-term outcomes with drug-eluting and bare-metal coronary stents: a mixed-treatment comparison analysis of 117 762 patient-years of follow-up from randomized trials.

Authors:  Sripal Bangalore; Sunil Kumar; Mario Fusaro; Nicholas Amoroso; Michael J Attubato; Frederick Feit; Deepak L Bhatt; James Slater
Journal:  Circulation       Date:  2012-05-14       Impact factor: 29.690

9.  Long-term survival and repeat coronary revascularization in dialysis patients after surgical and percutaneous coronary revascularization with drug-eluting and bare metal stents in the United States.

Authors:  Gautam R Shroff; Craig A Solid; Charles A Herzog
Journal:  Circulation       Date:  2013-04-09       Impact factor: 29.690

10.  Development of a novel biocompatible polymer system for extended drug release in a next-generation drug-eluting stent.

Authors:  Kishore Udipi; Mingfei Chen; Peiwen Cheng; Kevin Jiang; Diane Judd; Alejandra Caceres; Robert J Melder; Josiah N Wilcox
Journal:  J Biomed Mater Res A       Date:  2008-06-15       Impact factor: 4.396

View more
  4 in total

Review 1.  Small vessel coronary artery disease: How small can we go with myocardial revascularization?

Authors:  Maciej T Wybraniec; Paweł Bańka; Tomasz Bochenek; Tomasz Roleder; Katarzyna Mizia-Stec
Journal:  Cardiol J       Date:  2020-09-28       Impact factor: 2.737

2.  Comparison of short-term clinical outcomes between Resolute Onyx zotarolimus-eluting stents and everolimus-eluting stent in patients with acute myocardial infarction: Results from the Korea Acute Myocardial infarction Registry (KAMIR).

Authors:  Yongcheol Kim; Sung Sik Oh; Myung Ho Jeong; Youngkeun Ahn; Ju Han Kim; Young Joon Hong; Doo Sun Sim; Min Chul Kim; Hyo-Soo Kim; Kyeong Ho Yun; Seok Kyu Oh; Chong Jin Kim; Myeong Chan Cho
Journal:  Cardiol J       Date:  2018-05-10       Impact factor: 2.737

3.  Double-vessel very late stent thrombosis following Resolute Onyx zotarolimus eluting stents implantation in an octogenarian.

Authors:  George Kassimis; Tushar Raina
Journal:  J Geriatr Cardiol       Date:  2018-10       Impact factor: 3.327

4.  One-Year Clinical Outcome of Inspiron Stent in All-Comers Population (Analysis from 790 Consecutive Patients).

Authors:  Felipe Falcão; Fabiano Cantarelli; Rodrigo Cantarelli; Flávio Mota; Manuela Navarro; Henrique Mota; Martinelly Santos; Daniel Cruz; André Sansônio; Marcelo Parente; Flávio Oliveira
Journal:  J Interv Cardiol       Date:  2020-12-17       Impact factor: 2.279

  4 in total