Effect of cleaning status on accuracy and precision of oxygen flowmeters of various ages.

Stephanie Fissekis1, David S Hodgson2, Nora M Bello3.   

Abstract

OBJECTIVE: To evaluate oxygen flowmeters for accuracy and precision, assess the effects of cleaning and assess conformity to the American Society for Testing Materials (ASTM) standards. STUDY
DESIGN: Experimental study.
METHODS: The flow of oxygen flowmeters from 31 anesthesia machines aged 1-45 years was measured before and after cleaning using a volumetric flow analyzer set at 0.5, 1.0, 2.0, 3.0, and 4.0 L minute-1. A general linear mixed models approach was used to assess flow accuracy and precision.
RESULTS: Flowmeters 1 year of age delivered accurate mean oxygen flows at all settings regardless of cleaning status. Flowmeters ≥5 years of age underdelivered at flows of 3.0 and 4.0 L minute-1. Flowmeters ≥12 years underdelivered at flows of 2.0, 3.0 and 4.0 L minute-1 prior to cleaning. There was no evidence of any beneficial effect of cleaning on accuracy of flowmeters 5-12 years of age (p > 0.22), but the accuracy of flowmeters ≥15 years of age was improved by cleaning (p < 0.05). Regardless of age, cleaning increased precision, decreasing flow variability by approximately 17%. Nine of 31 uncleaned flowmeters did not meet ASTM standards. After cleaning, a different set of nine flowmeters did not meet standards, including three that had met standards prior to cleaning.
CONCLUSIONS: Older flowmeters were more likely to underdeliver oxygen, especially at higher flows. Regardless of age, cleaning decreased flow variability, improving precision. However, flowmeters still may fail to meet ASTM standards, regardless of cleaning status. CLINICAL RELEVANCE: Cleaning anesthesia machine oxygen flowmeters improved precision for all tested machines and partially corrected inaccuracies in flowmeters ≥15 years old. A notable proportion of flowmeters did not meet ASTM standards. Cleaning did not ensure that they subsequently conformed to ASTM standards. We recommend annual flow output validation to identify whether flowmeters are acceptable for continued clinical use.
Copyright © 2017 Association of Veterinary Anaesthetists and American College of Veterinary Anesthesia and Analgesia. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  anesthesia; durable medical equipment; equipment maintenance; oxygen flowmeters

Mesh:

Substances:

Year:  2017        PMID: 28751133     DOI: 10.1016/j.vaa.2016.10.002

Source DB:  PubMed          Journal:  Vet Anaesth Analg        ISSN: 1467-2987            Impact factor:   1.648


  1 in total

1.  Novel strategy to decrease the fraction of inspired oxygen (FiO2) of an anesthetic circuit using a commercially available fish tank air pump: An in vitro study.

Authors:  Allison Goldberg; Emily McCobb; Elizabeth Rozanski
Journal:  Can Vet J       Date:  2021-03       Impact factor: 1.008

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.