Thermal Effect of J-Plasma® Energy in a Porcine Tissue Model: Implications for Minimally Invasive Surgery.

Jasmine D Pedroso1, Melissa M Gutierrez1, K Warren Volker1, David L Howard1.   

Abstract

OBJECTIVE: To evaluate tissue effect of J-Plasma® (Bovie Medical Corporation, Clearwater, Florida) in porcine liver, kidney, muscle, ovarian, and uterine tissue blocks.
DESIGN: Prospective study utilizing porcine tissue blocks to evaluate the thermal spread of J-Plasma® device on liver, kidney, muscle, ovarian, and uterine tissue at various power settings, gas flow, and exposure times.
MATERIALS AND METHODS: J-Plasma® helium was used in porcine liver, kidney, and muscle tissue at 20%, 50%, and 100% power, and 1 L/min, 3 L/min, and 5 L/min gas flow at one, five, and 10-second intervals. J-Plasma® was then used in ovarian and uterine tissue at maximum power and gas flow settings in intervals of one, five, 10, and 30 seconds. Histologic evaluation of each tissue was then performed to measure thermal spread.
RESULTS: Regardless of tissue type, increased power setting, gas flow rate, and exposure time correlated with greater depth of thermal spread in liver, kidney, and muscle tissue. J-Plasma® did not exceed 2 mm thermal spread on liver, kidney, muscle, ovarian, and uterine tissue, even at a maximum setting of 100% power and 5 L/min gas flow after five seconds. Prolonged exposure to J-Plasma® of up to 30 seconds resulted in increased length and width of thermal spread of up to 12 mm, but did not result in significantly increased depth at 2.84 mm.
CONCLUSIONS: The J-Plasma® helium device has minimal lateral and depth of thermal spread in a variety of tissue types and can likely be used for a multitude of gynecologic surgical procedures. However, further studies are needed to demonstrate device safety in a clinical setting.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28693047

Source DB:  PubMed          Journal:  Surg Technol Int        ISSN: 1090-3941


  5 in total

1.  A Prospective Trial: Handsfree Thermoregulated Bipolar Radiofrequency for Face and Neck Contouring.

Authors:  Erez Dayan; Anne Chapas; Joseph Marte; Christopher Chia; Spero Theodorou
Journal:  Plast Reconstr Surg Glob Open       Date:  2022-03-28

Review 2.  Targeting Strategies for the Combination Treatment of Cancer Using Drug Delivery Systems.

Authors:  Janel Kydd; Rahul Jadia; Praveena Velpurisiva; Aniket Gad; Shailee Paliwal; Prakash Rai
Journal:  Pharmaceutics       Date:  2017-10-14       Impact factor: 6.321

3.  Use of Laparoscopic and Laparotomic J-Plasma Handpiece in Gynecological Malignancies: Results From A Pilot Study in A Tertiary Care Center.

Authors:  Salvatore Gueli Alletti; Andrea Rosati; Vito Andrea Capozzi; Matteo Pavone; Alessandro Gioè; Stefano Cianci; Vito Chiantera; Giuseppe Vizzielli; Giulia Scaglione; Anna Fagotti; Giovanni Scambia
Journal:  Front Oncol       Date:  2022-06-28       Impact factor: 5.738

4.  Helium Plasma Skin Regeneration: Evaluation of Skin Tissue Effects in a Porcine Model and Comparison to Nitrogen Plasma Skin Regeneration.

Authors:  J David Holcomb; Adrienne Schucker
Journal:  Lasers Surg Med       Date:  2019-10-06       Impact factor: 4.025

5.  Utilization of low-temperature helium plasma (J-Plasma) for dissection and hemostasis during carotid endarterectomy.

Authors:  Konstantinos Filis; George Galyfos; Fragiska Sigala; Georgios Zografos
Journal:  J Vasc Surg Cases Innov Tech       Date:  2020-03-03