Split-VMAT technique to control the expiratory breath-hold time in liver stereotactic body radiation therapy.

Yen Hwa Lin1, Shuichi Ozawa2, Hideharu Miura3, Katsunori Yogo3, Takeo Nakashima4, Kentaro Miki5, Shintaro Tsuda4, Yusuke Ochi4, Daisuke Kawahara6, Tomoki Kimura5, Akito Saito5, Yasushi Nagata7.   

Abstract

PURPOSE: In this study, we demonstrate the feasibility of using split-arcs in volumetric modulated arc therapy (VMAT), tailored for expiratory breath-hold in stereotactic body radiation therapy (SBRT) for liver tumors. We compare it with three-dimensional conformal radiation therapy (3D-CRT) and continuous-VMAT, for ten randomly selected hepatocellular carcinoma cases.
METHODS: Four coplanar and four non-coplanar beams were used for the 3D-CRT plans. A pair of partial arcs, chosen using a back-and-forth rotating motion, were used for the continuous-VMAT plans. Split-VMAT plans were created using the same arc range as the continuous-VMAT plans, but were split into smaller arcs (<90°), to simulate an expiratory breath hold of <15s. The dose distribution, treatment delivery efficiency, and patient specific quality assurance of the split-VMAT, were verified to ensure that the outcomes were equal, or better than, those for 3D-CRT and continuous-VMAT. The prescription was 48Gy/4 fractions, to 95% of the PTV, using 10MV FFF X-ray beams.
RESULTS: The mean dose of the liver-GTV was lower in the split-VMAT compared with that of 3D-CRT. Split-VMAT was more conformal compared with 3D-CRT. The total treatment time for split-VMAT was shorter than that of 3D-CRT. Similar dosimetric indices were observed for split-VMAT and continuous-VMAT. All VMAT plans passed the gamma acceptance test.
CONCLUSIONS: Split-VMAT designed to accommodate an expiratory breath-hold period of 15s is a feasible and efficient use of liver SBRT, because it does not compromise the quality of the plan, when compared with 3D-CRT or continuous-VMAT.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Expiratory breath-hold; Liver stereotactic body radiation therapy; Split-VMAT; Volumetric modulated arc therapy

Mesh:

Year:  2017        PMID: 28712715     DOI: 10.1016/j.ejmp.2017.06.022

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  Positional repeatability and variation in internal and external markers during volumetric-modulated arc therapy under end-exhalation breath-hold conditions for pancreatic cancer patients.

Authors:  Makoto Sasaki; Mitsuhiro Nakamura; Tomohiro Ono; Ryo Ashida; Michio Yoshimura; Manabu Nakata; Takashi Mizowaki; Naozo Sugimoto
Journal:  J Radiat Res       Date:  2020-09-08       Impact factor: 2.724

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.