| Literature DB >> 28729905 |
Takahiro Hirayama1,2, Nobuyuki Nosaka3,4, Yasumasa Okawa5, Soichiro Ushio5, Yoshihisa Kitamura5, Toshiaki Sendo5, Toyomu Ugawa2, Atsunori Nakao2.
Abstract
BACKGROUND: In Japan, nafamostat mesylate (NM) is frequently used as an anticoagulant during continuous renal replacement therapy (CRRT). The dialyzer membrane AN69ST has been reported to adsorb NM and affect the management of anticoagulant therapy. However, the adsorbed amount has not yet been quantitatively assessed. Therefore, in this study, we evaluated the pre- and post-hemofilter prolongation of the activated clotting time (ACT) in patients with AN69ST and PS membranes. We also measured the adsorption of NM in three types of CRRT membranes using an experimental model.Entities:
Keywords: AN69ST; Adsorption; Anticoagulant therapy; Continuous renal replacement therapy; Nafamostat mesylate
Year: 2017 PMID: 28729905 PMCID: PMC5516335 DOI: 10.1186/s40560-017-0244-x
Source DB: PubMed Journal: J Intensive Care ISSN: 2052-0492
Fig. 1Activating clotting time (ACT) sampling sites
Characteristics of the three hemofilters
| Dialysis membrane | Material | Structure | Surface area (m2) | Inner diameter (μm) | Wall thickness (μm) | Priming volume (mL) |
|---|---|---|---|---|---|---|
| sepXiris100 | Polyacrylonitrile (surface treated) | Symmetric | 1.0 | 240 | 50 | 69 |
| Hemofeel SHG-1.0 | Polysulfone | Asymmetric | 1.0 | 200 | 40 | 67 |
| Hemofeel CH-1.0N | Polymethylmethacrylate | Symmetric | 1.0 | 200 | 30 | 58 |
Fig. 2Diagram of in vitro experimental setup (single pass). The heights of all sites of measurement of the inlet pressure, filtration pressure, and return pressure were aligned with the heights of the sites of connection between the dialyzer and the patient. Adjustments were made to maintain a fixed/constant pressure inside the circuit. During sampling, caution was taken not to apply negative pressure. Extracorporeal ultrafiltration method (ECUM) model
Clearance formulas
| Solution clearance (mL/min) | CLs = (CBi − CBo)/CBi × (QB − QF) + QF |
| Filtrate clearance (mL/min) | CLf = CF/CBi × QF |
| Adsorption clearance (mL/min) | CLad = CLs − CLf |
| Sieving coefficient | SC = 2CF/(CBi + CBo) |
C inlet (pre-hemofilter) solute concentration, C outlet (post-hemofilter) solute concentration, CF filtrate solute concentration, QB blood flow, QF filtration flow
Patient summary
| Dialysis membrane | AN69ST | PS |
|
|---|---|---|---|
| Age (years) | 69.8 ± 11.4 | 67.4 ± 19.6 | N.S. |
| Sex (male:female) | 13:6 | 6:5 | N.S. |
| Body weight | 58.2 ± 14.6 | 62.3 ± 16.9 | N.S. |
| SOFA | 11.1 ± 4.0 | 9.5 ± 4.4 | N.S. |
| APACHE II | 27.6 ± 10.0 | 26.9 ± 10.4 | N.S. |
| Mortality | 6/19 (32%) | 3/11 (27%) | N.S. |
| Sepsis | 13/19 (68%) | 3/11 (27%) | 0.03 |
| QB (mL/min) | 95.9 ± 8.9 | 95.1 ± 15.4 | N.S. |
| QD (mL/h) | 934.8 ± 329.6 | 1422.4 ± 602.5 | <0.01 |
| QS (mL/h) | 517.7 ± 152.5 | 602.6 ± 312.8 | N.S. |
| QF (mL/h) | 1471.7 ± 414.7 | 1852.4 ± 662.2 | <0.01 |
Data are the mean and standard deviation or number. For intergroup testing, the Mann-Whitney U test or the χ 2 test was performed at suitable locations. AN69ST polyacrylonitrile (surface treated), PS polysulfone, N.S. not significant, SOFA Sequential Organ Failure Assessment, APACHE II Acute Physiology and Chronic Health Evaluation II score, QB blood pump flow rate, QD dialysate pump flow rate, QF filtration pump flow rate, QS substitution fluid pump flow rate
Fig. 3Comparison of ACT difference. a Pre- and post-filter ACT for the AN69ST (n = 122) and PS (n = 37) hemofilters. b Pre- and post-hemofilter ACT difference. Data represent the mean (SD) of four independent experiments. Asterisk indicates p < 0.01 by Mann-Whitney U test
Fig. 4Results of the in vitro assay (single pass). AN69ST polyacrylonitrile (surface treated), PS polysulfone, PMMA polymethylmethacrylate, conc. concentration. Study of the adsorbed/filtered amounts of NM (ECUM). Data represent the mean (SD) of four independent experiments. Asterisk indicates Tukey’s post-tests that were used to determine the significant differences between groups
Fig. 5Nafamostat mesylate (NM) clearance. Data represent the mean (SD of total clearance) of four independent experiments. Asterisk indicates Tukey’s post-tests that were used to determine differences between the groups. The theoretical maximum value of the filtration clearance is shown as a dotted line, and the theoretical maximum value of the adsorption clearance is shown as a gray line. AN69ST polyacrylonitrile (surface treated), PS polysulfone, PMMA polymethylmethacrylate