Technical Note: Evaluation of plastic scintillator detector for small field stereotactic patient-specific quality assurance.

Yujiao Qin1, Stephen J Gardner1, Joshua Kim1, Yimei Huang1, Ning Wen1, Anthony Doemer1, Indrin J Chetty1.   

Abstract

PURPOSE: To evaluate the performance of a commercial plastic scintillator detector (PSD) for small-field stereotactic patient-specific quality assurance (QA) measurements using flattening-filter-free beam.
METHODS: A total of 10 spherical targets [volume range: (0.03 cc-2 cc)] were planned with two techniques: (a) dynamic conformal arc (DCA-10 plans) and (b) volumetric modulated arc therapy (VMAT-10 plans). All plans were generated using Varian Eclipse treatment planning system, and AcurosXB v.13 algorithm in 1.0 mm grid size. Additionally, 14 previously treated cranial and spine SRS plans were evaluated [6 DCA, 8 VMAT, volume range: (0.04 cc-119.02 cc)]. Plan modulation was quantified via two metrics: MU per prescription dose (MU/Rx) and Average Leaf Pair Opening (ALPO). QA was performed on the Varian Edge linear accelerator equipped with HDMLC. Three detectors were used: (a) PinPoint ion chamber (PTW; active volume 0.015 cc), (b) Exradin W1 PSD (Standard Imaging; active volume 0.002 cc), and (c) Gafchromic EBT3 film (Ashland). PinPoint chamber and PSD were positioned perpendicular to beam axis in a Lucy phantom (Standard Imaging); films were placed horizontally capturing the coronal plane.
RESULTS: PSD, film, and PinPoint chamber measured average differences of 1.00 ± 1.54%, 1.30 ± 1.69%, and -0.66 ± 2.36%, respectively, compared to AcurosXB dose calculation. As the target volume decreased, PinPoint chamber measured lower doses (maximum -5.07% at 0.07 cc target), while PSD and film measured higher doses (2.87% and 2.54% at 0.03 cc target) than AcurosXB. Film agreed with the benchmark detector PSD by an average difference of 0.31 ± 1.20%, but suffered from larger uncertainty; PinPoint chamber underestimated dose by more than 4% for targets smaller than 0.2 cc. Taking PSD as the measurement standard, DCA plans achieved good QA results across all volumes studied, with an average of -0.07 ± 0.89%; for VMAT plans, PSD measured consistently higher dose (1.95 ± 1.36%) than AcurosXB. Correlation study revealed that plan modulation quantified by both MU/Rx and ALPO correlated significantly with QA results.
CONCLUSION: Among all three detectors, PSD demonstrated superior performances in plans with small fields and heavy modulation. High consistency and low uncertainty made PSD a suitable detector for clinical routine SRS QA. PinPoint chamber should be avoided for targets smaller than 0.2 cc; film dosimetry can be utilized with careful evaluation of its uncertainty bracket. Compared to PSD measurements, AcurosXB calculation demonstrated high accuracy for nonmodulated small fields. The positive correlation between plan modulation and QA discrepancy calls for our attention for clinical SRS plans with high modulation.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  quality assurance; scintillator detector; small field; stereotactic radiosurgery

Mesh:

Substances:

Year:  2017        PMID: 28714067     DOI: 10.1002/mp.12471

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Beam commissioning of the first clinical biology-guided radiotherapy system.

Authors:  Bin Han; Dante Capaldi; Nataliya Kovalchuk; Eric Simiele; John White; Daniel Zaks; Lei Xing; Murat Surucu
Journal:  J Appl Clin Med Phys       Date:  2022-04-28       Impact factor: 2.243

2.  Use of a plastic scintillator detector for patient-specific quality assurance of VMAT SRS.

Authors:  Jesse D Snyder; Rodney J Sullivan; Xingen Wu; Elizabeth L Covington; Richard A Popple
Journal:  J Appl Clin Med Phys       Date:  2019-09       Impact factor: 2.102

3.  Quality Assurance for Small-Field VMAT SRS and Conventional-Field IMRT Using the Exradin W1 Scintillator.

Authors:  Zike Huang; Jian Qiao; Cui Yang; Ming Liu; Jiazhou Wang; Xu Han; Weigang Hu
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec