Literature DB >> 28742650

Reliable and Easy-To-Use Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Analysis of Fluconazole, Isavuconazole, Itraconazole, Hydroxy-Itraconazole, Posaconazole, and Voriconazole in Human Plasma and Serum.

Carsten Müller1, David Gehlen, Cornelia Blaich, Domenik Prozeller, Blasius Liss, Thomas Streichert, Martin H J Wiesen.   

Abstract

BACKGROUND: A fast and easy-to-use liquid chromatography-tandem mass spectrometry method for the determination and quantification of 6 triazoles [fluconazole (FLZ), isavuconazole (ISZ), itraconazole (ITZ), hydroxy-itraconazole (OH-ITZ), posaconazole (PSZ), and voriconazole (VRZ)] in human plasma and serum was developed and validated for therapeutic drug monitoring.
METHODS: Sample preparation was based on protein precipitation with acetonitrile and subsequent centrifugation. Isotope-labeled analogues for each analyte were used as internal standards. Chromatographic separation was achieved using a 50 × 2.1 mm, 1.9 μm polar Hypersil Gold C18 column and mobile phase consisting of 0.1% formic acid/acetonitrile (45%/55%, vol/vol) at a flow rate of 340 μL/min. The triazoles were simultaneously detected using a triple-stage quadrupole mass spectrometer operated in selected reaction monitoring mode with positive heated electrospray ionization within a single runtime of t = 3.00 minutes.
RESULTS: Linearity of all azole concentration ranges was verified by the Mandel test and demonstrated for all azoles. All calibration curves were linear and fitted using least squares regression with a weighting factor of the reciprocal concentration. Limits of detection (μg/L/L) were FLZ, 9.3; ISZ, 0.3; ITZ, 0.6; OH-ITZ, 8.6; PSZ, 3.4; and VRZ, 2.1. The lower limits of quantitation (μg/L/liter) were FLZ, 28.3; ISZ, 1.0; ITZ, 1.7; OH-ITZ, 26.2; PSZ, 10.3; and VRZ, 6.3. Intraday and interday precisions ranged from 0.6% to 6.6% for all azoles. Intraday and interday accuracies (%bias) of all analytes were within 10.5%. In addition, we report on a 29-year-old white woman (94 kg body weight) with a history of acute myeloid leukemia who underwent stem cell transplantation. Because of diagnosis of aspergillus pneumonia, antifungal pharmacotherapy was initiated with different application modes and dosages of ISZ, and plasma concentrations were monitored over a time period of 6 months.
CONCLUSIONS: A precise and highly sensitive liquid chromatography-tandem mass spectrometry method was developed that enables quantification of triazoles in plasma and serum matrix across therapeutically relevant concentration ranges. It was successfully implemented in our therapeutic drug monitoring routine service and is suitable for routine monitoring of antifungal therapy and in severely ill patients.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28742650     DOI: 10.1097/FTD.0000000000000438

Source DB:  PubMed          Journal:  Ther Drug Monit        ISSN: 0163-4356            Impact factor:   3.681


  3 in total

1.  Simultaneous quantitation of five triazole anti-fungal agents by paper spray-mass spectrometry.

Authors:  Christine L Skaggs; Greta J Ren; El Taher M Elgierari; Lillian R Sturmer; Run Z Shi; Nicholas E Manicke; Lindsey M Kirkpatrick
Journal:  Clin Chem Lab Med       Date:  2020-04-28       Impact factor: 3.694

2.  A sensitive method for analyzing fluconazole in extremely small volumes of neonatal serum.

Authors:  Jumpei Saito; Ayano Tanzawa; Yuka Kojo; Hidehiko Maruyama; Tetsuya Isayama; Kensuke Shoji; Yushi Ito; Akimasa Yamatani
Journal:  J Pharm Health Care Sci       Date:  2020-07-01

3.  Interlaboratory Analysis of Isavuconazole Plasma Concentration Assays Among European Laboratories.

Authors:  Federico Pea; Robert Krause; Carsten Müller; Benjamin Hennart; Malcolm Richardson; Andreas Meinitzer; Martin H J Wiesen; Tatiana Wiktorowicz; Jochen Spickermann; Anne S Henriksen
Journal:  Ther Drug Monit       Date:  2019-10       Impact factor: 3.681

  3 in total