Design and evaluation of two multi-pinhole collimators for brain SPECT.

Ling Chen1, Benjamin M W Tsui2, Greta S P Mok3,4.   

Abstract

OBJECTIVE: SPECT is a powerful tool for diagnosing or staging brain diseases such as Alzheimer's disease (AD) and Parkinson's disease (PD) but is limited by its inferior resolution and sensitivity. At the same time, pinhole SPECT provides superior resolution and detection efficiency trade-off as compared to the conventional parallel-hole collimator for imaging small field-of-view (FOV), which fits for the case of brain imaging. In this study, we propose to develop and evaluate two multi-pinhole (MPH) collimator designs to improve the imaging of cerebral blood flow and striatum.
METHODS: We set the target resolutions to be 12 and 8 mm, respectively, and the FOV at 200 mm which is large enough to cover the whole brain. The constraints for system optimization include maximum and minimum detector-to-center-of-FOV (CFOV) distances of 344 and 294 mm, respectively, and minimal radius-of-rotation (ROR) of 135 mm to accommodate patients' shoulder. According to the targeted FOV, resolutions, and constraints, we determined the pinhole number, ROR, focal length, aperture acceptance angle, and aperture diameter which maximized the system sensitivity. We then assessed the imaging performance of the proposed MPH and standard low-energy high-resolution (LEHR) collimators using analytical simulations of a digital NCAT brain phantom with 99mTc-HMPAO/99mTc-TRODAT-1 distributions; Monte Carlo simulations of a hot-rod phantom; and a Defrise phantom using GATE v6.1. Projections were generated over 360° and reconstructed using the 3D MPH/LEHR OS-EM methods with up to 720 updates. The normalized mean square error (NMSE) was calculated over the cerebral and striatal regions extracted from the reconstructed images for 99mTc-HMPAO and 99mTc-TRODAT-1 simulations, respectively, and average normalized standard deviation (NSD) based on 20 noise realizations was assessed on selected uniform 3D regions as the noise index. Visual assessment and image profiles were applied to the results of Monte Carlo simulations.
RESULTS: The optimized design parameters of the MPH collimators were 9 pinholes with 4.7 and 2.8 mm pinhole diameter, 73° acceptance angle, 127 mm focal length, 167 mm ROR for 12 mm and 8 mm target resolution, respectively. According to the optimization results, the detection efficiencies of the proposed collimators were 270 and 40% more as compared to LEHR. The Monte Carlo simulations showed that 7.9 and 6.4 mm rods can be discriminated for the MPH collimators with target resolutions of 12 and 8 mm, respectively. The eight 12 mm-thick discs of the Defrise phantom can also be resolved clearly in the axial plane as demonstrated by the image profiles generated with the MPH collimators.
CONCLUSION: The two collimator designs provide superior image quality as compared to the conventional LEHR, and shows potential to improve current brain SPECT imaging based on a conventional SPECT scanner.

Entities:  

Keywords:  Alzheimer’s disease (AD); Brain SPECT; Multi-pinhole; Parkinson’s disease (PD)

Mesh:

Year:  2017        PMID: 28755084     DOI: 10.1007/s12149-017-1195-y

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  4 in total

1.  Investigation of Axial and Angular Sampling in Multi-Detector Pinhole-SPECT Brain Imaging.

Authors:  Navid Zeraatkar; Kesava S Kalluri; Benjamin Auer; Arda Konik; Timothy J Fromme; Lars R Furenlid; Phillip H Kuo; Michael A King
Journal:  IEEE Trans Med Imaging       Date:  2020-11-30       Impact factor: 10.048

2.  Cerebral SPECT imaging with different acquisition schemes using varying levels of multiplexing versus sensitivity in an adaptive multi-pinhole brain-dedicated scanner.

Authors:  Navid Zeraatkar; Kesava S Kalluri; Benjamin Auer; Micaehla May; R Garrett Richards; Lars R Furenlid; Phillip H Kuo; Michael A King
Journal:  Biomed Phys Eng Express       Date:  2021-09-22

Review 3.  Role of Nuclear Imaging to Understand the Neural Substrates of Brain Disorders in Laboratory Animals: Current Status and Future Prospects.

Authors:  Annunziata D'Elia; Sara Schiavi; Andrea Soluri; Roberto Massari; Alessandro Soluri; Viviana Trezza
Journal:  Front Behav Neurosci       Date:  2020-12-11       Impact factor: 3.558

4.  Design Study of an Ultrahigh Resolution Brain SPECT System Using a Synthetic Compound-Eye Camera Design With Micro-Slit and Micro-Ring Apertures.

Authors:  Elena Maria Zannoni; Can Yang; Ling-Jian Meng
Journal:  IEEE Trans Med Imaging       Date:  2021-11-30       Impact factor: 10.048

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.