Misfit of Three Different Implant-Abutment Connections Before and After Cyclic Load Application: An In Vitro Study.

Sergio Alexandre Gehrke, Rafael Arcesio Delgado-Ruiz, Juan Carlos Prados Frutos, María Prados-Privado, Berenice Anina Dedavid, Jose Manuel Granero Marín, José Luiz Calvo Guirado.   

Abstract

PURPOSE: This study aimed to evaluate the misfit of three different implant-abutment connections before and after cycling load.
MATERIALS AND METHODS: One hundred twenty dental implants and correspondent prefabricated titanium abutments were used. Three different implant-abutment connections were evaluated: Morse taper (MT group), external hexagon (EH group), and internal hexagon (IH group). Forty implants and 40 abutments were used per group. The parameters for the mechanical evaluation were set as: 360,000 cycles, load of 150 N, and frequency of 4 Hz. Samples were sectioned in their longitudinal and transversal axes, and the misfit of the implant-abutment connection was evaluated by scanning electron microscopy analysis. One-way analyses of variance, Tukey post hoc analyses (α = .05), and t test (P < .05) were used to determine differences between groups.
RESULTS: At the longitudinal direction, all the groups showed the presence of microgaps before cycling load; after cycling load, microgaps were reduced in all groups (P > .05). Transversally, only the MT group showed full fitting after cycling load compared with the other groups (EH and IH) (P < .0001).
CONCLUSION: The application of cycling load produces an accommodation of the implant-abutment connection in internal, external, and Morse taper connections. In the longitudinal direction, the accommodation decreases and/or eliminates the gap observed initially (before load). In the horizontal direction, Morse cone implant-abutment connections experience a complete accommodation with the elimination of the gap.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28708914     DOI: 10.11607/jomi.5629

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  5 in total

Review 1.  Implant-abutment connections on single crowns: a systematic review.

Authors:  F M Ceruso; P Barnaba; S Mazzoleni; L Ottria; M Gargari; A Zuccon; G Bruno; A DI Fiore
Journal:  Oral Implantol (Rome)       Date:  2017-01-21

2.  Microgap and microleakage of a hybrid connection platform-switched implant system in the absence or presence of a silicone-based sealing agent.

Authors:  Pedro Ferrás Fernandes; Liliana Grenho; Maria Helena Fernandes; João Carlos Sampaio-Fernandes; Pedro Sousa Gomes
Journal:  Odontology       Date:  2021-08-27       Impact factor: 2.634

3.  A Comparative 3D Finite Element Computational Study of Three Connections.

Authors:  Davide Farronato; Mattia Manfredini; Andrea Stevanello; Veronica Campana; Lorenzo Azzi; Marco Farronato
Journal:  Materials (Basel)       Date:  2019-09-26       Impact factor: 3.623

4.  A Comparative Analysis of Implants Presenting Different Diameters: Extra-Narrow, Narrow and Conventional.

Authors:  Henrique Tuzzolo Neto; Alessandra Sayuri Tuzita; Sérgio Alexandre Gehrke; Renata de Vasconcellos Moura; Márcio Zaffalon Casati; Alfredo Mikail Melo Mesquita
Journal:  Materials (Basel)       Date:  2020-04-17       Impact factor: 3.623

5.  Bacterial infiltration and detorque at the implant abutment morse taper interface after masticatory simulation.

Authors:  Ana Paula Granja Scarabel Nogueira Bella; Alessandra Sayuri Tuzita; Ivana Barbosa Suffredini; Alberto Noriyuki Kojima; Elcio Magdalena Giovani; Alfredo Mikail Melo Mesquita
Journal:  Sci Rep       Date:  2022-10-12       Impact factor: 4.996

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.