Mateusz M Plucinski1,2, Eric Rogier2, Pedro Rafael Dimbu3, Filomeno Fortes3, Eric S Halsey1,2, Michael Aidoo2.
Abstract
Most malaria testing is by rapid diagnostic tests (RDTs) that detect Plasmodium falciparum histidine-rich protein 2 (HRP2). Recently, several RDT manufacturers have developed highly sensitive RDTs (hsRDTs), promising a limit of detection (LOD) orders of magnitude lower than conventional RDTs. To model the added utility of hsRDTs, HRP2 concentration in Angolan outpatients was measured quantitatively using an ultrasensitive bead-based assay. The distribution of HRP2 concentration was bimodal in both afebrile and febrile patients. The conventional RDT was able to detect 81% of all HRP2-positive febrile patients and 52-77% of HRP2-positive afebrile patients. The added utility of hsRDTs was estimated to be greater in afebrile patients, where an hsRDT with a LOD of 200 pg/mL would detect an additional 50-60% of HRP2-positive persons compared with a conventional RDT with a LOD of 3,000 pg/mL. In febrile patients, the hsRDT would detect an additional 10-20% of cases. Conventional RDTs already capture the vast majority of symptomatic HRP2-positive individuals, and hsRDTs would have to reach a sufficiently low LOD approaching 200 pg/mL to provide added utility in identifying HRP2-positive, asymptomatic individuals.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28722629 PMCID: PMC5637620 DOI: 10.4269/ajtmh.17-0262