Inter-rater reliability and double reading analysis of an automated three-dimensional breast ultrasound system: comparison of two independent examiners.

Anna Maier1, Joerg Heil1, Anna Lauer1, Aba Harcos1, Benedikt Schaefgen1, Alexandra von Au1, Julia Spratte1, Fabian Riedel1, Geraldine Rauch2,3, André Hennigs1, Christoph Domschke1, Sarah Schott1, Joachim Rom1, Florian Schuetz1, Christof Sohn1, Michael Golatta4.   

Abstract

PURPOSE: Breast ultrasound could be a valuable tool complementary to mammography in breast cancer screening. Automated 3D breast ultrasound (ABUS) addresses challenges of hand-held ultrasound and could allow double reading analysis of ultrasound images. This trial assesses the inter-rater reliability and double reading analysis of an ABUS system.
METHODS: To assess the reproducibility and diagnostic validity of the ABUS system, SomoV™, a blinded double reading analysis, was performed in 1019 patients (2038 breasts) by two examiners (examiner A/B) and compared to single reading results, as well as to the reference standard regarding its diagnostic validity. Cohen's kappa coefficients were calculated to measure the inter-rater reliability and agreement of the different diagnostic modalities. Patient comfort and time consumption for image acquisition and reading were analyzed descriptively as secondary objectives.
RESULTS: Analysis of inter-rater reliability yielded agreement in 81.6% (κ = 0.37; p < 0.0001) showing fair agreement. Single reading analysis of SomoV™ exams (examiner A/examiner B) compared to reference standard showed good specificity (examiner A: 88.3%/examiner B: 84.5%), fair inter-rater agreement (examiner A: κ = 0.31/examiner B: κ = 0.31), and adequate sensitivity (examiner A: 53.1%/examiner B: 64.2%). Double reading analysis yielded good sensitivity and specificity (73.7 and 77.7%). Mammography (n = 1911) alone detected 160 of 176 carcinomas (sensitivity 90.1%). Adding SomoV™ to mammography would have detected 12 additional carcinomas, resulting in a higher sensitivity of 97.7%.
CONCLUSION: SomoV™ is a promising technique with good sensitivity, high patient comfort, and fair inter-examiner reliability. It allows double reading analysis that, in combination with mammography, could increase detection rates in breast cancer screening.

Entities:  

Keywords:  Automated 3D breast ultrasound (ABUS); Breast cancer; Double reading analysis; Hand-held ultrasound; Inter-rater reliability

Mesh:

Year:  2017        PMID: 28748340     DOI: 10.1007/s00404-017-4473-y

Source DB:  PubMed          Journal:  Arch Gynecol Obstet        ISSN: 0932-0067            Impact factor:   2.344


  2 in total

1.  Deep learning applied to breast imaging classification and segmentation with human expert intervention.

Authors:  Rory Wilding; Vivek M Sheraton; Lysabella Soto; Niketa Chotai; Ern Yu Tan
Journal:  J Ultrasound       Date:  2022-01-09

2.  The value of coronal view as a stand-alone assessment in women undergoing automated breast ultrasound.

Authors:  Simone Schiaffino; Licia Gristina; Simona Tosto; Elena Massone; Sara De Giorgis; Alessandro Garlaschi; Alberto Tagliafico; Massimo Calabrese
Journal:  Radiol Med       Date:  2020-07-16       Impact factor: 3.469

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.