Lei Ouyang1,2, Zuyan Jiang3, Nan Wang4, Lihua Zhu5, Heqing Tang6.
Abstract
Sibutramine hydrochloride (Entities:
Keywords: Surface Enhanced Raman Scattering; polyvinyl alcohol; sibutramine hydrochloride; β-cyclodextrin
Mesh:
Substances:
Year: 2017 PMID: 28698502 PMCID: PMC5539472 DOI: 10.3390/s17071601
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Characterization of the β-CD-Ag/PVA substrate. (a) Photograph of the β-CD-Ag/PVA sol before hydrogel formation; (b) Photograph of the β-CD-Ag/PVA hydrogel; (c,d) SEM images of the β-CD-Ag/PVA hydrogel at different magnification.
Figure 2(a) Comparison of the SERS performance for SH detection with different substrates; (b) Adsorption kinetics of SH on PVA, Ag/PVA and β-CD-Ag/PVA (298 K).
Figure 3(a) UV-Vis spectra of Ag NP suspensions prepared with different synthesis pH values ranging from 9.5 to 13.35; (b) SERS spectra of SH on β-CD/Ag NPs synthesized at different pH values; (c) UV-Vis spectra of Ag NPs obtained at different ratio of AgNO3 and β-CD: 0.2:1, 0.5:1, 1:1, 3:1, 6:1, and 12:1; (d) TEM micrograph of β-CD/Ag obtained with AgNO3 and β-CD at a ratio of 1:1 ratio.
Figure 4SERS intensity change of SH at 1594 cm−1 using β-CD-Ag/PVA with different densities of β-CD-Ag NPs (from 1 to 9 mmol·L−1).
Figure 5Raman responses of the substrate β-CD-Ag/PVA after being soaked in water at different pH values.
Figure 6(a) SERS spectra of different concentrations of SH; (b) A correlation between the Raman intensity at 1594 cm−1 and the concentration of SH. The inset in (b) showed the linear range within 7–150 µg·mL−1 between the intensity at 1594 cm−1 and the concentration of SH.
Figure 7(a) Results of confirming the selectivity for SH detection; (b) SERS intensity (at the 1594 cm−1 peak) of 100 μg mL−1 SH on eight batches of β-CD-Ag/PVA substrates, five spots were randomly chosen for each batch of substrate.
Detection of positive samples with both SERS and HPLC method. ms: the detected mass of SH in capsules sample, ma: mass of SH added to capsules, md: the detected mass of SH. ω% represents the recovery of detected SH compared to its real concentration. ϖ% is the mean value of ω% for three times of detection.
| ms/mg | ma/mg | md/mg | ω%/SERS | ϖ%/SERS | RSD% | ϖ%/HPLC |
|---|---|---|---|---|---|---|
| 0.504 | 0.400 | 0.885 | 97.9 | 96.7 | 4.1 | 99.3 |
| 0.502 | 0.400 | 0.860 | 95.4 | |||
| 0.523 | 0.500 | 1.119 | 109.4 | 103.7 | 98.9 | |
| 0.582 | 0.500 | 1.060 | 98.0 | |||
| 0.528 | 0.600 | 1.089 | 96.6 | 104.2 | 100.2 | |
| 0.471 | 0.600 | 1.196 | 111.7 |
Detection results of six negative real samples with SH addition.
| Samples | Theoretical/mg | Detection/mg | ω%/SERS | ϖ%/SERS |
|---|---|---|---|---|
| 1 | 0.500 | 0.473 | 94.9 | 99.5 |
| 0.500 | 0.548 | 109.6 | ||
| 0.500 | 0.470 | 94.1 | ||
| 2 | 0.500 | 0.467 | 93.3 | 93.8 |
| 0.500 | 0.489 | 97.9 | ||
| 0.500 | 0.452 | 90.3 | ||
| 3 | 0.500 | 0.471 | 94.1 | 96.3 |
| 0.500 | 0.468 | 93.5 | ||
| 0.500 | 0.507 | 101.3 | ||
| 4 | 0.500 | 0.483 | 96.6 | 101.1 |
| 0.500 | 0.483 | 96.7 | ||
| 0.500 | 0.548 | 109.5 | ||
| 5 | 0.500 | 0.549 | 109.8 | 104.8 |
| 0.500 | 0.507 | 101.3 | ||
| 0.500 | 0.516 | 103.1 | ||
| 6 | 0.500 | 0.484 | 96.8 | 95.6 |
| 0.500 | 0.486 | 97.1 | ||
| 0.500 | 0.466 | 93.1 |