Cone-morse implant connection system significantly reduces bacterial leakage between implant and abutment: an in vitro study.

A Baj1,2, A Bolzoni1,2, A Russillo1,2, D Lauritano3, A Palmieri4, F Cura4, F J Silvestre5, A B Giannì1,2.   

Abstract

Osseointegrated implants are very popular dental treatments today in the world. In osseointegrated implants, the occlusal forces are transmitted from prosthesis through an abutment to a dental implant. The abutment is connected to the implant by mean of a screw. A screw is the most used mean for connecting an implant to an abutment. Frequently the screws break and are lost. There is an alternative to screw retained abutment systems: the cone-morse connection (CMC). The CMC, thanks to the absence of the abutment screw, guarantees no micro-gaps, no micro-movements, and a reduction of bacterial leakage between implant and abutment. As P. gingivalis and T. forsythia penetration might have clinical relevance, it was the purpose of this investigation to evaluate molecular leakage of these two bacteria in a new CMC implants systems (Leone Spa®, Florence, Italy). To identify the capability of the implant to protect the internal space from the external environment, the passage of genetically modified Escherichia coli across implant-abutment interface was evaluated. Four cone-morse Leone implants (Leone® Spa, Florence, Italy) were immerged in a bacterial culture for 24 h and bacteria amount was then measured inside implant-abutment interface with Real-time PCR. Bacteria were detected inside all studied implants, with a median percentage of 3% for P. gingivalis and 4% for T. forsythia. Cone-morse connection implant system has very low bacterial leakage percentage and is similar to one-piece implants.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28691474

Source DB:  PubMed          Journal:  J Biol Regul Homeost Agents        ISSN: 0393-974X            Impact factor:   1.711


  3 in total

Review 1.  Implant-abutment connections on single crowns: a systematic review.

Authors:  F M Ceruso; P Barnaba; S Mazzoleni; L Ottria; M Gargari; A Zuccon; G Bruno; A DI Fiore
Journal:  Oral Implantol (Rome)       Date:  2017-01-21

Review 2.  Review of the Mechanical Behavior of Different Implant-Abutment Connections.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Paula López-Jarana; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2020-11-23       Impact factor: 3.390

3.  Bacterial Microleakage at the Implant-Abutment Interface: An In Vitro Study.

Authors:  Simonetta D'Ercole; Tatiane Cristina Dotta; Marzieh Ramezani Farani; Niloofar Etemadi; Giovanna Iezzi; Luca Comuzzi; Adriano Piattelli; Morena Petrini
Journal:  Bioengineering (Basel)       Date:  2022-06-26
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.