Accuracy of Novel Computed Tomography-Guided Frameless Stereotactic Drilling and Catheter System in Human Cadavers.

Eric W Sankey1, Eric Butler1, John H Sampson2.   

Abstract

OBJECTIVE: To evaluate accuracy of a computed tomography (CT)-guided frameless stereotactic drilling and catheter system.
METHODS: A prospective, single-arm study was performed using human cadaver heads to evaluate placement accuracy of a novel, flexible intracranial catheter and stabilizing bone anchor system and drill kit. There were 20 catheter placements included in the analysis. The primary endpoint was accuracy of catheter tip location on intraoperative CT. Secondary endpoints included target registration error and entry and target point error before and after drilling. Measurements are reported as mean ± SD (median, range).
RESULTS: Target registration error was 0.46 mm ± 0.26 (0.50 mm, -1.00 to 1.00 mm). Two (10%) target point trajectories were negatively impacted by drilling. Intracranial catheter depth was 59.8 mm ± 9.4 (60.5 mm, 38.0-80.0 mm). Drilling angle was 22° ± 9 (21°, 7°-45°). Deviation between planned and actual entry point on CT was 1.04 mm ± 0.38 (1.00 mm, 0.40-2.00 mm). Deviation between planned and actual target point on CT was 1.60 mm ± 0.98 (1.40 mm, 0.40-4.00 mm). No correlation was observed between intracranial catheter depth and target point deviation (accuracy) (Pearson coefficient 0.018) or between technician experience and accuracy (Pearson coefficient 0.020). There was no significant difference in accuracy with trajectories performed for different cadaver heads (P = 0.362).
CONCLUSIONS: Highly accurate catheter placement is achievable using this novel flexible catheter and bone anchor system placed via frameless stereotaxy, with an average deviation between planned and actual target point of 1.60 mm ± 0.98 (1.40 mm, 0.40-4.00 mm).
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bone anchor; Catheter; Frameless stereotactic drilling; Image-guided neurosurgery; Placement accuracy; Target registration error

Mesh:

Year:  2017        PMID: 28754645     DOI: 10.1016/j.wneu.2017.07.098

Source DB:  PubMed          Journal:  World Neurosurg        ISSN: 1878-8750            Impact factor:   2.104


  1 in total

1.  A comparative study on the efficacy of robot of stereotactic assistant and frame-assisted stereotactic drilling, drainage for intracerebral hematoma in patients with hypertensive intracerebral hemorrhage.

Authors:  Liang Liang; Xin Li; Haiqing Dong; Xin Gong; Guanpeng Wang
Journal:  Pak J Med Sci       Date:  2022 Sep-Oct       Impact factor: 2.340

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.