Literature DB >> 28700083

Torsional resistance of XP-endo Shaper at body temperature compared with several nickel-titanium rotary instruments.

A M Elnaghy1, S E Elsaka2.   

Abstract

AIM: To compare the torsional resistance of XP-endo Shaper (XPS; size 30, .01 taper, FKG Dentaire, La Chaux-de-Fonds, Switzerland) instruments at body temperature with TRUShape (TRS; size 30, .06 taper, Dentsply Tulsa Dental Specialties, Tulsa, OK, USA), ProFile Vortex (PV; size 30, .04 taper, Dentsply Tulsa Dental Specialties) and FlexMaster (FM; size 30, .04 taper, VDW GmbH, Munich, Germany) nickel-titanium rotary instruments.
METHODOLOGY: A metal block with a square-shaped mould (5 mm × 5 mm × 5 mm) was positioned inside a glass container. Five millimetres of the tip of each instrument was held inside the metal block by filling the mould with a resin composite. The instruments were tested for torsional resistance in saline solution at 37 °C. Data were analysed using one-way analysis of variance (anova) and Tukey post hoc tests. The significance level was set at P < 0.05.
RESULTS: FM had the greatest torsional resistance amongst the instruments tested (P < 0.001). There was no significant difference between FM and PV instruments (P = 0.211). The ranking for torsional resistance values was: FM > PV > TRS > XPS.
CONCLUSIONS: FlexMaster and ProFile Vortex instruments were more resistant to torsional stress compared with TRUShape and XP-endo Shaper instruments. The manufacturing process used to produce XP-endo Shaper instruments did not enhance their resistance to torsional stress as compared with the other instruments.
© 2017 International Endodontic Journal. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  FlexMaster; ProFile Vortex; TRUShape; XP-endo Shaper; body temperature; torsional resistance

Mesh:

Substances:

Year:  2017        PMID: 28700083     DOI: 10.1111/iej.12815

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  7 in total

1.  Postendodontic pain in asymptomatic necrotic teeth prepared with different rotary instrumentation techniques.

Authors:  Talal Al-Nahlawi; Ahmad Alabdullah; Alaa Othman; Raghad Sukkar; Mazen Doumani
Journal:  J Family Med Prim Care       Date:  2020-07-30

2.  Comparison of antibacterial effectiveness of three rotary file system with different geometry in infected root canals before and after instrumentation-a double-blinded randomized controlled clinical trial.

Authors:  Riluwan Siddique; Malli Sureshbabu Nivedhitha; Manish Ranjan; Benoy Jacob; Pradeep Solete
Journal:  BDJ Open       Date:  2020-06-08

Review 3.  Nickel-Titanium Rotary File Systems: What's New?

Authors:  Sadia Tabassum; Kamil Zafar; Fahad Umer
Journal:  Eur Endod J       Date:  2019-10-18

4.  Evidence for reduced torsional resistance of rotary files under curved position.

Authors:  Ahmed Jamleh; Raghad Almedlej; Rasha Alomar; Nouf Almayouf; Abdulmohsen Alfadley; Khalid Alfouzan
Journal:  Saudi Dent J       Date:  2020-08-06

5.  Advancing Nitinol: From heat treatment to surface functionalization for nickel-titanium (NiTi) instruments in endodontics.

Authors:  Wai-Sze Chan; Karan Gulati; Ove A Peters
Journal:  Bioact Mater       Date:  2022-09-27

6.  Influence of NiTi Wire Diameter on Cyclic and Torsional Fatigue Resistance of Different Heat-Treated Endodontic Instruments.

Authors:  Eugenio Pedullà; Francesco Saverio Canova; Giusy Rita Maria La Rosa; Alfred Naaman; Franck Diemer; Luigi Generali; Walid Nehme
Journal:  Materials (Basel)       Date:  2022-09-22       Impact factor: 3.748

7.  Improved dentin disinfection by combining different-geometry rotary nickel-titanium files in preparing root canals.

Authors:  Marwa M Bedier; Ahmed Abdel Rahman Hashem; Yosra M Hassan
Journal:  Restor Dent Endod       Date:  2018-11-01
  7 in total