Abstract

Clavicle fracture reported incidence is about 5% of fractures in adult; among them, those located in the middle third of the shaft represent more than 80% from the total of cases. Due to the special morphological and biomechanical constraints of the clavicle, several methods for restoring morphological integrity in these fractures are described, including conservative, non-surgical treatment. The last 10 years of clinical studies in the field have favored the surgical treatment for selected cases; several osteosynthesis implants are in use - mostly anatomical plates with specific advantages and documented complications. A failed anatomical clavicle plate was explanted and analyzed after a protocol using stereomicroscopy, scanning electron microscopy and energy dispersive spectrometry. Based on the computed tomography (CT) scan determination of patient morphological parameters, a finite elements analysis of the failure scenario was completed. The failure analysis has proved that the plate breakage had occurred in the point of maximal elastic stress and minor deformation. The clinical implication is that no hole should remain free of screw during clavicle plate fixation and the implant should be chosen based on patient morphological parameters. In comminuted clavicle fracture, anatomic bridging with locked plate technique may lead to implant failure due to increase of the stress in the midshaft area. Thorough knowledge of anatomy and morphology of complex bones like the clavicle is necessary. Modern osteosynthesis anatomical implants are still to be improved.

Entities:  

Mesh:

Year:  2017        PMID: 28730248

Source DB:  PubMed          Journal:  Rom J Morphol Embryol        ISSN: 1220-0522            Impact factor:   1.033


  4 in total

1.  Bone Cements Used for Hip Prosthesis Fixation: The Influence of the Handling Procedures on Functional Properties Observed during In Vitro Study.

Authors:  Alina Robu; Robert Ciocoiu; Aurora Antoniac; Iulian Antoniac; Anca Daniela Raiciu; Horatiu Dura; Norin Forna; Mihai Bogdan Cristea; Ioana Dana Carstoc
Journal:  Materials (Basel)       Date:  2022-04-19       Impact factor: 3.748

2.  Rare Breast Carcinoma with Paradoxical Plasma Cell Immunoprofile: A Case Report.

Authors:  Eugen Sebastian Grădinaru; Mihai-Ciprian Stoicea; Liliana Mocanu; Iulian Antoniac; Daniela Gheorghiță; Alina Gabriela Mihaela Grigore
Journal:  Medicina (Kaunas)       Date:  2020-02-05       Impact factor: 2.430

3.  Clavicle nonunion and plate breakage after locking compression plate fixation of displaced midshaft clavicular fractures.

Authors:  Xiaoyan Huang; Haijun Xiao; Feng Xue
Journal:  Exp Ther Med       Date:  2019-11-19       Impact factor: 2.447

4.  Biomechanical effects of different numbers and locations of screw-in clavicle hook plates.

Authors:  Cheng-Chi Wang; Cheng-Hung Lee; Kun-Hui Chen; Chien-Chou Pan; Ming-Tzu Tsai; Kuo-Chih Su
Journal:  Front Bioeng Biotechnol       Date:  2022-09-09
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.